arXiv:2010.15244

Compact transportable 171Yb+ single-ion optical fully automated clock with 4.9E-16 relative instability

Timur Abbasov, Makarenko, Konstantin, I. Sherstov, Mikhail A. Aksenov, Zalivako, Ilya, Ilya A. Semerikov, Borisenko, Alexander, Ksenia Yu. Khabarova, N. Kolachevsky, Sergey Chepurov, Alexei Taichenachev, S. A. Bagaev, Anton V. Tausenev

Published Oct 24, 2020• 1 citation
arXivHTMLPDF

Abstract

The paper describes the results achieved in the development of the compact transportable fully automated optical clock based on a single 171Yb+ ion in a radiofrequency (RF) quadrupole trap. The resulted measurements demonstrated the 4.9E-16 output RF signal relative instability on 1000 s integration time with 298.1 kg weight, 0.921 volume, and 2.766 kW input power consumption of the device. A transformation of the ultrastable optical signal into the RF range was performed via the optical frequency comb with a supercontinuum fiber laser generator. The transformation was conducted without loss of initial stability and accuracy characteristics of the signal.

DOI
10.48550/arxiv.2010.15244

Related papers

Compact Transportable Optical Standard Based on a Single 171Yb+ Ion (“YBIS” Project)

Ilya A. Semerikov, Ksenia Yu. Khabarova, Ilia V. Zalivako +2 more

The schematic diagram of the compact transportable optical frequency standard based on the quadrupole transition in a single 171 Yb + ion with a relative instability at the levelx of 5 · 10 −16 , less than 1m 3 in volume and less than 300 kg in weight is presented. The ionic trap and laser systems are described. Trapping and laser cooling of the 174 Yb + ion are experimentally demonstrated.

Bulletin of the Lebedev Physics Institute•Nov 1, 2018•19 citations

Towards compact transportable optical clock based on 171Yb+

Ilia V. Zalivako, Ksenia Yu. Khabarova, Ilya A. Semerikov +7 more

We report on a new project aimed for development of a transportable compact optical clock based on single171Yb+ion trapped in a ion Paul trap. The goal is to reach relative frequency instability level of 5E-16 at 24 hours averaging time.

European Frequency and Time Forum•Apr 1, 2018•1 citation

Compact ultrastable laser system for spectroscopy of 2S1/2 → 2D3/2 quadrupole transition in 171Yb+ ion

Ilia V. Zalivako, Ilya A. Semerikov, Alexander S. Borisenko +7 more

Abstract We report the results of studying a compact laser system designed for manipulating a quantum state of the optical qubit based on the 2S1/2 → 2D3/2 quadrupole transition in the 171Yb+ ion at a wavelength of 435.5 nm. An emission power of the laser system reaches 500 μW at λ = 435.5 nm and the relative frequency instability of at most 3 × 10−15 is achieved at averaging intervals from 0.5 to 50 s with a subtracted linear frequency drift. The compactness of the developed system makes it possible to employ it in transportable systems including optical clocks.

Quantum Electronics•Aug 26, 2020•15 citations

Industrial 171Yb+ single-ion optical clock with systematic uncertainty below 2×10^−17

J. Stühler, A. Friedenauer, P. Thoumany +3 more

We present the setup of our optical frequency standard (OFS) “TOPTICLOCK”. It is based on the 2S1/2 (F=0) – 2D3/2 (F=2) electric quadrupole transition at 435.5 nm of a single laser-cooled and trapped 171Yb+ ion. All required subsystems are contained in two transportable 19” rack compartments. By comparing it with a laboratory-style OFS of the German National Metrology Institute (PTB), which operates on the 467 nm electric octupole transition, using a TOPTICA DFC optical frequency comb, a frequency instability of 5×10-15/√τ with systematic uncertainty below 2×10-17 was demonstrated. The performance of TOPTICLOCK was also benchmarked by directly comparing it to another OFS operating also at the 435.5 nm clock transition. An agreement between both systems with a total uncertainty well below 10-17 was obtained.

Optical and Quantum Sensing and Precision Metrology II•Mar 5, 2026•2 citations

Optical frequency standard based on a trapped171Yb+ion

Chr. Tamm, T. Schneider, E. Peik

We describe experimental investigations on an optical frequency standard based on a single laser cooled 171Yb+ ion confined in a radiofrequency Paul trap. The 435.5 nm electric-quadrupole transition from the 2S1/2(F=0) groudn state to the 2D3/2(F=2) state is probed using a diode laser system whose short-term frequency stability is derived from an environmentally isolated ULE cavity. The transition is resolved with a Foruier-limited full halfwidth of approximately 30 Hz. The absolute optical transition frequency is determined using a femtosecond frequency comb generator and a Cs fountain clock as the frequency reference. The measured frequency value is 688 358 979 321 Hz at an ambient blackbody radiation field of 300 K. The 1σ fractional uncertainty of this frequency measurement is 1×10-14.

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE•Oct 8, 2003•0 citations

AN OPTICAL CLOCK BASED ON A SINGLE TRAPPED 88 Sr + ION

Helen S. Margolis, G. P. Barwood, G. HUANG +6 more

The 5 s 2 S 1/2 -4 d 2 D 5/2 electric quadrupole transition in a single trapped and laser-cooled 88 Sr + ion provides the reference for a highly stable and accurate optical clock. This paper describes recent progress on the 88 Sr + optical clocks under development at NPL. An ultrastable laser at 674 nm is used to excite the clock transition and a fractional frequency resolution of 2 × 10 -14 has been achieved, corresponding to a linewidth of 9 Hz. A relative frequency instability of 2 × 10 -15 at 5000 s has been observed in the comparison between two similar systems and the fractional uncertainty in the absolute frequency of the clock transition is 3.8 × 10 -15 . Improvements to the experimental arrangement are in progress and are expected to lead to a 88 Sr + optical clock with stability and reproducibility exceeding that of the primary caesium standard.

Symposium on Frequency Standards and Metrology•May 1, 2009•0 citations
The paper describes the results achieved in the development of the compact transportable fully automated optical clock based on a single 171Yb+ ion in a radiofrequency (RF) quadrupole trap. The resulted measurements demonstrated the 4.9E-16 output RF signal relative instability on 1000 s integration time with 298.1 kg weight, 0.921 volume, and 2.766 kW input power consumption of the device. A transformation of the ultrastable optical signal into the RF range was performed via the optical frequency comb with a supercontinuum fiber laser generator. The transformation was conducted without loss of initial stability and accuracy characteristics of the signal.
Abstract We report the results of studying a compact laser system designed for manipulating a quantum state of the optical qubit based on the 2S1/2 → 2D3/2 quadrupole transition in the 171Yb+ ion at a wavelength of 435.5 nm. An emission power of the laser system reaches 500 μW at λ = 435.5 nm and the relative frequency instability of at most 3 × 10−15 is achieved at averaging intervals from 0.5 to 50 s with a subtracted linear frequency drift. The compactness of the developed system makes it possible to employ it in transportable systems including optical clocks.
We present the setup of our optical frequency standard (OFS) “TOPTICLOCK”. It is based on the 2S1/2 (F=0) – 2D3/2 (F=2) electric quadrupole transition at 435.5 nm of a single laser-cooled and trapped 171Yb+ ion. All required subsystems are contained in two transportable 19” rack compartments. By comparing it with a laboratory-style OFS of the German National Metrology Institute (PTB), which operates on the 467 nm electric octupole transition, using a TOPTICA DFC optical frequency comb, a frequency instability of 5×10-15/√τ with systematic uncertainty below 2×10-17 was demonstrated. The performance of TOPTICLOCK was also benchmarked by directly comparing it to another OFS operating also at the 435.5 nm clock transition. An agreement between both systems with a total uncertainty well below 10-17 was obtained.
We describe experimental investigations on an optical frequency standard based on a single laser cooled 171Yb+ ion confined in a radiofrequency Paul trap. The 435.5 nm electric-quadrupole transition from the 2S1/2(F=0) groudn state to the 2D3/2(F=2) state is probed using a diode laser system whose short-term frequency stability is derived from an environmentally isolated ULE cavity. The transition is resolved with a Foruier-limited full halfwidth of approximately 30 Hz. The absolute optical transition frequency is determined using a femtosecond frequency comb generator and a Cs fountain clock as the frequency reference. The measured frequency value is 688 358 979 321 Hz at an ambient blackbody radiation field of 300 K. The 1σ fractional uncertainty of this frequency measurement is 1×10-14.
The 5 s 2 S 1/2 -4 d 2 D 5/2 electric quadrupole transition in a single trapped and laser-cooled 88 Sr + ion provides the reference for a highly stable and accurate optical clock. This paper describes recent progress on the 88 Sr + optical clocks under development at NPL. An ultrastable laser at 674 nm is used to excite the clock transition and a fractional frequency resolution of 2 × 10 -14 has been achieved, corresponding to a linewidth of 9 Hz. A relative frequency instability of 2 × 10 -15 at 5000 s has been observed in the comparison between two similar systems and the fractional uncertainty in the absolute frequency of the clock transition is 3.8 × 10 -15 . Improvements to the experimental arrangement are in progress and are expected to lead to a 88 Sr + optical clock with stability and reproducibility exceeding that of the primary caesium standard.