Physical Review Applied

Quantum-logic-based 25 Mg + - 27 Al + optical frequency standard for the redefinition of the SI second

Zhiyu Ma, Ke Deng, Z. Y. Wang, W. Z. Wei, P. Hao, H. X. Zhang, Liren Pang, B. Wang, Fang-Fei Wu, H. L. Liu, W. H. Yuan, J. L. Chang, J. X. Zhang, Quanying Wu, Jie Zhang, Z. H. Lu

Published Apr 8, 2024• 20 citations
Publisher

Abstract

Optical clocks have been demonstrated to be good choices to redefine the SI ``second.'' With this goal, we have constructed a quantum-logic-based ${}^{27}$$\mathrm{Al}$${}^{+}$ ion optical clock with a fractional frequency uncertainty of $1.6\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}$. A ${}^{25}$$\mathrm{Mg}$${}^{+}$ ion is used as the logic ion for sympathetic cooling, state readout, and frequency-shift measurement of the ${}^{27}$$\mathrm{Al}$${}^{+}$ ion. The stability of the optical clock is $2.6\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}15}/\sqrt{\ensuremath{\tau}}$, measured with a self-comparison method.

DOI
10.1103/physrevapplied.21.044017

Related papers

Frequency Comparison of Two High-Accuracy Al + Optical Clocks

C. W. Chou, David B. Hume, Jeroen C. J. Koelemeij +2 more

We have constructed an optical clock with a fractional frequency inaccuracy of 8.6x10{-18}, based on quantum logic spectroscopy of an Al+ ion. A simultaneously trapped Mg+ ion serves to sympathetically laser cool the Al+ ion and detect its quantum state. The frequency of the {1}S{0}<-->{3}P{0} clock transition is compared to that of a previously constructed Al+ optical clock with a statistical measurement uncertainty of 7.0x10{-18}. The two clocks exhibit a relative stability of 2.8x10{-15}tau{-1/2}, and a fractional frequency difference of -1.8x10{-17}, consistent with the accuracy limit of the older clock.

arXiv:0911.4527•Physical Review Letters•Feb 17, 2010•917 citations

High-Stability Single-Ion Clock with 5.5 × 10 − 19 Systematic Uncertainty

Mason C. Marshall, Daniel A. Rodriguez Castillo, Willa J. Arthur-Dworschack +10 more

We report a single-ion optical atomic clock with a fractional frequency uncertainty of 5.5×10^{-19} and fractional frequency stability of 3.5×10^{-16}/sqrt[τ/s], based on quantum logic spectroscopy of a single ^{27}Al^{+} ion. A cotrapped ^{25}Mg^{+} ion provides sympathetic cooling and quantum logic readout of the ^{27}Al^{+}^{1}S_{0}↔^{3}P_{0} clock transition. A Rabi probe duration of 1 s, enabled by laser stability transfer from a remote cryogenic silicon cavity across a 3.6 km fiber link, results in a threefold reduction in instability compared to previous ^{27}Al^{+} clocks. Systematic uncertainties are lower due to an improved ion trap electrical design, which reduces excess micromotion, and a new vacuum system, which reduces collisional shifts. We also perform a direction-sensitive measurement of the ac magnetic field due to the rf ion trap, eliminating systematic uncertainty due to field orientation.

arXiv:2504.13071•Physical Review Letters•Jul 14, 2025•91 citations

Al + 27 Quantum-Logic Clock with a Systematic Uncertainty below 10 − 18

Samuel M. Brewer, Jwo-Sy Chen, Aaron M. Hankin +5 more

We describe an optical atomic clock based on quantum-logic spectroscopy of the ^{1}S_{0}↔^{3}P_{0} transition in ^{27}Al^{+} with a systematic uncertainty of 9.4×10^{-19} and a frequency stability of 1.2×10^{-15}/sqrt[τ]. A ^{25}Mg^{+} ion is simultaneously trapped with the ^{27}Al^{+} ion and used for sympathetic cooling and state readout. Improvements in a new trap have led to reduced secular motion heating, compared to previous ^{27}Al^{+} clocks, enabling clock operation with ion secular motion near the three-dimensional ground state. Operating the clock with a lower trap drive frequency has reduced excess micromotion compared to previous ^{27}Al^{+} clocks. Both of these improvements have led to a reduced time-dilation shift uncertainty. Other systematic uncertainties including those due to blackbody radiation and the second-order Zeeman effect have also been reduced.

arXiv:1902.07694•Physical Review Letters•Jul 15, 2019•736 citations

Stability Improvement of an Aluminum Ion Optical Clock

B. Wang, Z. Y. Ma, Liren Pang +5 more

Based on the current initiative to redefine the SI “second”, the optical clock has been demonstrated by many laboratories as a favorable candidate due to its excellent performance indicators. We describe the evaluation of uncertainty and optimization of stability in an aluminum-ion optical clock system based on quantum logic spectroscopy. The uncertainty evaluation of the current optical clock system is$1.6\times 10^{-18}$. The frequency stability of a single system has been improved from 3.9$\times 10^{-15} / \sqrt{\tau}$to 2.6$\times 10^{-15} / \sqrt{\tau}$by using the servo-controlled self-comparison locking method.

European Frequency and Time Forum•Jun 25, 2024•0 citations

Towards a transportable aluminium ion quantum logic optical clock

S. Hannig, L. Pelzer, N. Scharnhorst +7 more

With the advent of optical clocks featuring fractional frequency uncertainties on the order of 10−17 and below, new applications such as chronometric leveling with few-centimeter height resolution emerge. We are developing a transportable optical clock based on a single trapped aluminum ion, which is interrogated via quantum logic spectroscopy. We employ singly charged calcium as the logic ion for sympathetic cooling, state preparation, and readout. Here, we present a simple and compact physics and laser package for manipulation of 40Ca+. Important features are a segmented multilayer trap with separate loading and probing zones, a compact titanium vacuum chamber, a near-diffraction-limited imaging system with high numerical aperture based on a single biaspheric lens, and an all-in-fiber 40Ca+ repump laser system. We present preliminary estimates of the trap-induced frequency shifts on 27Al+, derived from measurements with a single calcium ion. The micromotion-induced second-order Doppler shift for 27Al+ has been determined to be δνEMMν=−0.4−0.3+0.4×10−18 and the black-body radiation shift is δνBBR/ν = (−4.0 ± 0.4) × 10−18. Moreover, heating rates of 30 (7) quanta per second at trap frequencies of ωrad,Ca+ ≈ 2π × 2.5 MHz (ωax,Ca+ ≈ 2π × 1.5 MHz) in radial (axial) direction have been measured, enabling interrogation times of a few hundreds of milliseconds.

arXiv:1901.02250•Review of Scientific Instruments•May 1, 2019•70 citations

Ticking near the Zero-Point Energy: Towards 1 x 10^-18 Accuracy in Al^+ Optical Clocks

Jwo-Sy Chen

The advancement of optical atomic clocks in the past two decades has motivated many potential applications in navigation, communication, and science. Accurate optical clocks that outperform the current cesium time standard also trigger a discussion about a possible redefinition of the SI second. The 27Al+ quantum logic clocks developed at the National Institute of Standards and Technology (NIST) were the first to achieve the clock fractional frequency uncertainty below 10-17 in 2010. To date, a few research groups around the world have demonstrated optical clocks based on various atomic transitions with fractional frequency uncertainty in the 10-18 range. The accuracy of the previous 27Al+ clocks was limited by the second order Doppler (time dilation) shift and the blackbody shift. Since 2010, the third 27Al+ optical clock is being developed at NIST to achieve a higher accuracy. The frequency uncertainties dominant in the previous clocks are controlled and reduced in the current optical clock. The new design of the ion trap system has reduced significantly both the blackbody radiation shift uncertainty and the time dilation shift uncertainty due to micromotion, while the time dilation shift uncertainty due to the secular motion has been reduced by more than an order of magnitude by operating the optical clock near the three dimensional zero-point energy. Despite those previously dominant uncertainties, several other systematic effects are being evaluated and some other efforts are being made to achieve a total uncertainty towards 1.0 x 10-18. In this thesis, I will document the detail of the construction and evaluation of the current 27Al+ optical clock developed at NIST.

•Jan 1, 2017•4 citations
Optical clocks have been demonstrated to be good choices to redefine the SI ``second.'' With this goal, we have constructed a quantum-logic-based {}^{27}$$\mathrm{Al}$${}^{+} ion optical clock with a fractional frequency uncertainty of 1.6\ifmmode×\else\texttimes\fi10\ensuremath−181.6\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}18}1.6\ifmmode×\else\texttimes\fi10\ensuremath−18. A {}^{25}$$\mathrm{Mg}$${}^{+} ion is used as the logic ion for sympathetic cooling, state readout, and frequency-shift measurement of the {}^{27}$$\mathrm{Al}$${}^{+} ion. The stability of the optical clock is 2.6\ifmmode×\else\texttimes\fi10\ensuremath−15/\ensuremathτ2.6\ifmmode\times\else\texttimes\fi{}{10}^{\ensuremath{-}15}/\sqrt{\ensuremath{\tau}}2.6\ifmmode×\else\texttimes\fi10\ensuremath−15/\ensuremathτ​, measured with a self-comparison method.
We have constructed an optical clock with a fractional frequency inaccuracy of 8.6x10{-18}, based on quantum logic spectroscopy of an Al+ ion. A simultaneously trapped Mg+ ion serves to sympathetically laser cool the Al+ ion and detect its quantum state. The frequency of the {1}S{0}<-->{3}P{0} clock transition is compared to that of a previously constructed Al+ optical clock with a statistical measurement uncertainty of 7.0x10{-18}. The two clocks exhibit a relative stability of 2.8x10{-15}tau{-1/2}, and a fractional frequency difference of -1.8x10{-17}, consistent with the accuracy limit of the older clock.
We report a single-ion optical atomic clock with a fractional frequency uncertainty of 5.5×10^{-19} and fractional frequency stability of 3.5×10^{-16}/sqrt[τ/s], based on quantum logic spectroscopy of a single ^{27}Al^{+} ion. A cotrapped ^{25}Mg^{+} ion provides sympathetic cooling and quantum logic readout of the ^{27}Al^{+}^{1}S_{0}↔^{3}P_{0} clock transition. A Rabi probe duration of 1 s, enabled by laser stability transfer from a remote cryogenic silicon cavity across a 3.6 km fiber link, results in a threefold reduction in instability compared to previous ^{27}Al^{+} clocks. Systematic uncertainties are lower due to an improved ion trap electrical design, which reduces excess micromotion, and a new vacuum system, which reduces collisional shifts. We also perform a direction-sensitive measurement of the ac magnetic field due to the rf ion trap, eliminating systematic uncertainty due to field orientation.
We describe an optical atomic clock based on quantum-logic spectroscopy of the ^{1}S_{0}↔^{3}P_{0} transition in ^{27}Al^{+} with a systematic uncertainty of 9.4×10^{-19} and a frequency stability of 1.2×10^{-15}/sqrt[τ]. A ^{25}Mg^{+} ion is simultaneously trapped with the ^{27}Al^{+} ion and used for sympathetic cooling and state readout. Improvements in a new trap have led to reduced secular motion heating, compared to previous ^{27}Al^{+} clocks, enabling clock operation with ion secular motion near the three-dimensional ground state. Operating the clock with a lower trap drive frequency has reduced excess micromotion compared to previous ^{27}Al^{+} clocks. Both of these improvements have led to a reduced time-dilation shift uncertainty. Other systematic uncertainties including those due to blackbody radiation and the second-order Zeeman effect have also been reduced.
Based on the current initiative to redefine the SI “second”, the optical clock has been demonstrated by many laboratories as a favorable candidate due to its excellent performance indicators. We describe the evaluation of uncertainty and optimization of stability in an aluminum-ion optical clock system based on quantum logic spectroscopy. The uncertainty evaluation of the current optical clock system is1.6×10−181.6\times 10^{-18}1.6×10−18. The frequency stability of a single system has been improved from 3.9×10−15/τ\times 10^{-15} / \sqrt{\tau}×10−15/τ​to 2.6×10−15/τ\times 10^{-15} / \sqrt{\tau}×10−15/τ​by using the servo-controlled self-comparison locking method.
With the advent of optical clocks featuring fractional frequency uncertainties on the order of 10−17 and below, new applications such as chronometric leveling with few-centimeter height resolution emerge. We are developing a transportable optical clock based on a single trapped aluminum ion, which is interrogated via quantum logic spectroscopy. We employ singly charged calcium as the logic ion for sympathetic cooling, state preparation, and readout. Here, we present a simple and compact physics and laser package for manipulation of 40Ca+. Important features are a segmented multilayer trap with separate loading and probing zones, a compact titanium vacuum chamber, a near-diffraction-limited imaging system with high numerical aperture based on a single biaspheric lens, and an all-in-fiber 40Ca+ repump laser system. We present preliminary estimates of the trap-induced frequency shifts on 27Al+, derived from measurements with a single calcium ion. The micromotion-induced second-order Doppler shift for 27Al+ has been determined to be δνEMMν=−0.4−0.3+0.4×10−18 and the black-body radiation shift is δνBBR/ν = (−4.0 ± 0.4) × 10−18. Moreover, heating rates of 30 (7) quanta per second at trap frequencies of ωrad,Ca+ ≈ 2π × 2.5 MHz (ωax,Ca+ ≈ 2π × 1.5 MHz) in radial (axial) direction have been measured, enabling interrogation times of a few hundreds of milliseconds.
The advancement of optical atomic clocks in the past two decades has motivated many potential applications in navigation, communication, and science. Accurate optical clocks that outperform the current cesium time standard also trigger a discussion about a possible redefinition of the SI second. The 27Al+ quantum logic clocks developed at the National Institute of Standards and Technology (NIST) were the first to achieve the clock fractional frequency uncertainty below 10-17 in 2010. To date, a few research groups around the world have demonstrated optical clocks based on various atomic transitions with fractional frequency uncertainty in the 10-18 range. The accuracy of the previous 27Al+ clocks was limited by the second order Doppler (time dilation) shift and the blackbody shift. Since 2010, the third 27Al+ optical clock is being developed at NIST to achieve a higher accuracy. The frequency uncertainties dominant in the previous clocks are controlled and reduced in the current optical clock. The new design of the ion trap system has reduced significantly both the blackbody radiation shift uncertainty and the time dilation shift uncertainty due to micromotion, while the time dilation shift uncertainty due to the secular motion has been reduced by more than an order of magnitude by operating the optical clock near the three dimensional zero-point energy. Despite those previously dominant uncertainties, several other systematic effects are being evaluated and some other efforts are being made to achieve a total uncertainty towards 1.0 x 10-18. In this thesis, I will document the detail of the construction and evaluation of the current 27Al+ optical clock developed at NIST.