Abstract

Ultrahigh sensitivity is essential for magnetometers used to detect extremely weak human magnetic fields in applications such as magnetoencephalography (MEG) and magnetocardiography (MCG), where device miniaturization is required to achieve high spatial resolution. Among ultraweak magnetic-field sensing technologies, spin-exchange relaxation-free (SERF) magnetometers offer the highest sensitivity, but their performance is strongly influenced by spin relaxation within the alkali-metal vapor cell. In millimeter-scale cells, the optimal buffer-gas pressure for suppressing spin relaxation usually lies in a relatively high-pressure regime, which causes pressure broadening, weakens light–atom interaction, and ultimately limits magnetometer performance. In this article, a collaborative method combining an inner-wall antirelaxation coating with buffer gas is proposed to suppress spin relaxation and optimize SERF magnetometer performance. The coating reduces depolarization during atom–wall collisions, while the buffer gas suppresses atomic diffusion to the wall. Their combined effect not only suppresses spin relaxation but also lowers the optimal buffer-gas pressure. Experiments using 4-mm cubic nitrogen-buffer-gas vapor cells, including coated and uncoated cells, show that the coating shifts the optimal pressure from approximately 1100 Torr for the uncoated cell to 700 Torr for the coated cell. The averaged sensitivity over the 10–100-Hz band is$13.82~\mathrm {fT}/\mathrm {Hz}^{1/2}$for the 700 Torr coated cell, comparable to$14.61~\mathrm {fT}/\mathrm {Hz}^{1/2}$for the 1100 Torr uncoated cell. These results provide a practical route toward high-performance miniaturized SERF magnetometers at reduced operating pressure and with reduced pump-power demand.

DOI
10.1109/tim.2026.3723501

Related papers

Buffer Gas Pressure Optimization for Atomic Spin Relaxation Suppression in Ultra-High-Sensitivity SERF Magnetometers

Siran Li, X.-G. Lu, Yinghui Zhang +3 more

Optically pumped magnetometers (OPM) are core quantum payloads for geomagnetic remote sensing. Among them, the spin-exchange relaxation-free (SERF) OPM with aT-level ultimate sensitivity stands as mainstream. While enlarging the alkali-metal vapor cell of the SERF OPM enhances sensitivity, it triggers complex atomic spin relaxation, notably intensified magnetic field gradient relaxation. To address the dilemma of atomic spin relaxation regulation and the engineering requirements of ultra-high-sensitivity SERF magnetometers, this paper constructs an analytical model of the total relaxation rate that comprehensively considers wall-collision relaxation, spin-destruction collision relaxation, and longitudinal/transverse magnetic field gradient relaxation, etc. The analytical relationship between buffer gas pressure and total relaxation rate for a commonly used spherical vapor cell is derived, revealing the intrinsic correlation among cell size, atomic spin relaxation, and optimal pressure. Based on the theoretical model, the filling parameters of the vapor cell are optimized, and experimental measurements are carried out. The theoretical relaxation results are highly consistent with the experimental ones, realizing the precise optimization of buffer gas pressure. The optimization method proposed in this paper provides a theoretical basis and parameter guidance for the engineering preparation of alkali-metal vapor cells for high-sensitivity SERF magnetometers in remote sensing applications.

Photonics•Jun 1, 2026•0 citations

Influence of Buffer-Gas Pressure Inside Micro Alkali Vapor Cells on the Performance of Chip-Scale SERF Magnetometers

Lei Wu, Jintang Shang, Yu Ji +2 more

Spin-exchange-relaxation-free (SERF) optical atomic magnetometers are facing the challenges of dramatic decrease of sensitivity when the alkali vapor cells become small. This paper focuses on the sensitivity improvement of the chip-scale atomic magnetometer (CSAM) by changing buffer-gas pressures inside the microfabricated vapor cells. The rubidium vapor cells with three different buffer-gas pressures were tested in the SERF magnetometer. The influences of vapor cell temperature and pumping laser power on the device sensitivity are also characterized for further optimization. Results show that the three atomic vapor cells, with N2buffer-gas pressure of 0.20, 0.88, and 2.35 amg, have sensitivities of 900, 500, and 150 fT/Hz1/2, respectively. Results indicated that when the alkali vapor cells continue to be smaller, high buffer-gas pressure would provide an effective way to reduce the wall collision and improve the performance of CSAMs.

IEEE Transactions on Components Packaging and Manufacturing Technology•Dec 21, 2017•18 citations

Coating-Enabled Buffer-Gas-Free Vapor Cell: A New Sensor Architecture for Single-Beam SERF Magnetometers

Yi-an Xu, Yuhao Wang, Jiaxin Liu +5 more

Alkali metal vapor cells are fundamental components in quantum sensors, and their relaxation characteristics critically influence both sensitivity and system stability. In Spin-Exchange Relaxation-Free (SERF) atomic magnetometers, although spin-exchange collisions are effectively suppressed, wall-collision-induced relaxation remains the primary limiting factor. A coating-enabled buffer-gas-free vapor cell (CBF-VC) is proposed as a novel sensor unit, realized through multilayer octadecyltrichlorosilane (OTS) selfassembled molecular coatings. The CBF-VC is integrated into a single-beam SERF atomic magnetometer, and experimental validation is conducted. The results indicate a significant extension of the wall collision relaxation time from 40.19 ms. This extended relaxation time directly enhances the magnetometer sensitivity by 48%, confirming the feasibility of CBF-VCs as a measurement-oriented vapor cell architecture. The findings demonstrate not only the effectiveness of anti-relaxation coatings, but also a new instrumentation pathway for advancing next-generation quantum sensing.

IEEE Transactions on Instrumentation and Measurement•Jan 1, 2026•0 citations

Ultrasensitive SERF atomic magnetometer with a miniaturized hybrid vapor cell

Yintao Ma, Yao Chen, Mingzhi Yu +8 more

Abstract The chip-scale hybrid optical pumping spin-exchange relaxation-free (SERF) atomic magnetometer with a single-beam arrangement has prominent applications in biomagnetic measurements because of its outstanding features, including ultrahigh sensitivity, an enhanced signal-to-noise ratio, homogeneous spin polarization and a much simpler optical configuration than other devices. In this work, a miniaturized single-beam hybrid optical pumping SERF atomic magnetometer based on a microfabricated atomic vapor cell is demonstrated. Although the optically thin Cs atoms are spin-polarized, the dense Rb atoms determine the experimental results. The enhanced signal strength and narrowed resonance linewidth are experimentally proven, which shows the superiority of the proposed magnetometer scheme. By using a differential detection scheme, we effectively suppress optical noise with an approximate five-fold improvement. Moreover, the cell temperature markedly affects the performance of the magnetometer. We systematically investigate the effects of temperature on the magnetometer parameters. The theoretical basis for these effects is explained in detail. The developed miniaturized magnetometer has an optimal magnetic sensitivity of 20 fT/Hz 1/2 . The presented work provides a foundation for the chip-scale integration of ultrahighly sensitive quantum magnetometers that can be used for forward-looking magnetocardiography (MCG) and magnetoencephalography (MEG) applications.

Microsystems & Nanoengineering•Aug 30, 2024•74 citations

Effect of OTS coating on spin polarization and spatial homogeneity in alkali metal vapor cells for SERF magnetometers

Liu Bowen, Xiangyang Zhou, Yuhao Liu +2 more

Abstract The spin polarization and spatial homogeneity of alkali metal atoms are critical characteristics that determine the performance of spin-exchange relaxation-free magnetometers. In buffer gas cells, these characteristics are influenced by optical broadening and alkali atomic motional restriction imposed by the buffer gas. In coated cells, the protective effect of the coating on atomic spins, combined with rapid atomic diffusion, leads to significantly superior spin polarization characteristics compared to those in high-pressure buffer gas cells. However, due to neglecting the influence of atom-coating interactions on atomic motion, the existing spin polarization and spatial distribution models fail to accurately describe the polarization characteristics in coated cells. In this paper, based on the Langevin–Bloch equation, a dynamic model considering interaction effects between alkali metal atoms and the coating as well as between the pump light and the hyperfine levels of alkali metal atoms is proposed to accurately describe the spin polarization and spatial distribution in coated cells. To validate the proposed model, a series of experiments were carried out. The results demonstrate that coated cells can maintain high spin polarization and spatial homogeneity even under conditions of low pump intensity and high temperature, which are in good agreement with the simulations. This work provides a theoretical reference for the design of high-performance and low-power quantum sensors.

Journal of Physics D Applied Physics•Mar 26, 2026•2 citations

Optimizing longitudinal spin relaxation in miniaturized optically pumped magnetometers

A. McWilliam, S. Dyer, Dominic Hunter +7 more

The microfabrication of cesium vapor cells for optically pumped magnetometry relies on optimization of buffer gas pressure in order to maximize atomic coherence time and sensitivity to external magnetic signals. We demonstrate post-bond nitrogen buffer gas pressure tuning through localized heating of an integrated micropill dispenser. We characterize the variation in the intrinsic longitudinal relaxation rate, γ 10 , and magnetic sensitivity, as a function of the resulting nitrogen buffer gas pressure. Measurements are conducted by employing an optically pumped magnetometer operating in a free-induction-decay configuration. γ 10 is extracted across a range of nitrogen pressures between 60 and 700 torr, measuring a minimum of 140 Hz at 115 torr. In addition, we achieve sensitivities as low as 130 fT / Hz at a bias field amplitude of about 50 μ T . With the optimal nitrogen buffer gas pressure now quantified and achievable post-fabrication, these mass-producible cells can be tailored to suit a variety of sensing applications, ensuring peak magnetometer performance. Published by the American Physical Society 2024

arXiv:2406.19178•Physical Review Applied•Dec 5, 2024•9 citations
Ultrahigh sensitivity is essential for magnetometers used to detect extremely weak human magnetic fields in applications such as magnetoencephalography (MEG) and magnetocardiography (MCG), where device miniaturization is required to achieve high spatial resolution. Among ultraweak magnetic-field sensing technologies, spin-exchange relaxation-free (SERF) magnetometers offer the highest sensitivity, but their performance is strongly influenced by spin relaxation within the alkali-metal vapor cell. In millimeter-scale cells, the optimal buffer-gas pressure for suppressing spin relaxation usually lies in a relatively high-pressure regime, which causes pressure broadening, weakens light–atom interaction, and ultimately limits magnetometer performance. In this article, a collaborative method combining an inner-wall antirelaxation coating with buffer gas is proposed to suppress spin relaxation and optimize SERF magnetometer performance. The coating reduces depolarization during atom–wall collisions, while the buffer gas suppresses atomic diffusion to the wall. Their combined effect not only suppresses spin relaxation but also lowers the optimal buffer-gas pressure. Experiments using 4-mm cubic nitrogen-buffer-gas vapor cells, including coated and uncoated cells, show that the coating shifts the optimal pressure from approximately 1100 Torr for the uncoated cell to 700 Torr for the coated cell. The averaged sensitivity over the 10–100-Hz band is13.82 fT/Hz1/213.82~\mathrm {fT}/\mathrm {Hz}^{1/2}13.82 fT/Hz1/2for the 700 Torr coated cell, comparable to14.61 fT/Hz1/214.61~\mathrm {fT}/\mathrm {Hz}^{1/2}14.61 fT/Hz1/2for the 1100 Torr uncoated cell. These results provide a practical route toward high-performance miniaturized SERF magnetometers at reduced operating pressure and with reduced pump-power demand.
Optically pumped magnetometers (OPM) are core quantum payloads for geomagnetic remote sensing. Among them, the spin-exchange relaxation-free (SERF) OPM with aT-level ultimate sensitivity stands as mainstream. While enlarging the alkali-metal vapor cell of the SERF OPM enhances sensitivity, it triggers complex atomic spin relaxation, notably intensified magnetic field gradient relaxation. To address the dilemma of atomic spin relaxation regulation and the engineering requirements of ultra-high-sensitivity SERF magnetometers, this paper constructs an analytical model of the total relaxation rate that comprehensively considers wall-collision relaxation, spin-destruction collision relaxation, and longitudinal/transverse magnetic field gradient relaxation, etc. The analytical relationship between buffer gas pressure and total relaxation rate for a commonly used spherical vapor cell is derived, revealing the intrinsic correlation among cell size, atomic spin relaxation, and optimal pressure. Based on the theoretical model, the filling parameters of the vapor cell are optimized, and experimental measurements are carried out. The theoretical relaxation results are highly consistent with the experimental ones, realizing the precise optimization of buffer gas pressure. The optimization method proposed in this paper provides a theoretical basis and parameter guidance for the engineering preparation of alkali-metal vapor cells for high-sensitivity SERF magnetometers in remote sensing applications.
Spin-exchange-relaxation-free (SERF) optical atomic magnetometers are facing the challenges of dramatic decrease of sensitivity when the alkali vapor cells become small. This paper focuses on the sensitivity improvement of the chip-scale atomic magnetometer (CSAM) by changing buffer-gas pressures inside the microfabricated vapor cells. The rubidium vapor cells with three different buffer-gas pressures were tested in the SERF magnetometer. The influences of vapor cell temperature and pumping laser power on the device sensitivity are also characterized for further optimization. Results show that the three atomic vapor cells, with N2buffer-gas pressure of 0.20, 0.88, and 2.35 amg, have sensitivities of 900, 500, and 150 fT/Hz1/2, respectively. Results indicated that when the alkali vapor cells continue to be smaller, high buffer-gas pressure would provide an effective way to reduce the wall collision and improve the performance of CSAMs.
Alkali metal vapor cells are fundamental components in quantum sensors, and their relaxation characteristics critically influence both sensitivity and system stability. In Spin-Exchange Relaxation-Free (SERF) atomic magnetometers, although spin-exchange collisions are effectively suppressed, wall-collision-induced relaxation remains the primary limiting factor. A coating-enabled buffer-gas-free vapor cell (CBF-VC) is proposed as a novel sensor unit, realized through multilayer octadecyltrichlorosilane (OTS) selfassembled molecular coatings. The CBF-VC is integrated into a single-beam SERF atomic magnetometer, and experimental validation is conducted. The results indicate a significant extension of the wall collision relaxation time from 40.19 ms. This extended relaxation time directly enhances the magnetometer sensitivity by 48%, confirming the feasibility of CBF-VCs as a measurement-oriented vapor cell architecture. The findings demonstrate not only the effectiveness of anti-relaxation coatings, but also a new instrumentation pathway for advancing next-generation quantum sensing.
Abstract The chip-scale hybrid optical pumping spin-exchange relaxation-free (SERF) atomic magnetometer with a single-beam arrangement has prominent applications in biomagnetic measurements because of its outstanding features, including ultrahigh sensitivity, an enhanced signal-to-noise ratio, homogeneous spin polarization and a much simpler optical configuration than other devices. In this work, a miniaturized single-beam hybrid optical pumping SERF atomic magnetometer based on a microfabricated atomic vapor cell is demonstrated. Although the optically thin Cs atoms are spin-polarized, the dense Rb atoms determine the experimental results. The enhanced signal strength and narrowed resonance linewidth are experimentally proven, which shows the superiority of the proposed magnetometer scheme. By using a differential detection scheme, we effectively suppress optical noise with an approximate five-fold improvement. Moreover, the cell temperature markedly affects the performance of the magnetometer. We systematically investigate the effects of temperature on the magnetometer parameters. The theoretical basis for these effects is explained in detail. The developed miniaturized magnetometer has an optimal magnetic sensitivity of 20 fT/Hz 1/2 . The presented work provides a foundation for the chip-scale integration of ultrahighly sensitive quantum magnetometers that can be used for forward-looking magnetocardiography (MCG) and magnetoencephalography (MEG) applications.
Abstract The spin polarization and spatial homogeneity of alkali metal atoms are critical characteristics that determine the performance of spin-exchange relaxation-free magnetometers. In buffer gas cells, these characteristics are influenced by optical broadening and alkali atomic motional restriction imposed by the buffer gas. In coated cells, the protective effect of the coating on atomic spins, combined with rapid atomic diffusion, leads to significantly superior spin polarization characteristics compared to those in high-pressure buffer gas cells. However, due to neglecting the influence of atom-coating interactions on atomic motion, the existing spin polarization and spatial distribution models fail to accurately describe the polarization characteristics in coated cells. In this paper, based on the Langevin–Bloch equation, a dynamic model considering interaction effects between alkali metal atoms and the coating as well as between the pump light and the hyperfine levels of alkali metal atoms is proposed to accurately describe the spin polarization and spatial distribution in coated cells. To validate the proposed model, a series of experiments were carried out. The results demonstrate that coated cells can maintain high spin polarization and spatial homogeneity even under conditions of low pump intensity and high temperature, which are in good agreement with the simulations. This work provides a theoretical reference for the design of high-performance and low-power quantum sensors.
The microfabrication of cesium vapor cells for optically pumped magnetometry relies on optimization of buffer gas pressure in order to maximize atomic coherence time and sensitivity to external magnetic signals. We demonstrate post-bond nitrogen buffer gas pressure tuning through localized heating of an integrated micropill dispenser. We characterize the variation in the intrinsic longitudinal relaxation rate, γ 10 , and magnetic sensitivity, as a function of the resulting nitrogen buffer gas pressure. Measurements are conducted by employing an optically pumped magnetometer operating in a free-induction-decay configuration. γ 10 is extracted across a range of nitrogen pressures between 60 and 700 torr, measuring a minimum of 140 Hz at 115 torr. In addition, we achieve sensitivities as low as 130 fT / Hz at a bias field amplitude of about 50 μ T . With the optimal nitrogen buffer gas pressure now quantified and achievable post-fabrication, these mass-producible cells can be tailored to suit a variety of sensing applications, ensuring peak magnetometer performance. Published by the American Physical Society 2024