IEEE MTT-S International Microwave Symposium digest

Automatic 4 K Cryogenic Probe Station for DC and Microwave Measurements on 150-mm and 200-mm Wafers

Joshua T. West, A. Kurlej, Alex Wynn, Chad Rogers, M.A. Gouker, Sergey K. Tolpygo

Published Jun 19, 2022• 3 citations
Publisher

Abstract

Robust maturation of cryogenic electronics has been limited by the lack of high-throughput measurement capabilities, especially for electronics that operate at 4 K. Automated on-wafer probing to characterize the dc and RF performance of components and complete circuits has been a critical element in the development of room temperature electronics. We report here the development of a fully automated wafer prober that operates with the wafer cooled to a temperature of 4 K and has the ability to null the ambient magnetic field to below 100 nT, which is important for measuring superconducting circuits with the wafer prober.

DOI
10.1109/ims37962.2022.9865525

Related papers

Wafer-Scale Characterization of a Superconductor Integrated Circuit Fabrication Process, Using a Cryogenic Wafer Prober

Joshua T. West, A. Kurlej, Alex Wynn +3 more

Using a fully automated cryogenic wafer prober, we measured superconductor fabrication process control monitors and simple integrated circuits on 200-mm wafers at 4.4 K, including SQIF-based magnetic field sensors, SQUID-based circuits for measuring inductors, Nb/Al-AlOx/Nb Josephson junctions, test structures for measuring critical current of superconducting wires and vias, resistors, etc., to demonstrate the feasibility of using the system for characterizing niobium superconducting devices and integrated circuits on a wafer scale. Data on the wafer-scale distributions of the residual magnetic field, junction tunnel resistance, energy gap, inductance of multiple Nb layers, and critical currents of interlayer vias are presented. A comparison with existing models is made. The wafers were fabricated in the SFQ5ee process, the fully planarized process with eight niobium layers and a layer of kinetic inductors, developed for superconductor electronics at MIT Lincoln Laboratory, Lexington, MA, USA. The cryogenic wafer prober was developed at HPD/FormFactor, Inc., Boulder, CO, USA.

arXiv:2112.00705•IEEE Transactions on Applied Superconductivity•May 5, 2022•18 citations

A compact and versatile cryogenic probe station for quantum device testing

Mathieu de Kruijf, Simon Geyer, Toni Berger +4 more

Fast feedback from cryogenic electrical characterization measurements is key for the development of scalable quantum computing technology. At room temperature, high-throughput device testing is accomplished with a probe-based solution, where electrical probes are repeatedly positioned onto devices for acquiring statistical data. In this work, we present a probe station that can be operated from room temperature down to below 2 K. Its small size makes it compatible with standard cryogenic measurement setups with a magnet. A large variety of electronic devices can be tested. Here, we demonstrate the performance of the prober by characterizing silicon fin field-effect transistors as a host for quantum dot spin qubits. Such a tool can massively accelerate the design-fabrication-measurement cycle and provide important feedback for process optimization toward building scalable quantum circuits.

arXiv:2212.12369•Review of Scientific Instruments•May 1, 2023•7 citations

Broadband four-channel cryogenic wafer probing system for automated coplanar on-chip measurements

G. J. Höfer, F. Jäger, H. A. Kratz

An automated measurement system for cryogenic wafer probing is described. The system has been designed for up to four simultaneous broadband on-chip measurements in the temperature range from 4 to 300 K. Special requirements for superconducting samples have been taken into consideration. Most of the measurements needed for characterization of superconducting thin films and circuits can be done with a high degree of precision in the low- and high-frequency region. Details of the design are presented and the system performance is discussed.

Review of Scientific Instruments•Mar 1, 1993•3 citations

Measurement setup for the characterization of integrated semiconductor circuits at cryogenic temperatures

Paul Julius Ritter, Marc-André Tucholke, Marius Neumann +8 more

Integrated semiconductor circuits operating at cryogenic temperatures can play a crucial role in scaling quantum computing architectures based on trapped-ion and superconducting qubits. Other technologies, such as low-temperature detectors, can also benefit significantly from these circuits. These applications demand high-frequency, cryogenic temperature measurement systems for the thorough characterization of semiconductor components and circuits. This work presents a customizable, high-frequency, fast, and reliable cryogenic measurement setup for measurement temperatures ranging from room temperature to 4.2 K. It features up to two dc probes or two ground-signal-ground probes for frequencies of up to 67 GHz, which can be configured in a 180° or 90° configuration. In addition, up to 48 BeCu wires configured as twisted pairs provide supply and control signals. Furthermore, up to four rf connections can be mounted directly to a printed circuit board. The setup features an optical microscope in the vacuum chamber to position the probes, which is further utilized by a machine vision algorithm, allowing the detection of pads and automatic positioning of the probes over the pads. The hardware is located in a 550 × 500 × 500 mm3 large vacuum chamber with two independent pulse tube cryocoolers with cooling powers of up to 0.9 and 0.4 W at 4.2 K. Exemplary room temperature and cryogenic S-parameter, transient frequency (fT), time-domain, capacitance-voltage, and dc measurements of single transistors and integrated circuits fabricated by Infineon are presented, demonstrating the system's measurement capability for quantum computing and other applications.

Review of Scientific Instruments•Apr 1, 2025•9 citations

Sub-kelvin cryogenic wafer probing verified by on-wafer primary Coulomb blockade thermometry

Lassi Lehtisyrjä, R. P. Loreto, Juho Luomahaara +11 more

Scalable cryogenic wafer level testing is emerging as an essential tool for the development of quantum devices and cryogenic microelectronics, but the main motivation, the actual operational on-wafer temperature, remains largely unverified. In this work, utilizing a cryogenic wafer prober (CWP) enables wafer-scale electrical testing of Coulomb blockade thermometers (CBTs), while the CBTs simultaneously provide an independent static and dynamic assessment of the prober's thermal performance through primary and in situ calibrated secondary electron thermometry, realizing a reciprocal characterization approach. Using recently introduced scalable normal metal TiW-based CBTs, which require no superconductivity suppressing magnetic fields, we have measured on-wafer electron temperatures of 630-660 mK across a full 150 mm wafer, representing, to our knowledge, record-low electron temperatures measured in a CWP. These exceed the nominal 600 mK chuck setpoint by approximately 40 mK on average, demonstrating the necessity of direct on-wafer thermometry. Self-calibrated secondary CBT thermometry resolves the thermal transients caused by chuck movement and reveals a thermal settling time within 5 minutes after probe touch down. This reciprocal wafer-scale approach simultaneously evaluates the device under test and the thermal performance of the measurement environment, providing a practical methodology for scalable sub-kelvin characterization of quantum devices and cryogenic microelectronics.

arXiv:2609.12596•Sep 11, 2026•0 citations

A Review of RF and Microwave Measurement Strategies for Characterizing Cryo-CMOS Technologies in Quantum Computing

Manoj Stanley, Murat Celep, Xiaobang Shang +1 more

Successful deployment of cryo-CMOS technologies supporting quantum computing systems requires radio frequency (RF) and microwave circuitry that operates at cryogenic temperatures. The performance of this circuitry needs to be validated through reliable in-situ measurement. Two measurement modalities exist for this purpose, using either dilution refrigerators or cryogenic on-wafer probe stations. In both cases, confidence is needed in the measurements so that reliable decisions can be made concerning the performance of the RF circuitry and related cryo-CMOS technology. This paper discusses recent advances and challenges in both measurement modalities and how these advances are facilitating developments with cryo-CMOS technologies.

International Conference on Electromagnetics in Advanced Applications (ICEAA)•Sep 8, 2025•0 citations
Robust maturation of cryogenic electronics has been limited by the lack of high-throughput measurement capabilities, especially for electronics that operate at 4 K. Automated on-wafer probing to characterize the dc and RF performance of components and complete circuits has been a critical element in the development of room temperature electronics. We report here the development of a fully automated wafer prober that operates with the wafer cooled to a temperature of 4 K and has the ability to null the ambient magnetic field to below 100 nT, which is important for measuring superconducting circuits with the wafer prober.
Using a fully automated cryogenic wafer prober, we measured superconductor fabrication process control monitors and simple integrated circuits on 200-mm wafers at 4.4 K, including SQIF-based magnetic field sensors, SQUID-based circuits for measuring inductors, Nb/Al-AlOx/Nb Josephson junctions, test structures for measuring critical current of superconducting wires and vias, resistors, etc., to demonstrate the feasibility of using the system for characterizing niobium superconducting devices and integrated circuits on a wafer scale. Data on the wafer-scale distributions of the residual magnetic field, junction tunnel resistance, energy gap, inductance of multiple Nb layers, and critical currents of interlayer vias are presented. A comparison with existing models is made. The wafers were fabricated in the SFQ5ee process, the fully planarized process with eight niobium layers and a layer of kinetic inductors, developed for superconductor electronics at MIT Lincoln Laboratory, Lexington, MA, USA. The cryogenic wafer prober was developed at HPD/FormFactor, Inc., Boulder, CO, USA.
Fast feedback from cryogenic electrical characterization measurements is key for the development of scalable quantum computing technology. At room temperature, high-throughput device testing is accomplished with a probe-based solution, where electrical probes are repeatedly positioned onto devices for acquiring statistical data. In this work, we present a probe station that can be operated from room temperature down to below 2 K. Its small size makes it compatible with standard cryogenic measurement setups with a magnet. A large variety of electronic devices can be tested. Here, we demonstrate the performance of the prober by characterizing silicon fin field-effect transistors as a host for quantum dot spin qubits. Such a tool can massively accelerate the design-fabrication-measurement cycle and provide important feedback for process optimization toward building scalable quantum circuits.
An automated measurement system for cryogenic wafer probing is described. The system has been designed for up to four simultaneous broadband on-chip measurements in the temperature range from 4 to 300 K. Special requirements for superconducting samples have been taken into consideration. Most of the measurements needed for characterization of superconducting thin films and circuits can be done with a high degree of precision in the low- and high-frequency region. Details of the design are presented and the system performance is discussed.
Integrated semiconductor circuits operating at cryogenic temperatures can play a crucial role in scaling quantum computing architectures based on trapped-ion and superconducting qubits. Other technologies, such as low-temperature detectors, can also benefit significantly from these circuits. These applications demand high-frequency, cryogenic temperature measurement systems for the thorough characterization of semiconductor components and circuits. This work presents a customizable, high-frequency, fast, and reliable cryogenic measurement setup for measurement temperatures ranging from room temperature to 4.2 K. It features up to two dc probes or two ground-signal-ground probes for frequencies of up to 67 GHz, which can be configured in a 180° or 90° configuration. In addition, up to 48 BeCu wires configured as twisted pairs provide supply and control signals. Furthermore, up to four rf connections can be mounted directly to a printed circuit board. The setup features an optical microscope in the vacuum chamber to position the probes, which is further utilized by a machine vision algorithm, allowing the detection of pads and automatic positioning of the probes over the pads. The hardware is located in a 550 × 500 × 500 mm3 large vacuum chamber with two independent pulse tube cryocoolers with cooling powers of up to 0.9 and 0.4 W at 4.2 K. Exemplary room temperature and cryogenic S-parameter, transient frequency (fT), time-domain, capacitance-voltage, and dc measurements of single transistors and integrated circuits fabricated by Infineon are presented, demonstrating the system's measurement capability for quantum computing and other applications.
Scalable cryogenic wafer level testing is emerging as an essential tool for the development of quantum devices and cryogenic microelectronics, but the main motivation, the actual operational on-wafer temperature, remains largely unverified. In this work, utilizing a cryogenic wafer prober (CWP) enables wafer-scale electrical testing of Coulomb blockade thermometers (CBTs), while the CBTs simultaneously provide an independent static and dynamic assessment of the prober's thermal performance through primary and in situ calibrated secondary electron thermometry, realizing a reciprocal characterization approach. Using recently introduced scalable normal metal TiW-based CBTs, which require no superconductivity suppressing magnetic fields, we have measured on-wafer electron temperatures of 630-660 mK across a full 150 mm wafer, representing, to our knowledge, record-low electron temperatures measured in a CWP. These exceed the nominal 600 mK chuck setpoint by approximately 40 mK on average, demonstrating the necessity of direct on-wafer thermometry. Self-calibrated secondary CBT thermometry resolves the thermal transients caused by chuck movement and reveals a thermal settling time within 5 minutes after probe touch down. This reciprocal wafer-scale approach simultaneously evaluates the device under test and the thermal performance of the measurement environment, providing a practical methodology for scalable sub-kelvin characterization of quantum devices and cryogenic microelectronics.
Successful deployment of cryo-CMOS technologies supporting quantum computing systems requires radio frequency (RF) and microwave circuitry that operates at cryogenic temperatures. The performance of this circuitry needs to be validated through reliable in-situ measurement. Two measurement modalities exist for this purpose, using either dilution refrigerators or cryogenic on-wafer probe stations. In both cases, confidence is needed in the measurements so that reliable decisions can be made concerning the performance of the RF circuitry and related cryo-CMOS technology. This paper discusses recent advances and challenges in both measurement modalities and how these advances are facilitating developments with cryo-CMOS technologies.