Thermal-response time of superconducting transition-edge microcalorimeters

K. D. Irwin, Gene C Hilton, David A. Wollman, John M. Martinis

Published Apr 15, 1998• 176 citations
Publisher

Abstract

We investigate limits on the thermal-response time of superconducting transition-edge microcalorimeters. For operation at 0.1 K, we show that the lower limit on the response time of a superconducting transition-edge microcalorimeter is of order 1 μs due to the heat diffusion time, electrical instabilities, the amplifier noise, and the critical current of the superconducting film. The response time is not limited by self-heating effects and is independent of the intended photon energy. However, design constraints associated with the inductance of the bias circuit make it difficult to achieve the fastest response times for devices with heat capacities high enough for x-ray and gamma-ray detection.

DOI
10.1063/1.367153

Related papers

X-ray detection using a superconducting transition-edge sensor microcalorimeter with electrothermal feedback

K. D. Irwin, Gene C Hilton, David A. Wollman +1 more

We have developed a new type of x-ray detector based on a superconducting transition-edge thermometer operated near 100 mK. A superconducting quantum interference device is used to measure the current through the thermometer, and negative electrothermal feedback is used to improve the energy resolution and shorten the thermal time constant. We have used a detector mounted on a scanning electron microscope to measure the energy of titanium Kα (4.5 keV) fluorescence x rays with a resolution better than 14 eV full width at half-maximum. Using two other devices, we have measured an energy resolution for Joule heat pulses of 2.6 eV at 1 keV and 0.2 eV at 4 eV, the best reported for any calorimeter. An electrical noise equivalent power of 3×10−18 W/√Hz was also measured, suggesting the use of these detectors as infrared bolometers.

Applied Physics Letters•Sep 23, 1996•230 citations

A superconducting bolometer with strong electrothermal feedback

Adrian Tae-Jin Lee, Paul L. Richards, Sae Woo Nam +2 more

We present a theoretical analysis and experimental evaluation of a transition-edge superconducting bolometer for detecting infrared and millimeter waves. The superconducting film is voltage biased and the current is read by a superconducting quantum interference device ammeter. Strong electrothermal feedback maintains the sensor temperature within the transition, gives a current responsivity that is simply the inverse of the bias voltage, and reduces the response time by several orders of magnitude below the intrinsic time constant C/G. We evaluated a voltage-biased bolometer that operates on the Tc∼95 mK transition of a tungsten film with a thermal conductance of G∼1.2×10−9 W/K. As expected, the electrical noise equivalent power of 3.3×10−17/W√Hz is close to the thermal fluctuation noise limit and is lower than that of other technologies for these values of G and temperature. The measured time constant of 10 μs is ∼100 times faster than the intrinsic time constant.

Applied Physics Letters•Sep 16, 1996•203 citations

Faster recovery time of a hot-electron transition-edge sensor by use of normal metal heat-sinks

Brice Calkins, Adriana E. Lita, Anna E. Fox +1 more

Transition-edge sensor microcalorimeters with recovery times near 1 μs have become highly desirable in quantum science applications as near-infrared single-photon detectors with photon-number resolving capability. Previously, the recovery times of these devices could be decreased only by changing device material or modifying the superconducting-to-normal transition. We demonstrate a method for improving this speed that uses a normal-metal heat-sink. This demonstration with tungsten devices realizes a factor of 4 decrease in recovery time without significantly affecting energy resolution. Our approach may enable the creation of high-efficiency transition-edge sensors with decay times short enough to operate with 80 MHz pulsed single-photon sources.

Applied Physics Letters•Dec 12, 2011•31 citations

Titanium Based Transition Edge Microcalorimeters for Optical Photon Measurements

Daiji Fukuda, R. M. T. Damayanthi, Akio Yoshizawa +4 more

Transition edge sensor microcalorimeters can be used in many optical quantum measurements because of its low dark counts, high quantum efficiency, and high resolving power of a photon number in weak light pulses. In order to increase count rates up to a few MHz, we have developed a titanium transition edge sensor for the optical measurements, and its performances were analysed. Titanium is one of the ideal superconductor because of its higher transition temperature and lower optical reflectance at 1.5 wavelength. Our titanium film was fabricated with electron-beam evaporation, and showed high residual resistance ratio of 3.5. The sharp superconducting transition also was found at 359 mK, which is close to the critical temperature in bulk. The fabricated device showed a fast response to pulsed laser illumination of 1.5 wavelength with the fall time constant of 300 ns. These features are very promising for high-speed single photon detection in many quantum optical measurements.

IEEE Transactions on Applied Superconductivity•Jun 1, 2007•29 citations

Low-noise microwave SQUID multiplexed readout of 38 x-ray transition-edge sensor microcalorimeters

Y. Nakashima, Fuminori Hirayama, Satoshi Kohjiro +15 more

We report very-low-noise, fast-response, middle-scale multiplexing in a microwave superconducting quantum interference device multiplexer (MW-Mux) as a transition-edge sensor (TES) readout. Our MW-Mux is able to read 40 channels with 500 kHz sampling and has a low readout noise of 0.9 μΦ0/Hz (where Φ0 is the magnetic flux quantum), equivalent to 9 pA/Hz. By contrast, a multiplexer of less than 10 pixels with 500 kHz sampling and ∼2 μΦ0/Hz readout noise has so far been reported in the literature. Owing to the 500 kHz sampling, our MW-Mux exhibits a fast response to detect a TES pulse with a rise time around 12 μs. We demonstrated simultaneous readout of 38 pixels from an array of x-ray TES microcalorimeters. The measured full-width values at half-maximum spectral resolution ranged from 2.79 to 4.56 eV, with a median value of 3.30 eV at 5.9 keV, including a ∼10% contribution of readout noise, i.e., 0.9–1.7 eV.

Applied Physics Letters•Sep 21, 2020•30 citations

Thermal fluctuation noise in a voltage biased superconducting transition edge thermometer

Henk F. C. Hoevers, A. Bento, M. P. Bruijn +4 more

The current noise at the output of a microcalorimeter with a voltage biased superconducting transition edge thermometer is studied in detail. In addition to the two well-known noise sources: thermal fluctuation noise from the heat link to the bath and Johnson noise from the resistive thermometer, a third noise source strongly correlated with the steepness of the thermometer is required to fit the measured noise spectra. Thermal fluctuation noise, originating in the thermometer itself, fully explains the additional noise. A simple model provides quantitative agreement between the observed and calculated noise spectra for all bias points in the superconducting transition.

Applied Physics Letters•Dec 25, 2000•64 citations
We investigate limits on the thermal-response time of superconducting transition-edge microcalorimeters. For operation at 0.1 K, we show that the lower limit on the response time of a superconducting transition-edge microcalorimeter is of order 1 μs due to the heat diffusion time, electrical instabilities, the amplifier noise, and the critical current of the superconducting film. The response time is not limited by self-heating effects and is independent of the intended photon energy. However, design constraints associated with the inductance of the bias circuit make it difficult to achieve the fastest response times for devices with heat capacities high enough for x-ray and gamma-ray detection.
We have developed a new type of x-ray detector based on a superconducting transition-edge thermometer operated near 100 mK. A superconducting quantum interference device is used to measure the current through the thermometer, and negative electrothermal feedback is used to improve the energy resolution and shorten the thermal time constant. We have used a detector mounted on a scanning electron microscope to measure the energy of titanium Kα (4.5 keV) fluorescence x rays with a resolution better than 14 eV full width at half-maximum. Using two other devices, we have measured an energy resolution for Joule heat pulses of 2.6 eV at 1 keV and 0.2 eV at 4 eV, the best reported for any calorimeter. An electrical noise equivalent power of 3×10−18 W/√Hz was also measured, suggesting the use of these detectors as infrared bolometers.
We present a theoretical analysis and experimental evaluation of a transition-edge superconducting bolometer for detecting infrared and millimeter waves. The superconducting film is voltage biased and the current is read by a superconducting quantum interference device ammeter. Strong electrothermal feedback maintains the sensor temperature within the transition, gives a current responsivity that is simply the inverse of the bias voltage, and reduces the response time by several orders of magnitude below the intrinsic time constant C/G. We evaluated a voltage-biased bolometer that operates on the Tc∼95 mK transition of a tungsten film with a thermal conductance of G∼1.2×10−9 W/K. As expected, the electrical noise equivalent power of 3.3×10−17/W√Hz is close to the thermal fluctuation noise limit and is lower than that of other technologies for these values of G and temperature. The measured time constant of 10 μs is ∼100 times faster than the intrinsic time constant.
Transition-edge sensor microcalorimeters with recovery times near 1 μs have become highly desirable in quantum science applications as near-infrared single-photon detectors with photon-number resolving capability. Previously, the recovery times of these devices could be decreased only by changing device material or modifying the superconducting-to-normal transition. We demonstrate a method for improving this speed that uses a normal-metal heat-sink. This demonstration with tungsten devices realizes a factor of 4 decrease in recovery time without significantly affecting energy resolution. Our approach may enable the creation of high-efficiency transition-edge sensors with decay times short enough to operate with 80 MHz pulsed single-photon sources.
Transition edge sensor microcalorimeters can be used in many optical quantum measurements because of its low dark counts, high quantum efficiency, and high resolving power of a photon number in weak light pulses. In order to increase count rates up to a few MHz, we have developed a titanium transition edge sensor for the optical measurements, and its performances were analysed. Titanium is one of the ideal superconductor because of its higher transition temperature and lower optical reflectance at 1.5 wavelength. Our titanium film was fabricated with electron-beam evaporation, and showed high residual resistance ratio of 3.5. The sharp superconducting transition also was found at 359 mK, which is close to the critical temperature in bulk. The fabricated device showed a fast response to pulsed laser illumination of 1.5 wavelength with the fall time constant of 300 ns. These features are very promising for high-speed single photon detection in many quantum optical measurements.
We report very-low-noise, fast-response, middle-scale multiplexing in a microwave superconducting quantum interference device multiplexer (MW-Mux) as a transition-edge sensor (TES) readout. Our MW-Mux is able to read 40 channels with 500 kHz sampling and has a low readout noise of 0.9 μΦ0/Hz (where Φ0 is the magnetic flux quantum), equivalent to 9 pA/Hz. By contrast, a multiplexer of less than 10 pixels with 500 kHz sampling and ∼2 μΦ0/Hz readout noise has so far been reported in the literature. Owing to the 500 kHz sampling, our MW-Mux exhibits a fast response to detect a TES pulse with a rise time around 12 μs. We demonstrated simultaneous readout of 38 pixels from an array of x-ray TES microcalorimeters. The measured full-width values at half-maximum spectral resolution ranged from 2.79 to 4.56 eV, with a median value of 3.30 eV at 5.9 keV, including a ∼10% contribution of readout noise, i.e., 0.9–1.7 eV.
The current noise at the output of a microcalorimeter with a voltage biased superconducting transition edge thermometer is studied in detail. In addition to the two well-known noise sources: thermal fluctuation noise from the heat link to the bath and Johnson noise from the resistive thermometer, a third noise source strongly correlated with the steepness of the thermometer is required to fit the measured noise spectra. Thermal fluctuation noise, originating in the thermometer itself, fully explains the additional noise. A simple model provides quantitative agreement between the observed and calculated noise spectra for all bias points in the superconducting transition.