Design concepts for an improved integrated scanning SQUID

Nicholas C. Koshnick, John R. Kirtley, Kathryn A. Moler

Published Feb 8, 2010• 4 citations
arXivHTMLPDF

Abstract

In this paper we discuss design concepts for increasing the spatial resolution, improving the sensitivity, and reducing the invasiveness in scanning Superconducting Quantum Interference Device (SQUID) microscope sensors with integrated flux pickup loops. This can be done not only by reducing the ground-rule line widths and spacings, but also by taking advantage of planarization, reducing flux noise through reducing the SQUID inductance, and reducing back-action through dispersive readouts or on-chip filtering.

DOI
10.48550/arxiv.1002.1529

Related papers

A terraced scanning super conducting quantum interference device susceptometer with submicron pickup loops

Nicholas C. Koshnick, M. E. Huber, Julie A. Bert +4 more

Superconducting quantum interference devices (SQUIDs) can have excellent spin sensitivity depending on their magnetic flux noise, pickup loop diameter, and distance from the sample. We report a family of scanning SQUID susceptometers with terraced tips that position the pickup loops 300nm from the sample. The 600nm–2μm pickup loops, defined by focused ion beam, are integrated into a 12-layer optical lithography process allowing flux-locked feedback, in situ background subtraction and optimized flux noise. These features enable a sensitivity of ∼70 electron spins per root hertz at 4K.

arXiv:0901.0308•Applied Physics Letters•Dec 15, 2008•66 citations

A micro-SQUID with dispersive readout for magnetic scanning microscopy

F. Foroughi, J.-M. Mol, T. Müller +3 more

We have designed and characterized a micro-SQUID with dispersive readout for use in low temperature scanning probe microscopy systems. The design features a capacitively shunted RF SQUID with a tunable resonance frequency from 5 to 12 GHz, micrometer spatial resolution, and integrated superconducting coils for local application of magnetic fields. The SQUID is operated as a nonlinear oscillator with a flux- and power-dependent resonance frequency. Measurements for device characterization and noise benchmarking were carried out at 4 K. The measured flux noise above 10 kHz at 4 K is 80 nΦ0 Hz−1∕2 at a bandwidth of 200 MHz. Estimations suggest that one can benefit from parametric gain based on inherent nonlinearity of the Josephson junction and reduce the flux noise to 30 nΦ0Hz–1∕2 at 100 mK, which corresponds to 10.6 μBHz–1∕2 for a magnetic moment located at the center of the pickup loop.

arXiv:1803.07636•Applied Physics Letters•Jun 18, 2018•24 citations

dc SQUID Design with Femtotesla Sensitivity for Quantum-Ready Readouts

I. Sochnikov, D. Davino, Beena Kalisky

Among some of the current uses of dc superconducting quantum interference devices (SQUIDs) are qubit readouts and sensors for probing the properties of quantum materials. We present a gradiometric niobium SQUID design with state-of-the-art sensitivity in the femtotesla range, which can be tuned to specific readout requirements. The sensor is a next-generation fractional SQUID with a tightly optimized input coil and a combination of various measures known for restraining parasitic resonances and other detrimental effects. Our design combines the practical usefulness of well-defined pickup loops for well-defined imaging kernel and tunable probing applications with a fractionalization approach to reduce undesired inductances. In addition, our modeling predicts small dimensions for these planar sensors. These features make them of relevance for material studies and for the detection of magnetic fields in small volumes, e.g., as part of a cryogenic scanning quantum imaging apparatus for efficient diagnostics and quantum device readouts. This manuscript will benefit scientists and engineers working on quantum-computing technologies by clarifying potential general misconceptions about dc SQUID optimization alongside the introduction of the flexible compact dc SQUID design.

arXiv:2003.07702•Physical Review Applied•Jul 8, 2020•14 citations

Measurement of the spatial sensitivity of miniature SQUIDs using magnetic-tipped STM

Patrick Josephs-Franks, L Hao, Alexander Ya. Tzalenchuk +5 more

Large area superconducting quantum interference devices (SQUIDs) are known to be ultimate sensors of magnetic flux. By diminishing the size of a SQUID loop, the energy sensitivity of the device can be increased to near quantum-limited operation whilst also making it less sensitive to external magnetic field noise and improving its coupling to any nano-scale magnetic particle within the loop. This makes it an ideal nano-sensor for magnetic particles including single molecules. It is, however, important to optimize the coupling between any magnetic nano-particle placed within the SQUID and the SQUID itself in order to achieve maximum sensitivity for the spin detection. We have made some preliminary calculations of the expected SQUID response to a magnetic particle placed at different locations within the SQUID device, and using a magnetic tip in a scanning tunnelling microscope (STM) we are able to produce a spatial map of the SQUID sensitivity. Initial results on a 3 µm diameter loop SQUID are presented to demonstrate this method.

Superconductor Science and Technology•Nov 13, 2003•23 citations

Octagonal washer DC SQUIDs and integrated susceptometers fabricated in a planarized sub- mu m Nb-AlO/sub x/-Nb technology

M. B. Ketchen, Dale J. Pearson, Kevin G. Stawiasz +9 more

The authors have designed, fabricated, and measured a number of Nb-AlO/sub x/-Nb octagonal washer DC superconducting quantum interference devices (SQUIDs) and miniature DC SQUID susceptometers having minimum feature size down to 0.5 mu m. With SQUID inductance values on the order of 100 pH, typical noise performance is better than 1 mu Phi /sub 0// square root Hz. The small minimum feature size of input coils and pickup loop structures will facilitate tight coupling to a wide variety of systems ranging from submicron particles and structures to conventional approximately mu H input circuits. A single-washer SQUID with an 80-turn 0.5- mu m-linewidth, 630-nH input coil has user-friendly V- Phi (voltage-flux) curves and a coupled energy sensitivity of 20 h at 4.2 K. Susceptometers with pickup loops ranging from 20- mu m to 0.8- mu m across have very user-friendly V- Phi curves and a resolution of order 100 mu /sub B// square root Hz at 4.2 K for devices with the smallest loops.>

IEEE Transactions on Applied Superconductivity•Mar 1, 1993•27 citations

Low-noise SQUIDs with large transfer - two stage SQUIDs based on DROSs

M. Podt, Jakob Flokstra, Horst Rogalla

We have realized a two-stage integrated superconducting quantum interference device (SQUID) system with a closed loop bandwidth of 2.5 MHz, operated in a direct voltage readout mode. The corresponding flux slew rate was 1.3×105 Φ0/s and the measured white flux noise was 1.3 μΦ0/√Hz at 4.2 K. The system is based on a conventional dc SQUID with a double relaxation oscillation SQUID (DROS) as the second stage. Because of the large flux-to-voltage transfer, the sensitivity of the system is completely determined by the sensor SQUID and not by the DROS or the room-temperature preamplifier. Decreasing the Josephson junction area enables a further improvement of the sensitivity of the two-stage SQUID systems.

•Aug 26, 2001•0 citations
In this paper we discuss design concepts for increasing the spatial resolution, improving the sensitivity, and reducing the invasiveness in scanning Superconducting Quantum Interference Device (SQUID) microscope sensors with integrated flux pickup loops. This can be done not only by reducing the ground-rule line widths and spacings, but also by taking advantage of planarization, reducing flux noise through reducing the SQUID inductance, and reducing back-action through dispersive readouts or on-chip filtering.
Superconducting quantum interference devices (SQUIDs) can have excellent spin sensitivity depending on their magnetic flux noise, pickup loop diameter, and distance from the sample. We report a family of scanning SQUID susceptometers with terraced tips that position the pickup loops 300nm from the sample. The 600nm–2μm pickup loops, defined by focused ion beam, are integrated into a 12-layer optical lithography process allowing flux-locked feedback, in situ background subtraction and optimized flux noise. These features enable a sensitivity of ∼70 electron spins per root hertz at 4K.
We have designed and characterized a micro-SQUID with dispersive readout for use in low temperature scanning probe microscopy systems. The design features a capacitively shunted RF SQUID with a tunable resonance frequency from 5 to 12 GHz, micrometer spatial resolution, and integrated superconducting coils for local application of magnetic fields. The SQUID is operated as a nonlinear oscillator with a flux- and power-dependent resonance frequency. Measurements for device characterization and noise benchmarking were carried out at 4 K. The measured flux noise above 10 kHz at 4 K is 80 nΦ0 Hz−1∕2 at a bandwidth of 200 MHz. Estimations suggest that one can benefit from parametric gain based on inherent nonlinearity of the Josephson junction and reduce the flux noise to 30 nΦ0Hz–1∕2 at 100 mK, which corresponds to 10.6 μBHz–1∕2 for a magnetic moment located at the center of the pickup loop.
Among some of the current uses of dc superconducting quantum interference devices (SQUIDs) are qubit readouts and sensors for probing the properties of quantum materials. We present a gradiometric niobium SQUID design with state-of-the-art sensitivity in the femtotesla range, which can be tuned to specific readout requirements. The sensor is a next-generation fractional SQUID with a tightly optimized input coil and a combination of various measures known for restraining parasitic resonances and other detrimental effects. Our design combines the practical usefulness of well-defined pickup loops for well-defined imaging kernel and tunable probing applications with a fractionalization approach to reduce undesired inductances. In addition, our modeling predicts small dimensions for these planar sensors. These features make them of relevance for material studies and for the detection of magnetic fields in small volumes, e.g., as part of a cryogenic scanning quantum imaging apparatus for efficient diagnostics and quantum device readouts. This manuscript will benefit scientists and engineers working on quantum-computing technologies by clarifying potential general misconceptions about dc SQUID optimization alongside the introduction of the flexible compact dc SQUID design.
Large area superconducting quantum interference devices (SQUIDs) are known to be ultimate sensors of magnetic flux. By diminishing the size of a SQUID loop, the energy sensitivity of the device can be increased to near quantum-limited operation whilst also making it less sensitive to external magnetic field noise and improving its coupling to any nano-scale magnetic particle within the loop. This makes it an ideal nano-sensor for magnetic particles including single molecules. It is, however, important to optimize the coupling between any magnetic nano-particle placed within the SQUID and the SQUID itself in order to achieve maximum sensitivity for the spin detection. We have made some preliminary calculations of the expected SQUID response to a magnetic particle placed at different locations within the SQUID device, and using a magnetic tip in a scanning tunnelling microscope (STM) we are able to produce a spatial map of the SQUID sensitivity. Initial results on a 3 µm diameter loop SQUID are presented to demonstrate this method.
The authors have designed, fabricated, and measured a number of Nb-AlO/sub x/-Nb octagonal washer DC superconducting quantum interference devices (SQUIDs) and miniature DC SQUID susceptometers having minimum feature size down to 0.5 mu m. With SQUID inductance values on the order of 100 pH, typical noise performance is better than 1 mu Phi /sub 0// square root Hz. The small minimum feature size of input coils and pickup loop structures will facilitate tight coupling to a wide variety of systems ranging from submicron particles and structures to conventional approximately mu H input circuits. A single-washer SQUID with an 80-turn 0.5- mu m-linewidth, 630-nH input coil has user-friendly V- Phi (voltage-flux) curves and a coupled energy sensitivity of 20 h at 4.2 K. Susceptometers with pickup loops ranging from 20- mu m to 0.8- mu m across have very user-friendly V- Phi curves and a resolution of order 100 mu /sub B// square root Hz at 4.2 K for devices with the smallest loops.>
We have realized a two-stage integrated superconducting quantum interference device (SQUID) system with a closed loop bandwidth of 2.5 MHz, operated in a direct voltage readout mode. The corresponding flux slew rate was 1.3×105 Φ0/s and the measured white flux noise was 1.3 μΦ0/√Hz at 4.2 K. The system is based on a conventional dc SQUID with a double relaxation oscillation SQUID (DROS) as the second stage. Because of the large flux-to-voltage transfer, the sensitivity of the system is completely determined by the sensor SQUID and not by the DROS or the room-temperature preamplifier. Decreasing the Josephson junction area enables a further improvement of the sensitivity of the two-stage SQUID systems.