Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Log inSign up
Challenges
Datasets
Workspaces
Papers
Discussions
Leaderboard
Blog
Job Board
Q3AS

© 2026 Aqora Quantum S.A.S.

TermsPrivacyLegal Notice

IEEE Transactions on Instrumentation and Measurement

Single-Beam Optically Pumped Atomic Magnetometer With Suppression of Light Shift Based on Space-Division Multiplexing Light

Rongtong Zhu, Jianli Li, Shuo Sun, Huanyu Zhou, Yi Zhang, Pengcheng Du, Jin Li

Published Jan 1, 2026• 1 citation
Publisher

Abstract

Single-beam optically pumped atomic magnetometers (OPAMs) represent promising candidates for applications in geomagnetic field monitoring, biomedicine, and various other precision sensing domains owing to their exceptional sensitivity, high measurement accuracy, and cost-effectiveness. However, in orientation-based OPAMs utilizing circularly polarized light (CPL) pumping, the off-resonant light introduces light shift. The light shift acts as an additional magnetic field, reducing the measurement accuracy of OPAMs. Moreover, the dependence of the light shift on light intensity and frequency facilitates the conversion of light noise into magnetic noise, thereby degrading the sensitivity of OPAMs. In this study, we propose an optical pumping scheme based on space division multiplexing light (SDML) to suppress the light shift in OPAM. The approach employs two orthogonally polarized linearly polarized beams with a controlled phase difference to generate a spatially periodic distribution of left- and right-handed CPL. Optical pumping with the SDML induces oppositely directed light shifts in distinct spatial regions that effectively cancel each other. Finally, it is verified by experiments that the light shift is greatly suppressed based on SDML. Under detuned SDML pumping, both the light shift and light intensity noise are suppressed by over 50%, while maintaining a sensitivity comparable to that achieved with conventional CPL pumping. The proposed method demonstrates significant potential for magnetometers in light-noise-prone environments, and it can be generalized to other orientation-based magnetometers that rely on CPL pumping.

DOI
10.1109/tim.2026.3659643

Related papers

Light-shift suppression in an optically pumped magnetometer via dual-frequency optical pumping

Huanyu Zhou, J Q Li, Yi Zhang +5 more

The light shift represents a major source of measurement error and system noise in optically pumped atomic magnetometers (OPAMs). Tuning the laser frequency to a specific value is a common and effective approach to suppressing light shift in many optical-pumping systems. However, when this method is applied to OPAMs, a significant issue arises: at the zero-light-shift frequency, the magnetometer’s sensitivity decreases markedly. Consequently, it is not possible to optimize simultaneously both the measurement accuracy affected by the light shift and the overall sensitivity of the magnetometer. To address this contradiction, the underlying mechanism was analyzed using the theory of light absorption and the principles governing light–atom interactions, leading to the proposal of a dual-frequency optical-pumping scheme. Experimental results demonstrate that this approach can suppress the light shift by more than 95% while maintaining the optimal sensitivity of the OPAM. The proposed dual-frequency scheme can also be extended to other circularly polarized optical-pumping systems, thereby facilitating both the enhancement of OPAM performance and its practical deployment in real-world applications.

Advanced Photonics Nexus•Jun 19, 2026•0 citations

Dual-Axis Measurement and Suppression Method for Light Shift Based on a Single-Beam Compact SERF Atomic Magnetometer

Kexuan Miao, Tao Hu, Yan Chang +2 more

The light shift in a spin-exchange relaxation-free (SERF) atomic magnetometer (AM) generated by detuning of circularly polarized pump light, results in deviations of orthogonality measurement and attenuation of magnetometer response. In this paper, we propose and systematically investigate a dual-axis method to suppress light shift based on a self-developed single-beam compact SERF AM capable of measuring magnetic field components along two orthogonal sensitive axes. By sequentially observing the non-sensitive axis response and applying a magnetic field along the pump light direction, the virtual magnetic field induced by light shift is compensated and is rigorously demonstrated as a dominant component in this direction. Experimental results show a significant reduction in crosstalk between the dual sensitive axes, improving accuracy in resonant frequency determination of87Rb atoms and orthogonality evaluation of triaxial coils. Furthermore, the performance of dual-axis measurement in the SERF AM is tested, showing a more notable enhancement than that of single-axis measurement.

IEEE Sensors Journal•Jan 7, 2025•3 citations

Laser pumped 4He magnetometer with light shift suppression

Zaisheng Lin, He Wang, Xiang Peng +2 more

We report a laser-pumped 4He atomic magnetometer with light shift suppression through the atomic sensor itself. A linearly polarized light is used to optically align the 4He metastable atoms and we monitor the magneto-optical double resonance (MODR) signals produced by the left- and right-circularly orthogonal components. It is shown that light shift leads to the atomic alignment to orientation conversion effect, and thus, the difference between the two MODR signals. One of these two MODR signals is locked at the Larmor frequency and is used to measure the ambient magnetic field, while the differential signal is, simultaneously, fed back to suppress the light shift. The scheme could be of the advantage to the design of compact magnetometers by reducing the systematic errors due to light shift.

Review of Scientific Instruments•Nov 1, 2016•6 citations

Measurement and cancellation of light shift in optically pumped magnetometers

Rujie Li, Wei Quan, Wenfeng Fan +1 more

Light shift arising from the circularly polarized pump beam in atomic magnetometers can induce an undesired magnetic response, and thus affect their accuracy. Here, the light shift and the corresponding cross-talk effect in the magnetometer have been investigated, and a scheme with an additional off-resonant pump beam with an appropriate polarization and intensity has been proposed to effectively reject the light shift. And the experimental results reproduce the expected behavior. This scheme provides a powerful tool to reject the light shift in optical pumping systems, in particular, the hybrid optically pumped systems.

Europhysics Letters (EPL)•Mar 1, 2018•13 citations

Space division multiplexing metasurface for single-beam atomic magnetometers

Xuejing Liu, Meiling Tang, Yanhui Hu +3 more

A significant drive has emerged to miniaturize and integrate atomic magnetometers for various ultra-high precision quantum sensing applications, extending from laboratory research to industrial use. We introduce a novel configuration of a single-beam atomic magnetometer based on space division multiplexing (SDM), offering a promising approach to this field of research. This design utilizes a custom-designed metasurface to manipulate two generated modes, enabling the implementation of a pump-probe configuration within a single-beam atomic magnetometer. This approach achieves higher spatial resolution and reduces the system's volume. The metasurface centrally generates circularly polarized light carrying spin angular momentum, while linearly polarized light is emitted from the outer areas. This design not only enhances the miniaturization of chip-scale atomic magnetometers but also opens new avenues for exploring interactions between hot atomic spin ensembles and vector beams.

Optics Express•Nov 18, 2024•3 citations

Detuned square-wave optical modulation zero-field atomic magnetometer

Xiaoyu Li, Bangcheng Han, Jixi Lu +2 more

In recent years, arrayed atomic magnetometers have been continuously applied in biomagnetic scenarios. However, the crosstalk between adjacent sensors introduced by the modulated magnetic fields significantly compromises the imaging accuracy of arrayed atomic magnetometers. To solve this problem, a detuned square-wave optical modulation zero-field atomic magnetometer configuration is developed and we provide an analytical solution of this method. In this scheme, the circularly polarized pump light is used to polarize alkali metal atoms longitudinally, and a beam of detuned square-wave modulated circularly polarized light is applied transversely to generate a light-shift modulation instead of the coil-generated modulated magnetic field. Then the transverse magnetic field is measured by the optical absorption method of longitudinally circularly polarized pump light. This scheme maintains the inherent advantage of high sensitivity in spinexchange relaxation-free (SERF) atomic magnetometers. Additionally, the all-optical configuration eliminates the need for additional modulated magnetic fields or radio frequency (RF) fields, thereby mitigating crosstalk issues associated with modulated magnetic fields in array applications. Consequently, this scheme exhibits great potential for arraying and is expected to be employed in magnetocardiography (MCG) and magnetoencephalography (MEG).

Optics Young Scientist Summit (OYSS)•Dec 29, 2023•0 citations
Single-beam optically pumped atomic magnetometers (OPAMs) represent promising candidates for applications in geomagnetic field monitoring, biomedicine, and various other precision sensing domains owing to their exceptional sensitivity, high measurement accuracy, and cost-effectiveness. However, in orientation-based OPAMs utilizing circularly polarized light (CPL) pumping, the off-resonant light introduces light shift. The light shift acts as an additional magnetic field, reducing the measurement accuracy of OPAMs. Moreover, the dependence of the light shift on light intensity and frequency facilitates the conversion of light noise into magnetic noise, thereby degrading the sensitivity of OPAMs. In this study, we propose an optical pumping scheme based on space division multiplexing light (SDML) to suppress the light shift in OPAM. The approach employs two orthogonally polarized linearly polarized beams with a controlled phase difference to generate a spatially periodic distribution of left- and right-handed CPL. Optical pumping with the SDML induces oppositely directed light shifts in distinct spatial regions that effectively cancel each other. Finally, it is verified by experiments that the light shift is greatly suppressed based on SDML. Under detuned SDML pumping, both the light shift and light intensity noise are suppressed by over 50%, while maintaining a sensitivity comparable to that achieved with conventional CPL pumping. The proposed method demonstrates significant potential for magnetometers in light-noise-prone environments, and it can be generalized to other orientation-based magnetometers that rely on CPL pumping.
The light shift represents a major source of measurement error and system noise in optically pumped atomic magnetometers (OPAMs). Tuning the laser frequency to a specific value is a common and effective approach to suppressing light shift in many optical-pumping systems. However, when this method is applied to OPAMs, a significant issue arises: at the zero-light-shift frequency, the magnetometer’s sensitivity decreases markedly. Consequently, it is not possible to optimize simultaneously both the measurement accuracy affected by the light shift and the overall sensitivity of the magnetometer. To address this contradiction, the underlying mechanism was analyzed using the theory of light absorption and the principles governing light–atom interactions, leading to the proposal of a dual-frequency optical-pumping scheme. Experimental results demonstrate that this approach can suppress the light shift by more than 95% while maintaining the optimal sensitivity of the OPAM. The proposed dual-frequency scheme can also be extended to other circularly polarized optical-pumping systems, thereby facilitating both the enhancement of OPAM performance and its practical deployment in real-world applications.
The light shift in a spin-exchange relaxation-free (SERF) atomic magnetometer (AM) generated by detuning of circularly polarized pump light, results in deviations of orthogonality measurement and attenuation of magnetometer response. In this paper, we propose and systematically investigate a dual-axis method to suppress light shift based on a self-developed single-beam compact SERF AM capable of measuring magnetic field components along two orthogonal sensitive axes. By sequentially observing the non-sensitive axis response and applying a magnetic field along the pump light direction, the virtual magnetic field induced by light shift is compensated and is rigorously demonstrated as a dominant component in this direction. Experimental results show a significant reduction in crosstalk between the dual sensitive axes, improving accuracy in resonant frequency determination of87Rb atoms and orthogonality evaluation of triaxial coils. Furthermore, the performance of dual-axis measurement in the SERF AM is tested, showing a more notable enhancement than that of single-axis measurement.
We report a laser-pumped 4He atomic magnetometer with light shift suppression through the atomic sensor itself. A linearly polarized light is used to optically align the 4He metastable atoms and we monitor the magneto-optical double resonance (MODR) signals produced by the left- and right-circularly orthogonal components. It is shown that light shift leads to the atomic alignment to orientation conversion effect, and thus, the difference between the two MODR signals. One of these two MODR signals is locked at the Larmor frequency and is used to measure the ambient magnetic field, while the differential signal is, simultaneously, fed back to suppress the light shift. The scheme could be of the advantage to the design of compact magnetometers by reducing the systematic errors due to light shift.
Light shift arising from the circularly polarized pump beam in atomic magnetometers can induce an undesired magnetic response, and thus affect their accuracy. Here, the light shift and the corresponding cross-talk effect in the magnetometer have been investigated, and a scheme with an additional off-resonant pump beam with an appropriate polarization and intensity has been proposed to effectively reject the light shift. And the experimental results reproduce the expected behavior. This scheme provides a powerful tool to reject the light shift in optical pumping systems, in particular, the hybrid optically pumped systems.
A significant drive has emerged to miniaturize and integrate atomic magnetometers for various ultra-high precision quantum sensing applications, extending from laboratory research to industrial use. We introduce a novel configuration of a single-beam atomic magnetometer based on space division multiplexing (SDM), offering a promising approach to this field of research. This design utilizes a custom-designed metasurface to manipulate two generated modes, enabling the implementation of a pump-probe configuration within a single-beam atomic magnetometer. This approach achieves higher spatial resolution and reduces the system's volume. The metasurface centrally generates circularly polarized light carrying spin angular momentum, while linearly polarized light is emitted from the outer areas. This design not only enhances the miniaturization of chip-scale atomic magnetometers but also opens new avenues for exploring interactions between hot atomic spin ensembles and vector beams.
In recent years, arrayed atomic magnetometers have been continuously applied in biomagnetic scenarios. However, the crosstalk between adjacent sensors introduced by the modulated magnetic fields significantly compromises the imaging accuracy of arrayed atomic magnetometers. To solve this problem, a detuned square-wave optical modulation zero-field atomic magnetometer configuration is developed and we provide an analytical solution of this method. In this scheme, the circularly polarized pump light is used to polarize alkali metal atoms longitudinally, and a beam of detuned square-wave modulated circularly polarized light is applied transversely to generate a light-shift modulation instead of the coil-generated modulated magnetic field. Then the transverse magnetic field is measured by the optical absorption method of longitudinally circularly polarized pump light. This scheme maintains the inherent advantage of high sensitivity in spinexchange relaxation-free (SERF) atomic magnetometers. Additionally, the all-optical configuration eliminates the need for additional modulated magnetic fields or radio frequency (RF) fields, thereby mitigating crosstalk issues associated with modulated magnetic fields in array applications. Consequently, this scheme exhibits great potential for arraying and is expected to be employed in magnetocardiography (MCG) and magnetoencephalography (MEG).