Symposium on Frequency Standards and Metrology
AN OPTICAL CLOCK BASED ON A SINGLE TRAPPED 88 Sr + ION
Helen S. Margolis, G. P. Barwood, G. HUANG, Hugh A. Klein, S.N. Lea, Giuseppe Marra, Veronika Tsatourian, Barney Walton, P. GILL
Abstract
The 5 s 2 S 1/2 -4 d 2 D 5/2 electric quadrupole transition in a single trapped and laser-cooled 88 Sr + ion provides the reference for a highly stable and accurate optical clock. This paper describes recent progress on the 88 Sr + optical clocks under development at NPL. An ultrastable laser at 674 nm is used to excite the clock transition and a fractional frequency resolution of 2 × 10 -14 has been achieved, corresponding to a linewidth of 9 Hz. A relative frequency instability of 2 × 10 -15 at 5000 s has been observed in the comparison between two similar systems and the fractional uncertainty in the absolute frequency of the clock transition is 3.8 × 10 -15 . Improvements to the experimental arrangement are in progress and are expected to lead to a 88 Sr + optical clock with stability and reproducibility exceeding that of the primary caesium standard.