IEICE Technical Report; IEICE Tech. Rep.

Study on single-flux-quantum readout circuits for time-of-flight mass spectrometry systems using superconducting stripline ion detectors

信行 吉川, 京佑 佐野, 裕希 山梨, 伸幸 全, 雅隆 大久保

Published Sep 25, 2013• 0 citations
OpenAlex

Related papers

Numerical Analysis of Serially-Connected Superconducting Stripline Detectors with Single-Flux-Quantum Readout Circuits

Fujimaki Akira, Bozbey Ali, Kyohei Kamiya +2 more

IEICE Technical Report; IEICE Tech. Rep.•Oct 1, 2015•0 citations

Single-flux-quantum readout and digital signal processing of serially-connected superconducting stripline detection array

Ryosuke Naito, Kyohei Kamiya, Misaki Kozaka +2 more

IEICE Technical Report; IEICE Tech. Rep.•Aug 1, 2016•0 citations

Single-Flux Quantum Readout Circuits for Photon-Number-Resolving Superconducting Nanowire Single-Photon Detectors

Hiroaki Myoren, Shunsuke Takeda, Masato Naruse +3 more

Superconducting nanowire single-photon detectors (SNSPDs) with series-parallel meander-type configurations are demonstrated to have photon-number-resolving capabilities and are expected to be used for an evaluation system of single-photon generators and/or an energy-resolved fast detector for a new-generation time-of-flight mass spectrometry. In this paper, we present the design, fabrication, and confirmation of correct operations of two kinds of single-flux-quantum (SFQ) digital readout circuits for an SNSPD with a series-parallel meander-type configuration. Assuming a six-parallel meander-type configuration, we designed: a six-level current discriminator using six different magnetically coupled DC/SFQ converter circuits; and a Δ-type A/D converter using a quasi-one-junction SQUID (QOS) comparator with an SFQ feedback loop. For the design of the magnetically coupled inputs, we used a 3-D inductance extraction program FastHenry and the designed circuits were fabricated using a HYPRES 4.5-kA/cm2Nb process and a National Institute of Advanced Industrial Science and Technology (AIST) standard 2.5-kA/cm2Nb process.

IEEE Transactions on Applied Superconductivity•Oct 28, 2014•10 citations

Design and Demonstration of a Single-Flux-Quantum Multi-Stop Time-to-Digital Converter for Time-of-Flight Mass Spectrometry

Kyosuke Sano, Yuki Yamanashi, Nobuyuki Yoshikawa

We have been developing a superconducting time-of-flight mass spectrometry (TOF-MS) system, which utilizes a superconductive strip ion detector (SSID) and a single-flux-quantum (SFQ) multi-stop time-to-digital converter (TDC). The SFQ multi-stop TDC can measure the time intervals between multiple input signals and directly convert them into binary data. In this study, we designed and implemented 24-bit SFQ multi-stop TDCs with a 3×24-bit FIFO buffer using the AIST Nb standard process (STP2), whose time resolution and dynamic range are 100ps and 1.6ms, respectively. The timing jitter of the TDC was investigated by comparing two types of TDCs: one uses an on-chip SFQ clock generator (CG) and the other uses a microwave oscillator at room temperature. We confirmed the correct operation of both TDCs and evaluated their timing jitter. The experimentally-obtained timing jitter is about 40ns and 700ps for the TDCs with and without the on-chip SFQ CG, respectively, for the measured time interval of 50µs, which linearly increases with increase of the measured time interval.

IEICE Transactions on Electronics•Jan 1, 2014•10 citations

Multiplexing Techniques for Read-Out Superconductive Sensor System Based on Single Flux Quantum Circuits

一希 青木, 裕希 山梨, 信行 吉川

IEICE Technical Report; IEICE Tech. Rep.•Jul 12, 2012•0 citations

New Fields Opened Up by Superconducting Detector Technology

Nobuyuki Yoshikawa

Recent developments and their applications of superconducting detectors that directly detect the incidence of photons and particles will be reviewed. Transition-edge sensors, superconducting single-photon detectors and superconducting stripline sensors are included in this category. Their recent progresses in new application fields, such as quantum information communication and time-of-flight mass spectrometry, will be shown.

IEEJ Transactions on Fundamentals and Materials•Jan 1, 2011•0 citations
Superconducting nanowire single-photon detectors (SNSPDs) with series-parallel meander-type configurations are demonstrated to have photon-number-resolving capabilities and are expected to be used for an evaluation system of single-photon generators and/or an energy-resolved fast detector for a new-generation time-of-flight mass spectrometry. In this paper, we present the design, fabrication, and confirmation of correct operations of two kinds of single-flux-quantum (SFQ) digital readout circuits for an SNSPD with a series-parallel meander-type configuration. Assuming a six-parallel meander-type configuration, we designed: a six-level current discriminator using six different magnetically coupled DC/SFQ converter circuits; and a Δ-type A/D converter using a quasi-one-junction SQUID (QOS) comparator with an SFQ feedback loop. For the design of the magnetically coupled inputs, we used a 3-D inductance extraction program FastHenry and the designed circuits were fabricated using a HYPRES 4.5-kA/cm2Nb process and a National Institute of Advanced Industrial Science and Technology (AIST) standard 2.5-kA/cm2Nb process.
We have been developing a superconducting time-of-flight mass spectrometry (TOF-MS) system, which utilizes a superconductive strip ion detector (SSID) and a single-flux-quantum (SFQ) multi-stop time-to-digital converter (TDC). The SFQ multi-stop TDC can measure the time intervals between multiple input signals and directly convert them into binary data. In this study, we designed and implemented 24-bit SFQ multi-stop TDCs with a 3×24-bit FIFO buffer using the AIST Nb standard process (STP2), whose time resolution and dynamic range are 100ps and 1.6ms, respectively. The timing jitter of the TDC was investigated by comparing two types of TDCs: one uses an on-chip SFQ clock generator (CG) and the other uses a microwave oscillator at room temperature. We confirmed the correct operation of both TDCs and evaluated their timing jitter. The experimentally-obtained timing jitter is about 40ns and 700ps for the TDCs with and without the on-chip SFQ CG, respectively, for the measured time interval of 50µs, which linearly increases with increase of the measured time interval.