Sub-kelvin cryogenic wafer probing verified by on-wafer primary Coulomb blockade thermometry

Lassi Lehtisyrjä, R. P. Loreto, Juho Luomahaara, Jarno Järvinen, Tuure Rantanen, Juha Vikstedt, Janne S. Lehtinen, Matti Remes, Timo Salminen, Juuso Helander, Aki Junes, Pasi Aaltonen, Vesa Henttonen, Mika Prunnila

Published Sep 11, 2026• 0 citations
arXivHTMLPDF

Abstract

Scalable cryogenic wafer level testing is emerging as an essential tool for the development of quantum devices and cryogenic microelectronics, but the main motivation, the actual operational on-wafer temperature, remains largely unverified. In this work, utilizing a cryogenic wafer prober (CWP) enables wafer-scale electrical testing of Coulomb blockade thermometers (CBTs), while the CBTs simultaneously provide an independent static and dynamic assessment of the prober's thermal performance through primary and in situ calibrated secondary electron thermometry, realizing a reciprocal characterization approach. Using recently introduced scalable normal metal TiW-based CBTs, which require no superconductivity suppressing magnetic fields, we have measured on-wafer electron temperatures of 630-660 mK across a full 150 mm wafer, representing, to our knowledge, record-low electron temperatures measured in a CWP. These exceed the nominal 600 mK chuck setpoint by approximately 40 mK on average, demonstrating the necessity of direct on-wafer thermometry. Self-calibrated secondary CBT thermometry resolves the thermal transients caused by chuck movement and reveals a thermal settling time within 5 minutes after probe touch down. This reciprocal wafer-scale approach simultaneously evaluates the device under test and the thermal performance of the measurement environment, providing a practical methodology for scalable sub-kelvin characterization of quantum devices and cryogenic microelectronics.

DOI
10.48550/arxiv.2609.12596

Related papers

Automatic 4 K Cryogenic Probe Station for DC and Microwave Measurements on 150-mm and 200-mm Wafers

Joshua T. West, A. Kurlej, Alex Wynn +3 more

Robust maturation of cryogenic electronics has been limited by the lack of high-throughput measurement capabilities, especially for electronics that operate at 4 K. Automated on-wafer probing to characterize the dc and RF performance of components and complete circuits has been a critical element in the development of room temperature electronics. We report here the development of a fully automated wafer prober that operates with the wafer cooled to a temperature of 4 K and has the ability to null the ambient magnetic field to below 100 nT, which is important for measuring superconducting circuits with the wafer prober.

IEEE MTT-S International Microwave Symposium digest•Jun 19, 2022•3 citations

Radio-Frequency Coulomb-Blockade Thermometry

Florian Blanchet, Yu-Cheng Chang, Bayan Karimi +2 more

A Coulomb-blockade thermometer (CBT) provides calibration-free measurement of low temperatures, even in strong magnetic fields---but slowly. A standard conductance measurement with a CBT takes typically minutes, which is useless in applications where relatively fast changes in temperature are monitored. Besides, a slow measurement is prone to drifts, and low-frequency noise. The authors solve this issue by hooking a CBT sensor into a radio-frequency tank circuit, making the measurement about 1000 times faster. Their approach is expected to impact low-temperature thermometry, for $e.g.$ applications in quantum technology.

arXiv:2109.09677•Physical Review Applied•Jan 26, 2022•9 citations

Localized Thermometry via Dayem Bridges Integrated on Superconducting Qubit Chips

Ella O. Lachman, Dave P. Pappas, Jayss Marshall +1 more

Accurate knowledge of the on-chip temperature is essential for understanding and optimizing the performance of superconducting qubits, yet direct thermometry at millikelvin temperatures remains challenging. While qubits themselves are sensitive to the temperature of their environment, other factors may affect the qubits` effective temperature, and using them as thermometers with any accuracy requires specialized measurement protocols and qubit designs, limiting their practicality for routine diagnostics and adding complex infrastructure to any hardware testing apparatus. Here we demonstrate a complementary on-chip thermometry method based on superconducting Dayem bridges that are integrated on the same chip as transmon qubits. By extracting the critical current of the Dayem bridge from I-V measurements, we obtain a local, quantitative measure of the chip temperature without the need for microwave calibration or qubit-specific control sequences. To demonstrate the utility of the Dayem bridges as thermometers, we fabricate them in-situ with qubits on the same chip, calibrate the Dayem bridge critical current as a function of temperature, and characterize its resolution and stability at cryogenic temperatures. We additionally perform simultaneous measurements of the Dayem bridge thermometer and qubit excited-state population, and show agreement over the relevant temperature range, validating the method against established qubit thermometry. Furthermore, we correlate the independently measured chip temperature with qubit energy relaxation and dephasing times, demonstrating the utility of this approach for diagnosing temperature-dependent decoherence mechanisms. These results establish integrated Dayem bridges as a simple, non-invasive, and scalable tool for cryogenic hardware development, and on chip thermometry in superconducting quantum circuits.

quant-pharXiv:2607.06670•Jul 9, 2026•1 citation

Wafer-Scale Characterization of a Superconductor Integrated Circuit Fabrication Process, Using a Cryogenic Wafer Prober

Joshua T. West, A. Kurlej, Alex Wynn +3 more

Using a fully automated cryogenic wafer prober, we measured superconductor fabrication process control monitors and simple integrated circuits on 200-mm wafers at 4.4 K, including SQIF-based magnetic field sensors, SQUID-based circuits for measuring inductors, Nb/Al-AlOx/Nb Josephson junctions, test structures for measuring critical current of superconducting wires and vias, resistors, etc., to demonstrate the feasibility of using the system for characterizing niobium superconducting devices and integrated circuits on a wafer scale. Data on the wafer-scale distributions of the residual magnetic field, junction tunnel resistance, energy gap, inductance of multiple Nb layers, and critical currents of interlayer vias are presented. A comparison with existing models is made. The wafers were fabricated in the SFQ5ee process, the fully planarized process with eight niobium layers and a layer of kinetic inductors, developed for superconductor electronics at MIT Lincoln Laboratory, Lexington, MA, USA. The cryogenic wafer prober was developed at HPD/FormFactor, Inc., Boulder, CO, USA.

arXiv:2112.00705•IEEE Transactions on Applied Superconductivity•May 5, 2022•18 citations

A compact and versatile cryogenic probe station for quantum device testing

Mathieu de Kruijf, Simon Geyer, Toni Berger +4 more

Fast feedback from cryogenic electrical characterization measurements is key for the development of scalable quantum computing technology. At room temperature, high-throughput device testing is accomplished with a probe-based solution, where electrical probes are repeatedly positioned onto devices for acquiring statistical data. In this work, we present a probe station that can be operated from room temperature down to below 2 K. Its small size makes it compatible with standard cryogenic measurement setups with a magnet. A large variety of electronic devices can be tested. Here, we demonstrate the performance of the prober by characterizing silicon fin field-effect transistors as a host for quantum dot spin qubits. Such a tool can massively accelerate the design-fabrication-measurement cycle and provide important feedback for process optimization toward building scalable quantum circuits.

arXiv:2212.12369•Review of Scientific Instruments•May 1, 2023•7 citations

Low-temperature benchmarking of qubit control wires by primary electron thermometry

Elias Roos Hansen, Ferdinand Kuemmeth, Joost van der Heijden

Low-frequency qubit control wires require non-trivial thermal anchoring and low-pass filtering. The resulting electron temperature serves as a quality benchmark for these signal lines. In this technical note, we make use of a primary electron thermometry technique, using a Coulomb blockade thermometer, to establish the electron temperature in the millikelvin regime. The experimental four-probe measurement setup, the data analysis, and the measurement limitations are discussed in detail. We verify the results by also using another electron thermometry technique, based on a superconductor-insulator-normal metal junction. Our comparison of signal lines with QDevil's QFilter to unfiltered signal lines demonstrates that the filter significantly reduces both the rms noise and electron temperature, which is measured to be 22 $\pm$ 1 mK.

arXiv:2403.17720•Jan 1, 2024•0 citations
Scalable cryogenic wafer level testing is emerging as an essential tool for the development of quantum devices and cryogenic microelectronics, but the main motivation, the actual operational on-wafer temperature, remains largely unverified. In this work, utilizing a cryogenic wafer prober (CWP) enables wafer-scale electrical testing of Coulomb blockade thermometers (CBTs), while the CBTs simultaneously provide an independent static and dynamic assessment of the prober's thermal performance through primary and in situ calibrated secondary electron thermometry, realizing a reciprocal characterization approach. Using recently introduced scalable normal metal TiW-based CBTs, which require no superconductivity suppressing magnetic fields, we have measured on-wafer electron temperatures of 630-660 mK across a full 150 mm wafer, representing, to our knowledge, record-low electron temperatures measured in a CWP. These exceed the nominal 600 mK chuck setpoint by approximately 40 mK on average, demonstrating the necessity of direct on-wafer thermometry. Self-calibrated secondary CBT thermometry resolves the thermal transients caused by chuck movement and reveals a thermal settling time within 5 minutes after probe touch down. This reciprocal wafer-scale approach simultaneously evaluates the device under test and the thermal performance of the measurement environment, providing a practical methodology for scalable sub-kelvin characterization of quantum devices and cryogenic microelectronics.
Robust maturation of cryogenic electronics has been limited by the lack of high-throughput measurement capabilities, especially for electronics that operate at 4 K. Automated on-wafer probing to characterize the dc and RF performance of components and complete circuits has been a critical element in the development of room temperature electronics. We report here the development of a fully automated wafer prober that operates with the wafer cooled to a temperature of 4 K and has the ability to null the ambient magnetic field to below 100 nT, which is important for measuring superconducting circuits with the wafer prober.
A Coulomb-blockade thermometer (CBT) provides calibration-free measurement of low temperatures, even in strong magnetic fields---but slowly. A standard conductance measurement with a CBT takes typically minutes, which is useless in applications where relatively fast changes in temperature are monitored. Besides, a slow measurement is prone to drifts, and low-frequency noise. The authors solve this issue by hooking a CBT sensor into a radio-frequency tank circuit, making the measurement about 1000 times faster. Their approach is expected to impact low-temperature thermometry, for e.g.e.g.e.g. applications in quantum technology.
Accurate knowledge of the on-chip temperature is essential for understanding and optimizing the performance of superconducting qubits, yet direct thermometry at millikelvin temperatures remains challenging. While qubits themselves are sensitive to the temperature of their environment, other factors may affect the qubits` effective temperature, and using them as thermometers with any accuracy requires specialized measurement protocols and qubit designs, limiting their practicality for routine diagnostics and adding complex infrastructure to any hardware testing apparatus. Here we demonstrate a complementary on-chip thermometry method based on superconducting Dayem bridges that are integrated on the same chip as transmon qubits. By extracting the critical current of the Dayem bridge from I-V measurements, we obtain a local, quantitative measure of the chip temperature without the need for microwave calibration or qubit-specific control sequences. To demonstrate the utility of the Dayem bridges as thermometers, we fabricate them in-situ with qubits on the same chip, calibrate the Dayem bridge critical current as a function of temperature, and characterize its resolution and stability at cryogenic temperatures. We additionally perform simultaneous measurements of the Dayem bridge thermometer and qubit excited-state population, and show agreement over the relevant temperature range, validating the method against established qubit thermometry. Furthermore, we correlate the independently measured chip temperature with qubit energy relaxation and dephasing times, demonstrating the utility of this approach for diagnosing temperature-dependent decoherence mechanisms. These results establish integrated Dayem bridges as a simple, non-invasive, and scalable tool for cryogenic hardware development, and on chip thermometry in superconducting quantum circuits.
Using a fully automated cryogenic wafer prober, we measured superconductor fabrication process control monitors and simple integrated circuits on 200-mm wafers at 4.4 K, including SQIF-based magnetic field sensors, SQUID-based circuits for measuring inductors, Nb/Al-AlOx/Nb Josephson junctions, test structures for measuring critical current of superconducting wires and vias, resistors, etc., to demonstrate the feasibility of using the system for characterizing niobium superconducting devices and integrated circuits on a wafer scale. Data on the wafer-scale distributions of the residual magnetic field, junction tunnel resistance, energy gap, inductance of multiple Nb layers, and critical currents of interlayer vias are presented. A comparison with existing models is made. The wafers were fabricated in the SFQ5ee process, the fully planarized process with eight niobium layers and a layer of kinetic inductors, developed for superconductor electronics at MIT Lincoln Laboratory, Lexington, MA, USA. The cryogenic wafer prober was developed at HPD/FormFactor, Inc., Boulder, CO, USA.
Fast feedback from cryogenic electrical characterization measurements is key for the development of scalable quantum computing technology. At room temperature, high-throughput device testing is accomplished with a probe-based solution, where electrical probes are repeatedly positioned onto devices for acquiring statistical data. In this work, we present a probe station that can be operated from room temperature down to below 2 K. Its small size makes it compatible with standard cryogenic measurement setups with a magnet. A large variety of electronic devices can be tested. Here, we demonstrate the performance of the prober by characterizing silicon fin field-effect transistors as a host for quantum dot spin qubits. Such a tool can massively accelerate the design-fabrication-measurement cycle and provide important feedback for process optimization toward building scalable quantum circuits.
Low-frequency qubit control wires require non-trivial thermal anchoring and low-pass filtering. The resulting electron temperature serves as a quality benchmark for these signal lines. In this technical note, we make use of a primary electron thermometry technique, using a Coulomb blockade thermometer, to establish the electron temperature in the millikelvin regime. The experimental four-probe measurement setup, the data analysis, and the measurement limitations are discussed in detail. We verify the results by also using another electron thermometry technique, based on a superconductor-insulator-normal metal junction. Our comparison of signal lines with QDevil's QFilter to unfiltered signal lines demonstrates that the filter significantly reduces both the rms noise and electron temperature, which is measured to be 22 ±\pm± 1 mK.