IEEE Sensors Journal

Effects of Multiple Factors on the Performance of Portable Self-Oscillating Laser Pumped Atomic Magnetometer

Kai Jin, Xu‐xing Geng, Jianfeng Xiao, Zhi Liang, Wangwang Tang, Guangming Huang, Gao‐xiang Li, Shaoping Wu, Guoqing Yang, Shangqing Liang

Published Feb 19, 2024• 1 citation
Publisher

Abstract

Self-oscillating magnetometers with high sensitivity are widely used in fields such as biomedicine and remote sensing. However, few studies comprehensively consider the effect of different parameters on their performance. We examined the effects of laser intensity, resonance detuning, radio frequency (RF) magnetic field, and atomic cell on the performance of portable self-oscillating magnetometers. To obtain the atomic absorption spectrum, we solved the system’s master equation in the multipole moment form. In theory, we analyse the influencing factors on the atomic magnetic resonance (AMR) signal. In the experiment, we artificially varied the relevant parameters to control the trends of changes in the atomic absorption spectra and evaluated their performance. The results are beneficial for optimizing the parameters of portable self-oscillating magnetometers with a laser pump-probe structure.

DOI
10.1109/jsen.2024.3365701

Related papers

Multi-Parameter Optimization of Rubidium Laser Optically Pumped Magnetometers with Geomagnetic Field Intensity

Kun Xu, Xiuyan Ren, Yujie Xiang +7 more

Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.

Sensors•Nov 2, 2023•7 citations

A laser pump-re-pump atomic magnetometer

Ailin Yang, Guoqing Yang, Xunming Cai +2 more

We demonstrate experimentally an atomic magnetometer based on optical pumping theory, a magnetic resonance that is induced by a radio frequency field and dependent on the magnetic field strength. Compared with the conventional method using one radiation field, which is used not only as the probe beam but also as a pump beam, the additional re-pump beam can increase remarkably the amplitude of the signal. It is shown that the amplitude of the magnetic field resonance signal can increase more than 55% by using an additional re-pump beam, which makes the sensitivity of the magnetometer higher. Finally, we investigate the relation between amplitude of the signal and re-pump laser power, and calculate the atomic population in the trapping states with rate equations.

Chinese Physics B•Dec 1, 2013•9 citations

Effect of RF field intensity on resonance spectrum of rubidium optically pumped magnetometer

Kun Xu, Xiu-Yan Ren

As a weak magnetic sensor, optical pump magnetometer is widely used in resource exploration, space exploration, and other fields. According to the principle of measuring weak magnetic anomaly signal under geomagnetic field, a rubidium optical pump magnetometer experimental system is built based on the methods of laser pumping and RF(radio frequency) field depolarization, The influence of RF field intensity on the resonance spectrum of rubidium optical pump magnetometer is studied. The results show that the intensity and linewidth of the resonance spectrum increase with the increase of RF field intensity, but the ratio of linewidth to intensity has a minimum. This paper provides a technical basis for the sensitivity optimization of rubidium optically pumped magnetometer.

International Conference on Sensors and Information Technology (ICSI)•May 13, 2022•0 citations

The suppression of the probe laser pumping effect in SERF atomic magnetometer

Qian Cao, Yueyang Zhai, Binyue Huang +2 more

Introduction: A spin exchange relaxation free atomic magnetometer, as an ultra-highly sensitive magnetic field sensor, is limited by the performance of the probe laser system. The probe laser pumping effect (PLPE) hinders the increase in the performance of probe laser system. Methods: This study investigated the PLPE and proposed a method for suppressing the same. Through changes to the angle of a quarter wave plate and the addition of a triangular modulated magnetic field to the alkali atoms, the suppression point was determined. Results and discussion: Further, related parameters were measured for different degree of polarizations of the probe laser, which confirmed that the influence of PLPE on the magnetic field was the least at the suppressed point.

Frontiers in Physics•Jun 1, 2023•1 citation

Subpicotesla scalar atomic magnetometer with a microfabricated cell

Rui Zhang, Terry Dyer, Nathan Brockie +2 more

We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/Hz, which shows that the magnetometer by itself can achieve sub-100fT/Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields.

arXiv:1910.01768•Journal of Applied Physics•Sep 26, 2019•17 citations

Signal correlation in the tandem of a spin oscillator and microwave frequency discriminator with laser-pumped alkali atoms

A. A. Baranov, Ермак Сергей Викторович, E. A. Sagitov +2 more

We have studied the influence of low-frequency noise on the stability of resonance frequency of a self-oscillating magnetometer on 87Rb vapor with simultaneous monitoring of the signal of radio-optical resonance on the magnetic-field-dependent microwave transition under laser pumping at the D 2 line of the head doublet. The difference of synchronous records of detected signals reduced to the same scale in magnetic field units was processed to determine the Allan variance as a function of the averaging time. The correlation coefficient characterizing the coupling of detected signals determined by the pumping rate and intensity of radio fields generated in the region of the absorption chamber. The self-oscillating magnetometer can only operate provided that there is laser tuning to the long-wavelength component of the electric-dipole transition.

Technical Physics Letters•Feb 1, 2016•9 citations
Self-oscillating magnetometers with high sensitivity are widely used in fields such as biomedicine and remote sensing. However, few studies comprehensively consider the effect of different parameters on their performance. We examined the effects of laser intensity, resonance detuning, radio frequency (RF) magnetic field, and atomic cell on the performance of portable self-oscillating magnetometers. To obtain the atomic absorption spectrum, we solved the system’s master equation in the multipole moment form. In theory, we analyse the influencing factors on the atomic magnetic resonance (AMR) signal. In the experiment, we artificially varied the relevant parameters to control the trends of changes in the atomic absorption spectra and evaluated their performance. The results are beneficial for optimizing the parameters of portable self-oscillating magnetometers with a laser pump-probe structure.
Rubidium laser optically pumped magnetometers (OPMs) are widely used magnetic sensors based on the Zeeman effect, laser pumping, and magnetic resonance principles. They measure the magnetic field by measuring the magnetic resonance signal passing through a rubidium atomic gas cell. The quality of the magnetic resonance signal is a necessary condition for a magnetometer to achieve high sensitivity. In this research, to obtain the best magnetic resonance signal of rubidium laser OPMs in the Earth’s magnetic field intensity, the experiment system of rubidium laser OPMs is built with a rubidium atomic gas cell as the core component. The linewidth and amplitude ratio (LAR) of magnetic resonance signals is utilized as the optimization objective function. The magnetic resonance signals of the magnetometer experiment system are experimentally measured for different laser frequencies, radio frequency (RF) intensities, laser powers, and atomic gas cell temperatures in a background magnetic field of 50,765 nT. The experimental results indicate that optimizing these parameters can reduce the LAR by one order of magnitude. This shows that the optimal parameter combination can effectively improve the sensitivity of the magnetometer. The sensitivity defined using the noise spectral density measured under optimal experimental parameters is 1.5 pT/Hz1/2@1 Hz. This work will provide key technical support for rubidium laser OPMs’ product development.
We demonstrate experimentally an atomic magnetometer based on optical pumping theory, a magnetic resonance that is induced by a radio frequency field and dependent on the magnetic field strength. Compared with the conventional method using one radiation field, which is used not only as the probe beam but also as a pump beam, the additional re-pump beam can increase remarkably the amplitude of the signal. It is shown that the amplitude of the magnetic field resonance signal can increase more than 55% by using an additional re-pump beam, which makes the sensitivity of the magnetometer higher. Finally, we investigate the relation between amplitude of the signal and re-pump laser power, and calculate the atomic population in the trapping states with rate equations.
As a weak magnetic sensor, optical pump magnetometer is widely used in resource exploration, space exploration, and other fields. According to the principle of measuring weak magnetic anomaly signal under geomagnetic field, a rubidium optical pump magnetometer experimental system is built based on the methods of laser pumping and RF(radio frequency) field depolarization, The influence of RF field intensity on the resonance spectrum of rubidium optical pump magnetometer is studied. The results show that the intensity and linewidth of the resonance spectrum increase with the increase of RF field intensity, but the ratio of linewidth to intensity has a minimum. This paper provides a technical basis for the sensitivity optimization of rubidium optically pumped magnetometer.
Introduction: A spin exchange relaxation free atomic magnetometer, as an ultra-highly sensitive magnetic field sensor, is limited by the performance of the probe laser system. The probe laser pumping effect (PLPE) hinders the increase in the performance of probe laser system. Methods: This study investigated the PLPE and proposed a method for suppressing the same. Through changes to the angle of a quarter wave plate and the addition of a triangular modulated magnetic field to the alkali atoms, the suppression point was determined. Results and discussion: Further, related parameters were measured for different degree of polarizations of the probe laser, which confirmed that the influence of PLPE on the magnetic field was the least at the suppressed point.
We demonstrated a scalar atomic magnetometer using a microfabricated Cs vapor cell. The atomic spin precession is driven by an amplitude-modulated circularly-polarized pump laser resonant on the D1 transition of Cs atoms and detected by an off-resonant linearly-polarized probe laser using a balanced polarimeter setup. Under a magnetic field with amplitude in the Earth's magnetic field range, the magnetometer in the gradiometer mode can reach sensitivities below 150fT/Hz, which shows that the magnetometer by itself can achieve sub-100fT/Hz sensitivities. In addition to its high sensitivity, the magnetometer has a bandwidth close to 1 kHz due to the broad magnetic resonance inside the small vapor cell. Our experiment suggests the feasibility of a portable low-power and high-performance magnetometer which can be operated in the Earth's magnetic field. Such a device will greatly reduce the restrictions on the operating environment and expand the range of applications for atomic magnetometers, such as detection of nuclear magnetic resonance in low magnetic fields.
We have studied the influence of low-frequency noise on the stability of resonance frequency of a self-oscillating magnetometer on 87Rb vapor with simultaneous monitoring of the signal of radio-optical resonance on the magnetic-field-dependent microwave transition under laser pumping at the D 2 line of the head doublet. The difference of synchronous records of detected signals reduced to the same scale in magnetic field units was processed to determine the Allan variance as a function of the averaging time. The correlation coefficient characterizing the coupling of detected signals determined by the pumping rate and intensity of radio fields generated in the region of the absorption chamber. The self-oscillating magnetometer can only operate provided that there is laser tuning to the long-wavelength component of the electric-dipole transition.