Slow Light Imaging Spectroscopy (SLIS): A Comparison of Different Filters

Amirhossein Abbasszadeh Rad, Anuj Rekhy, Arthur Dogariu, Richard B. Miles

Published Jun 8, 2020• 7 citations
Publisher

Abstract

In this study, we model four atomic filters useful in slow light imaging spectroscopy. We calculate delay using the group velocity in rubidium at 780 nm, mercury at 254 nm, and cesium at 852 nm and 387.6 nm at different temperatures and path lengths of the atomic filter. We also calculate the output of a 5 ns pulse by propagating its Fourier components through cells. Then we compare the results of these two numerical models with the ideal limit set only by the natural linewidth of the atomic transition. This ideal limit corresponds to cases where the frequency is right between two equally strong absorption lines. In this case, delay and transmittance are as high as possible. By tuning the frequency to peaks of transmittance, the filter can work close to ideal. Cesium cell at 387.6, though unable to reach the ideal limit at normal temperatures and path lengths, can still separate this frequency at the cost of less than ideal transmittance.

DOI
10.2514/6.2020-3242

Related papers

Slow Light Imaging Spectroscopy: Sensitivity of the Instrument Function to Optical Thickness and Gate Delay

Amirhossein Abbasszadehrad, Jason M. Meyers, Kevin A. Brown +4 more

In this study, we investigate the effect of the optical thickness of an atomic vapor cell and the delay of the gated camera on the instrument function in Slow Light Imaging Spectroscopy. In an atomic vapor medium, the slow light effect happens at certain frequency windows where there is a large slope of the index of refraction versus frequency. The extent of these windows depends on the important parameters of the vapor cell (i.e., the optical thickness). Furthermore, the gate delay determines the minimum amount of delay that a signal needs to experience before it is detected. These two parameters determine the shape of the instrument function of the spectrometer in slow light imaging spectroscopy. Simulation and experimental results show that when absorption lines of cesium vapor at 852.1 nm are used as a delay medium, the instrument function at low temperature consists of four peaks corresponding to the steep refractive index gradients at either side of the two hyperfine absorption lines. By increasing the temperature, these peaks move away from the absorption lines. When the two peaks between the two absorption lines move away from their corresponding absorption lines, they move closer to each other until they blend, forming a strong peak. By increasing the gate delay, the four peaks of the instrument function reappear.

AIAA SciTech Forum•Jan 4, 2024•1 citation

Slow Light Imaging Spectroscopy (SLIS): A Simulation of Broad Linewidth Signals

Amirhossein Abbasszadehrad, Anuj Rekhy, Arthur Dogariu +1 more

View Video Presentation: https://doi.org/10.2514/6.2022-1747.vid In this study, we calculate the delayed output of light passing an atomic vapor cell where the input signal is relatively broad. We use a deconvolution/convolution method to decompose a signal into Gaussian portions. We calculate the outputs of these portions and add them together to find the output signal. Results show that a mercury cell with a longer path length operating at a lower temperature performs better due to reduced pressure broadening.

AIAA SciTech Forum•Jan 3, 2022•3 citations

Controllable ultrabroadband slow light in a warm rubidium vapor

Rui Zhang, Joel A. Greenberg, Martin C. Fischer +1 more

We study ultrabroadband slow light in a warm rubidium vapor cell. By working between the D1 and D2 transitions, we find a several-nanometer window centered at 788.4 nm in which the group index is highly uniform and the absorption is small (<1%). We demonstrate that we can control the group delay by varying the temperature of the cell, and we observe a tunable fractional delay of 18 for pulses as short as 250 fs (6.9 nm bandwidth) with a fractional broadening of only 0.65 and a power leakage of 55%. We find that a simple theoretical model is in excellent agreement with the experimental results. Using this model, we discuss the impact of the pulse’s spectral characteristics on the distortion it incurs during propagation through the vapor.

arXiv:1107.5432•Journal of the Optical Society of America B•Oct 4, 2011•5 citations

Seeded Optical Parametric Oscillator as a Light Source for Slow-Light Imaging Spectroscopy

Boris S. Leonov, Robert T. Randolph, Anuj Rekhy +5 more

View Video Presentation: https://doi.org/10.2514/6.2023-1861.vid In this article, the authors discuss an experimental demonstration of a seeded Optical Parametric Oscillator as a light source for Slow Light Imaging Spectroscopy applications. An advantage of such a system, namely narrow linewidth of less than 400 MHz, allows for precise profiling of slow light features of atomic rubidium vapor, produced by high refractive index gradients in the vicinity of resonant transitions. Time delays of up to 48 ns were observed with a clear definition of the delayed pulse from the parasitic Mie and Rayleigh scattering. A comparison between the experimental data and numerical simulation over a frequency detuning range from -1 GHz to 4 GHz shows a good agreement for both qualitative and quantitative results.

AIAA SciTech Forum•Jan 19, 2023•1 citation

Experimental evaluation of an atomic line filter for spaceborne application

B. P. Oehry, Walter Schupita, Brigitte Sumetsberger +2 more

Atomic line filters take advantage of the sharp spectral features offered by electronic transitions in free atoms. The basic idea is to convert a narrow atomic absorption profile into an equally narrow optical transmission filter by detecting the fluorescence radiation that is emitted after a photon has been absorbed. We developed the metastable thallium atomic line filter from the basic physical idea to a laboratory prototype in order to evaluate the usefulness of the technology for satellite borne backscatter LIDARs. Atomic thallium provides for an unsurpassed combination of features for this purpose. The input wavelength of the filter is 535 nm, which matches the frequency doubled Nd:BEL laser. The filter is an active device, it upconverts the 535 nm input into 378 nm output that is detected by a PMT. The experimental results were as expected concerning the characteristics directly related to atomic properties, i.e. the filter has only 10 ns response time, its optical bandwidth is 0.004 nm, and its acceptance angle is only limited by the device geometry. An optimized setup, with the size constraints of a space-borne system, displays a total quantum efficiency of 2%, i.e. from input photons to detected output photons. This constitutes a remarkable and unsurpassed value for atomic line filters but is too low for the application in mind. (A ground-based, scaled-up version of our prototype would reach about 10% quantum efficiency). In addition, the 500 degree(s)C operating temperature of the vapor cell requires sophisticated thermal layout and thermal shielding, which means a lot of added mass and volume. In summary, we found that atomic line filters can be built that offer characteristics not found with other technologies, but their applicability for space borne systems is questionable.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Dec 9, 1994•0 citations

Slow Light Prism Enhanced Spectroscopy (SLOPES)

Amirhossein Abbasszadehrad, Jason M. Meyers, Arthur Dogariu +1 more

We report a new spectroscopy method achieved by combined refraction and propagation delay of light pulses passing through a prism cell filled with an atomic vapor. The atomic vapor medium has strong refractive index gradients near the atomic resonant absorption features, and these features enable the separation in time and space of different spectral features of a light pulse propagating through the vapor cell. The gradient of the refractive index leads to a slowing of the light pulse, while simultaneously, the magnitude of the refractive index and the prismatic windows or internal prism elements of the atomic vapor cell spatially separate the frequencies due to dispersion. The combination of these two effects leads to high-resolution spectral selectivity and strong suppression of out-of-band light. Applications are achieved for Rayleigh, Raman, and Thomson spectroscopy for which the wavelength of the scattering spectra tracks the wavelength of the excitation laser source. The laser is tuned such that the spectral line of interest falls near the resonance of the atomic vapor. Applications for rotational Raman spectroscopy are of particular interest since specific Raman lines can be selected while other Raman lines as well as background Rayleigh and other scattering are strongly rejected.

AIAA AVIATION Forum•Jul 27, 2024•0 citations
In this study, we model four atomic filters useful in slow light imaging spectroscopy. We calculate delay using the group velocity in rubidium at 780 nm, mercury at 254 nm, and cesium at 852 nm and 387.6 nm at different temperatures and path lengths of the atomic filter. We also calculate the output of a 5 ns pulse by propagating its Fourier components through cells. Then we compare the results of these two numerical models with the ideal limit set only by the natural linewidth of the atomic transition. This ideal limit corresponds to cases where the frequency is right between two equally strong absorption lines. In this case, delay and transmittance are as high as possible. By tuning the frequency to peaks of transmittance, the filter can work close to ideal. Cesium cell at 387.6, though unable to reach the ideal limit at normal temperatures and path lengths, can still separate this frequency at the cost of less than ideal transmittance.
In this study, we investigate the effect of the optical thickness of an atomic vapor cell and the delay of the gated camera on the instrument function in Slow Light Imaging Spectroscopy. In an atomic vapor medium, the slow light effect happens at certain frequency windows where there is a large slope of the index of refraction versus frequency. The extent of these windows depends on the important parameters of the vapor cell (i.e., the optical thickness). Furthermore, the gate delay determines the minimum amount of delay that a signal needs to experience before it is detected. These two parameters determine the shape of the instrument function of the spectrometer in slow light imaging spectroscopy. Simulation and experimental results show that when absorption lines of cesium vapor at 852.1 nm are used as a delay medium, the instrument function at low temperature consists of four peaks corresponding to the steep refractive index gradients at either side of the two hyperfine absorption lines. By increasing the temperature, these peaks move away from the absorption lines. When the two peaks between the two absorption lines move away from their corresponding absorption lines, they move closer to each other until they blend, forming a strong peak. By increasing the gate delay, the four peaks of the instrument function reappear.
View Video Presentation: https://doi.org/10.2514/6.2022-1747.vid In this study, we calculate the delayed output of light passing an atomic vapor cell where the input signal is relatively broad. We use a deconvolution/convolution method to decompose a signal into Gaussian portions. We calculate the outputs of these portions and add them together to find the output signal. Results show that a mercury cell with a longer path length operating at a lower temperature performs better due to reduced pressure broadening.
We study ultrabroadband slow light in a warm rubidium vapor cell. By working between the D1 and D2 transitions, we find a several-nanometer window centered at 788.4 nm in which the group index is highly uniform and the absorption is small (<1%). We demonstrate that we can control the group delay by varying the temperature of the cell, and we observe a tunable fractional delay of 18 for pulses as short as 250 fs (6.9 nm bandwidth) with a fractional broadening of only 0.65 and a power leakage of 55%. We find that a simple theoretical model is in excellent agreement with the experimental results. Using this model, we discuss the impact of the pulse’s spectral characteristics on the distortion it incurs during propagation through the vapor.
View Video Presentation: https://doi.org/10.2514/6.2023-1861.vid In this article, the authors discuss an experimental demonstration of a seeded Optical Parametric Oscillator as a light source for Slow Light Imaging Spectroscopy applications. An advantage of such a system, namely narrow linewidth of less than 400 MHz, allows for precise profiling of slow light features of atomic rubidium vapor, produced by high refractive index gradients in the vicinity of resonant transitions. Time delays of up to 48 ns were observed with a clear definition of the delayed pulse from the parasitic Mie and Rayleigh scattering. A comparison between the experimental data and numerical simulation over a frequency detuning range from -1 GHz to 4 GHz shows a good agreement for both qualitative and quantitative results.
Atomic line filters take advantage of the sharp spectral features offered by electronic transitions in free atoms. The basic idea is to convert a narrow atomic absorption profile into an equally narrow optical transmission filter by detecting the fluorescence radiation that is emitted after a photon has been absorbed. We developed the metastable thallium atomic line filter from the basic physical idea to a laboratory prototype in order to evaluate the usefulness of the technology for satellite borne backscatter LIDARs. Atomic thallium provides for an unsurpassed combination of features for this purpose. The input wavelength of the filter is 535 nm, which matches the frequency doubled Nd:BEL laser. The filter is an active device, it upconverts the 535 nm input into 378 nm output that is detected by a PMT. The experimental results were as expected concerning the characteristics directly related to atomic properties, i.e. the filter has only 10 ns response time, its optical bandwidth is 0.004 nm, and its acceptance angle is only limited by the device geometry. An optimized setup, with the size constraints of a space-borne system, displays a total quantum efficiency of 2%, i.e. from input photons to detected output photons. This constitutes a remarkable and unsurpassed value for atomic line filters but is too low for the application in mind. (A ground-based, scaled-up version of our prototype would reach about 10% quantum efficiency). In addition, the 500 degree(s)C operating temperature of the vapor cell requires sophisticated thermal layout and thermal shielding, which means a lot of added mass and volume. In summary, we found that atomic line filters can be built that offer characteristics not found with other technologies, but their applicability for space borne systems is questionable.
We report a new spectroscopy method achieved by combined refraction and propagation delay of light pulses passing through a prism cell filled with an atomic vapor. The atomic vapor medium has strong refractive index gradients near the atomic resonant absorption features, and these features enable the separation in time and space of different spectral features of a light pulse propagating through the vapor cell. The gradient of the refractive index leads to a slowing of the light pulse, while simultaneously, the magnitude of the refractive index and the prismatic windows or internal prism elements of the atomic vapor cell spatially separate the frequencies due to dispersion. The combination of these two effects leads to high-resolution spectral selectivity and strong suppression of out-of-band light. Applications are achieved for Rayleigh, Raman, and Thomson spectroscopy for which the wavelength of the scattering spectra tracks the wavelength of the excitation laser source. The laser is tuned such that the spectral line of interest falls near the resonance of the atomic vapor. Applications for rotational Raman spectroscopy are of particular interest since specific Raman lines can be selected while other Raman lines as well as background Rayleigh and other scattering are strongly rejected.