Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Log inSign up
Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Blog
Job Board
Q3AS

© 2026 Aqora Quantum S.A.S.

TermsPrivacyLegal Notice

arXiv:2606.23151

An Al$^+$ clock with $1.6\times10^{-18}$ systematic uncertainty and its frequency ratios

Fabian Dawel, Johannes Krämer, D. Drapier, Lennart Pelzer, Kai Dietze, M Haider Ali, Marek B. Hild, Vincent Barbé, S A King, J. Klose, Kilian Stahl, Johannes Rahm, Navraj Poudel, Sören Dörscher, S. Weyers, Erik Benkler, Christian Lisdat, Piet O. Schmidt

Published Jun 22, 2026• 3 citations
arXivHTMLPDF

Abstract

Advances in optical clocks motivate a redefinition of the second, requiring rigorous evaluations of systematic uncertainties and robust consistency among the clocks. Here, we report the full evaluation of the systematic frequency shifts of an $^{27}\mathrm{Al}^+$ single-ion clock, and the measurement of its absolute frequency and frequency ratio with a $^{87}$Sr optical lattice clock at PTB. The evaluated total systematic fractional frequency uncertainty is $1.6\times10^{-18}$, mainly limited by the accuracy of the relevant atomic coefficients and by background gas collisions. The absolute frequency of the clock has been measured to be $ν_{\mathrm{Al}^+}=1 121 015 393 207 859.19(24)\,$Hz, obtained by comparison with two primary caesium fountain clocks at PTB. The frequency ratio between the Al and Sr optical clocks has been determined to be $ν_{\mathrm{Al}^+}/ν_{^{87}\mathrm{Sr}}=2.611 701 431 781 462 668(36)$, limited by the accuracy of the Sr clock. This ratio differs by $8.6σ$ and $1.2σ$ from the 2021 and 2025 frequency ratio published by the BACON collaboration, respectively. These results represent an important contribution toward a future redefinition of the second using optical clocks, and underscore the importance of independent measurements of clock-candidate frequency ratios across different institutions.

DOI
10.48550/arxiv.2606.23151

Related papers

Frequency ratio measurement of Al⁺ at the 10⁻¹⁸ level

Fabian Dawel

The invention of the atomic clock has enabled applications like the global navigation satellite system, are important for telecommunication and are used for the definition of the SI second. Nowadays, optical clocks allow for measurement of height differences on the centimeter scale and can be employed for tests of the standard model of particle physics. Among the various implementations of optical clocks, this thesis focuses on an Al⁺ clock. For this work an optical resonator is examined, which enables the simultaneous frequency stabilization of two laser on the same mirror pair through the use of a dual-wavelength coating. Various contributions on the frequency stability are characterized and correlations are investigated. This cavity enables high fidelities in coherent manipulation of trapped ions, as required to operate the Al⁺ clock. This thesis partly builds upon the findings presented in Ref. Kramer, 2023. The setup and the systematic shifts are revisited and a detailed evaluation of the shifts is conducted, resulting in a total fractional frequency uncertainty of 1.6×10⁻¹⁸. The biggest contribution is from an ac-Stark shift induced by cooling during the clock interrogation, resulting in a shift of −9.3(1.1) × 10⁻¹⁸. The cooling reduces the second order Doppler shift to −1.6(0.2)×10⁻¹⁸. Additionally, a position dependent micromotion of the ion was investigated, which contributes an uncertainty of 0.38 × 10⁻¹⁸. Moreover, the first-order Doppler shift, servo error, ac-Stark shift of the clock laser and the phase chirp are determined for the first time, showing uncertainties below 10⁻¹⁸. The first frequency comparison between the Al⁺ clock and the ¹¹⁵In⁺, ¹⁷¹Yb⁺ and ⁸⁷Sr clocks at PTB are demonstrated. Measurements against Sr reveal a stability of the Al+ clock of 5.9 × 10⁻¹⁶ at 1s , which is one of the best single ion clock stabilities to date. All measured frequency ratios fall within the 10⁻¹⁸ uncertainty regime. The ratio between Sr/Al⁺ exhibits a significant deviation of more than 8.6𝜎 from published measurement data, underscoring the importance of inter-institutional frequency ratio measurements to verify the uncertainty budgets and reproducibility of optical clocks.

•Jan 1, 2026•0 citations

Atomic clock frequency ratios with fractional uncertainty $\leq 3.2 \times 10^{-18}$

Alexander Aeppli, Willa J. Arthur-Dworschack, Kyle Beloy +24 more

We report high-precision frequency ratio measurements between optical atomic clocks based on $^{27}$Al$^+$, $^{171}$Yb, and $^{87}$Sr. With total fractional uncertainties at or below $3.2 \times 10^{-18}$, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in $^{87}$Sr ratios at approximately $1\times10^{-16}$ and the Al$^+$/Yb ratio at $1.6\times10^{-17}$ in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key innovation in this work is the use of a common ultrastable reference delivered to all clocks via a 3.6 km phase-stabilized fiber link between two institutions. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2 to 3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of $1.3 \times 10^{-16}$ at 1 second. By enabling faster comparisons, this stability will improve sensitivity to non-white noise processes and other underlying limits of state-of-the-art optical frequency standards.

physics.atom-phCross-listedarXiv:2512.21428•Physical Review Letters•Dec 26, 2025•11 citations

Frequency Ratio of Hg 199 and Sr 87 Optical Lattice Clocks beyond the SI Limit

K. Yamanaka, Noriaki Ohmae, Ichiro Ushijima +2 more

We report on a frequency ratio measurement of a $^{199}\mathrm{Hg}$-based optical lattice clock referencing a $^{87}\mathrm{Sr}$-based clock. Evaluations of lattice light shift, including atomic-motion-dependent shift, enable us to achieve a total systematic uncertainty of $7.2\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$ for the Hg clock. The frequency ratio is measured to be ${\ensuremath{\nu}}_{\mathrm{Hg}}/{\ensuremath{\nu}}_{\mathrm{Sr}}=2.629\text{ }314\text{ }209\text{ }898\text{ }909\text{ }60(22)$ with a fractional uncertainty of $8.4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$, which is smaller than the uncertainty of the realization of the International System of Units (SI) second, i.e., the SI limit.

arXiv:1503.07941•Physical Review Letters•Jun 10, 2015•106 citations

Improved Frequency Measurement of a One-Dimensional Optical Lattice Clock with a Spin-Polarized Fermionic87Sr Isotope

Masao Takamoto, Feng-Lei Hong, Ryoichi Higashi +3 more

We demonstrate a one-dimensional optical lattice clock with a spin-polarized fermionic isotope designed to realize a collision-shift-free atomic clock with neutral atom ensembles. To reduce systematic uncertainties, we developed both Zeeman shift and vector light-shift cancellation techniques. By introducing both an H-maser and a Global Positioning System (GPS) carrier phase link, the absolute frequency of the $^1S_0(F=9/2) - {}^3P_0(F=9/2)$ clock transition of the $^{87}$Sr optical lattice clock is determined as 429,228,004,229,875(4) Hz, where the uncertainty is mainly limited by that of the frequency link. The result indicates that the Sr lattice clock will play an important role in the scope of the redefinition of the ``second'' by optical frequency standards.

arXiv:physics/0608212•Journal of the Physical Society of Japan•Oct 10, 2006•129 citations

Evaluation of a Sr+88 Optical Clock with a Direct Measurement of the Blackbody Radiation Shift and Determination of the Clock Frequency

Martin Steinel, H. Shao, Melina Filzinger +7 more

We report on an evaluation of an optical clock that uses the ^{2}S_{1/2}→^{2}D_{5/2} transition of a single ^{88}Sr^{+} ion as the reference. In contrast to previous work, we estimate the effective temperature of the blackbody radiation that shifts the reference transition directly during operation from the corresponding frequency shift and the well-characterized sensitivity to thermal radiation. We measure the clock output frequency against an independent ^{171}Yb^{+} ion clock, based on the ^{2}S_{1/2}(F=0)→^{2}F_{7/2}(F=3) electric octupole (E3) transition, and determine the frequency ratio with a total fractional uncertainty of 2.3×10^{-17}. Relying on a previous measurement of the ^{171}Yb^{+} (E3) clock frequency, we find the absolute frequency of the ^{88}Sr^{+} clock transition to be 444 779 044 095 485.277(59) Hz. Our result reduces the uncertainty by a factor of 3 compared with the previously most accurate measurement and may help to resolve so far inconsistent determinations of this value. We also show that for three simultaneously interrogated ^{88}Sr^{+} ions, the increased number causes the expected improvement of the short-term frequency instability of the optical clock without degrading its systematic uncertainty.

arXiv:2212.08687•Physical Review Letters•Aug 23, 2023•31 citations

Direct comparison of multi-ion optical clocks based on 40 Ca + and 88 Sr +

Yosef Sokolik, Roee Ozeri, Nitzan Akerman

We report the first direct frequency comparison between two multi-ion optical clocks based on the S$_{1/2}$ to D$_{5/2}$ transition in \Ca and \Sr ions. Using linear chains of up to nine \Ca ions and six \Sr ions, we demonstrate improved stability as a function of the number of ions that are contributing to the laser frequency stabilization servo. The measured joint fractional frequency stability of the two clocks reaches $1.37(12)\times 10^{-15}$ at one second, placing an upper bound on the same stability of one of the clocks at $9.6(8)\times 10^{-16}$ in one second. We measured the frequency ratio of the two clocks to be $R_{\text{Sr/Ca}}=1.082076536381896986(18)$, where the systematic uncertainty is primarily limited by the room temperature blackbody radiation. Our direct measurement represents an order of magnitude improvement compared to existing indirect frequency ratio measurements. Furthermore, by combining our results with recent absolute frequency measurements of the \Sr transition, referenced to a primary frequency standard, we refined the absolute frequency of the \Ca transition to $ν_{\text{Ca}^+}=411042129776400.21(4)$ Hz, reducing its uncertainty by a factor of three. This study presents the first direct comparison between two multi-ion optical clocks, highlighting their significant potential for future applications in fundamental physics tests, geodesy, and precision metrology.

arXiv:2602.05626•Physical Review Letters•Nov 9, 2026•3 citations
An Al+^++ clock with 1.6×10−181.6\times10^{-18}1.6×10−18 systematic uncertainty and its frequency ratios
Advances in optical clocks motivate a redefinition of the second, requiring rigorous evaluations of systematic uncertainties and robust consistency among the clocks. Here, we report the full evaluation of the systematic frequency shifts of an 27Al+^{27}\mathrm{Al}^+27Al+ single-ion clock, and the measurement of its absolute frequency and frequency ratio with a 87^{87}87Sr optical lattice clock at PTB. The evaluated total systematic fractional frequency uncertainty is 1.6×10−181.6\times10^{-18}1.6×10−18, mainly limited by the accuracy of the relevant atomic coefficients and by background gas collisions. The absolute frequency of the clock has been measured to be νAl+=1121015393207859.19(24) ν_{\mathrm{Al}^+}=1 121 015 393 207 859.19(24)\,νAl+​=1121015393207859.19(24)Hz, obtained by comparison with two primary caesium fountain clocks at PTB. The frequency ratio between the Al and Sr optical clocks has been determined to be νAl+/ν87Sr=2.611701431781462668(36)ν_{\mathrm{Al}^+}/ν_{^{87}\mathrm{Sr}}=2.611 701 431 781 462 668(36)νAl+​/ν87Sr​=2.611701431781462668(36), limited by the accuracy of the Sr clock. This ratio differs by 8.6σ8.6σ8.6σ and 1.2σ1.2σ1.2σ from the 2021 and 2025 frequency ratio published by the BACON collaboration, respectively. These results represent an important contribution toward a future redefinition of the second using optical clocks, and underscore the importance of independent measurements of clock-candidate frequency ratios across different institutions.
The invention of the atomic clock has enabled applications like the global navigation satellite system, are important for telecommunication and are used for the definition of the SI second. Nowadays, optical clocks allow for measurement of height differences on the centimeter scale and can be employed for tests of the standard model of particle physics. Among the various implementations of optical clocks, this thesis focuses on an Al⁺ clock. For this work an optical resonator is examined, which enables the simultaneous frequency stabilization of two laser on the same mirror pair through the use of a dual-wavelength coating. Various contributions on the frequency stability are characterized and correlations are investigated. This cavity enables high fidelities in coherent manipulation of trapped ions, as required to operate the Al⁺ clock. This thesis partly builds upon the findings presented in Ref. Kramer, 2023. The setup and the systematic shifts are revisited and a detailed evaluation of the shifts is conducted, resulting in a total fractional frequency uncertainty of 1.6×10⁻¹⁸. The biggest contribution is from an ac-Stark shift induced by cooling during the clock interrogation, resulting in a shift of −9.3(1.1) × 10⁻¹⁸. The cooling reduces the second order Doppler shift to −1.6(0.2)×10⁻¹⁸. Additionally, a position dependent micromotion of the ion was investigated, which contributes an uncertainty of 0.38 × 10⁻¹⁸. Moreover, the first-order Doppler shift, servo error, ac-Stark shift of the clock laser and the phase chirp are determined for the first time, showing uncertainties below 10⁻¹⁸. The first frequency comparison between the Al⁺ clock and the ¹¹⁵In⁺, ¹⁷¹Yb⁺ and ⁸⁷Sr clocks at PTB are demonstrated. Measurements against Sr reveal a stability of the Al+ clock of 5.9 × 10⁻¹⁶ at 1s , which is one of the best single ion clock stabilities to date. All measured frequency ratios fall within the 10⁻¹⁸ uncertainty regime. The ratio between Sr/Al⁺ exhibits a significant deviation of more than 8.6𝜎 from published measurement data, underscoring the importance of inter-institutional frequency ratio measurements to verify the uncertainty budgets and reproducibility of optical clocks.
Atomic clock frequency ratios with fractional uncertainty ≤3.2×10−18\leq 3.2 \times 10^{-18}≤3.2×10−18
We report high-precision frequency ratio measurements between optical atomic clocks based on 27^{27}27Al+^++, 171^{171}171Yb, and 87^{87}87Sr. With total fractional uncertainties at or below 3.2×10−183.2 \times 10^{-18}3.2×10−18, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in 87^{87}87Sr ratios at approximately 1×10−161\times10^{-16}1×10−16 and the Al+^++/Yb ratio at 1.6×10−171.6\times10^{-17}1.6×10−17 in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key innovation in this work is the use of a common ultrastable reference delivered to all clocks via a 3.6 km phase-stabilized fiber link between two institutions. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2 to 3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of 1.3×10−161.3 \times 10^{-16}1.3×10−16 at 1 second. By enabling faster comparisons, this stability will improve sensitivity to non-white noise processes and other underlying limits of state-of-the-art optical frequency standards.
We report on a frequency ratio measurement of a 199Hg^{199}\mathrm{Hg}199Hg-based optical lattice clock referencing a 87Sr^{87}\mathrm{Sr}87Sr-based clock. Evaluations of lattice light shift, including atomic-motion-dependent shift, enable us to achieve a total systematic uncertainty of $7.2\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$ for the Hg clock. The frequency ratio is measured to be ${\ensuremath{\nu}}_{\mathrm{Hg}}/{\ensuremath{\nu}}_{\mathrm{Sr}}=2.629\text{ }314\text{ }209\text{ }898\text{ }909\text{ }60(22)$ with a fractional uncertainty of $8.4\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}17}$, which is smaller than the uncertainty of the realization of the International System of Units (SI) second, i.e., the SI limit.
We demonstrate a one-dimensional optical lattice clock with a spin-polarized fermionic isotope designed to realize a collision-shift-free atomic clock with neutral atom ensembles. To reduce systematic uncertainties, we developed both Zeeman shift and vector light-shift cancellation techniques. By introducing both an H-maser and a Global Positioning System (GPS) carrier phase link, the absolute frequency of the 1S0(F=9/2)−3P0(F=9/2)^1S_0(F=9/2) - {}^3P_0(F=9/2)1S0​(F=9/2)−3P0​(F=9/2) clock transition of the 87^{87}87Sr optical lattice clock is determined as 429,228,004,229,875(4) Hz, where the uncertainty is mainly limited by that of the frequency link. The result indicates that the Sr lattice clock will play an important role in the scope of the redefinition of the ``second'' by optical frequency standards.
We report on an evaluation of an optical clock that uses the ^{2}S_{1/2}→^{2}D_{5/2} transition of a single ^{88}Sr^{+} ion as the reference. In contrast to previous work, we estimate the effective temperature of the blackbody radiation that shifts the reference transition directly during operation from the corresponding frequency shift and the well-characterized sensitivity to thermal radiation. We measure the clock output frequency against an independent ^{171}Yb^{+} ion clock, based on the ^{2}S_{1/2}(F=0)→^{2}F_{7/2}(F=3) electric octupole (E3) transition, and determine the frequency ratio with a total fractional uncertainty of 2.3×10^{-17}. Relying on a previous measurement of the ^{171}Yb^{+} (E3) clock frequency, we find the absolute frequency of the ^{88}Sr^{+} clock transition to be 444 779 044 095 485.277(59) Hz. Our result reduces the uncertainty by a factor of 3 compared with the previously most accurate measurement and may help to resolve so far inconsistent determinations of this value. We also show that for three simultaneously interrogated ^{88}Sr^{+} ions, the increased number causes the expected improvement of the short-term frequency instability of the optical clock without degrading its systematic uncertainty.
We report the first direct frequency comparison between two multi-ion optical clocks based on the S1/2_{1/2}1/2​ to D5/2_{5/2}5/2​ transition in \Ca and \Sr ions. Using linear chains of up to nine \Ca ions and six \Sr ions, we demonstrate improved stability as a function of the number of ions that are contributing to the laser frequency stabilization servo. The measured joint fractional frequency stability of the two clocks reaches 1.37(12)×10−151.37(12)\times 10^{-15}1.37(12)×10−15 at one second, placing an upper bound on the same stability of one of the clocks at 9.6(8)×10−169.6(8)\times 10^{-16}9.6(8)×10−16 in one second. We measured the frequency ratio of the two clocks to be RSr/Ca=1.082076536381896986(18)R_{\text{Sr/Ca}}=1.082076536381896986(18)RSr/Ca​=1.082076536381896986(18), where the systematic uncertainty is primarily limited by the room temperature blackbody radiation. Our direct measurement represents an order of magnitude improvement compared to existing indirect frequency ratio measurements. Furthermore, by combining our results with recent absolute frequency measurements of the \Sr transition, referenced to a primary frequency standard, we refined the absolute frequency of the \Ca transition to νCa+=411042129776400.21(4)ν_{\text{Ca}^+}=411042129776400.21(4)νCa+​=411042129776400.21(4) Hz, reducing its uncertainty by a factor of three. This study presents the first direct comparison between two multi-ion optical clocks, highlighting their significant potential for future applications in fundamental physics tests, geodesy, and precision metrology.