IEEE International Instrumentation and Measurement Technology Conference
Spatial-Light-Modulator-Based Axial Energy- Shaping Optimization Method for Probe Light in SERF Atomic Magnetometers
Fangzhou Ling, Xiujie Fang, Kun CHEN, Yueshen Wang, Jiayi Wei, Xin Xu, Bangcheng Han, Danyue Ma
Abstract
This paper proposes a wavefront-shaping approach for weak magnetic-field detection in spin-exchangerelaxation-free (SERF) atomic magnetometers. A spatial light modulator (SLM) is employed to tailor the axial energy distribution of the probe beam inside the vapor cell, aiming to enhance the effective sensitivity to ultra-weak magneto-optical rotation signals. In near-resonant SERF operation, the probe beam undergoes non-negligible absorption along propagation, while the atomic spin polarization typically exhibits a “low–highextensible route to improving thelow” longitudinal profile. Motivated by this mismatch, we programmatically design the incident wavefront so that the incell probe intensity follows a similar “low–high–low” axial distribution, thereby maximizing the path-integrated Faraday rotation under a fixed photon budget. Theoretical modeling and numerical simulations show that, compared with a uniformly incident Gaussian probe, the wavefront-shaped probe can increase the path-integrated rotation signal by up to ~30% without increasing optical power. To provide experimental evidence that spatial intensity engineering of the probe directly impacts the readout chain, we further perform a proof-of-concept transverse validation experiment by comparing Gaussian and flat-top probes generated using a beam-homogenizing waveplate. Under matched probe spot size, sensitivity spectra are measured at probe powers of 2, 4, 6, 8, and 10 mW. Across all tested powers, the flat-top probe consistently reduces the band-averaged sensitivity over 20–45 Hz by approximately 8%–17% relative to the Gaussian probe, indicating a lower effective noise floor and improved readout signal-to-noise ratio. These results support spatial beam shaping as a feasible and extensible route to improving the signal-to-noise ratio and effective sensitivity of SERF atomic magnetometers for ultra-weak magnetic-field measurements.