Atomic resonances had long been used as a frequency reference. We present resonant tuning of coherent population trapping in 87Rb with an intracavity Fabry-Perot etalon in a mode-locked laser.
We investigate the use of phase modulation to generate an error signal for frequency stabilization to a reference atomic transition. The study focuses on the high-frequency modulation regime. Our theoretical analysis shows that, in the case of coherent population trapping, the maximal slope of the error signal remains constant as the modulation frequency increases, whereas it steadily decreases for other types of resonances. Experimental results with 87Rb atoms confirm this behavior under comparison with the double radio-optical resonance.
We present a microwave frequency standard based on coherent population trapping (CPT) in the 87Rb D1 line. The CPT spectrum is obtained using two Raman lasers with a 6.8 GHz frequency offset by injection locking of a master laser to a slave laser. We have constructed an atomic clock employing a 5 cm long Rb vapor cell confined with 6.67 kPa neon buffer gas at 70 °C. Using this system, we improve the CPT contrast through the elimination of undesired off-resonant fields created by the direct modulation method. We measured the frequency shift of the CPT signal as a function of the temperature of the Rb cell and estimated it to be approximately 1.3 ×10−9/K. The frequency of a 10 MHz crystal oscillator has been stabilized to the CPT spectrum between the two ground states in 87Rb. The relative frequency stability is approximately 2.3×10−12 for an average time of 68 s.