On-Wafer Penetration Depth Measurements of Superconducting Films
Michael E. Cyberey, Robert M. Weikle, Arthur W. Lichtenberger
Abstract
An on-wafer measurement technique allowing measurement of a superconducting film's penetration depth without post device-fabrication separation of the chips has been investigated. Coplanar waveguide (CPW) probes and a CPW to microstrip transition are utilized, allowing the use of commercially available LakeShore® CPX-1.5K cryogenic CPW probing station capable of microwave measurements to 50 GHz. The use of CPW probes allows for rapid prototyping and is nondestructive to the device wafers, and hence compatible with many device fabrication technologies. Measurement techniques utilizing resonant microwave circuits have been reported (1),(2), but require mounting to custom-made test fixtures and are not performed on-wafer, but instead on individual diced devices. However, such a dicing requirement is incompatible with our ultra-thin Si, Au beam-lead chip architecture (3). In addition to a traditional resonant circuit measurement technique, a technique that uses on-wafer calibration standards is discussed. The on-wafer calibration eliminates possible effects introduced by the testing setup and potentially offers additional advantages such as a size reduction and a direct measurement of RF loss. The designs and results for both measurement techniques are reported for NbTiN films.