Single Ion Addressing for Long Ion Strings
GIORGIO CIOCCA
Abstract
This Master’s thesis was dedicated to the design, assembly, and experimental characterisation of a novel single-ion addressing system for the linear Paul trap of the QSim experiment at IQOQI Innsbruck. The recently upgraded linear trap in the experiment, which is capable of confining chains of 100 or more Calcium-40 ions, made it necessary to build a suitable single-site addressing apparatus. Said new addressing system should be capable of handling an expected chain length of at least 460 µm and an inter-ion spacing as small as 2.6 µm, all while maintaining a resonant crosstalk level below 0.3. To overcome the limitations of addressing such extended strings with a single beam, a novel structural geometry was implemented: the system utilizes two independent Acousto-Optic Deflectors (AODs), each responsible for steering a laser beam over one half of the ion chain, the outputs of which are combined via a beam splitter onto a common path. The optical design was guided by geometrical and practical constraints, together with ray and physical optics principles, relying on Zemax OpticStudio simulations to verify the performance of the design. A Keplerian beam expander, featuring a Hastings triplet and a composite collimation stage, was chosen to achieve the desired beam size right before the final focusing onto the ions. Such a design was chosen in order to reduce aberrations of the beam to a minimum. Furthermore, custom mechanical mounts were designed in Solidworks to ensure maximum control and modularity of the resulting setup within the tight constraints of the main experiment. Following its construction, the system was thoroughly characterised on a test bench, using a substitute objective for preliminary tests. The output beam was reconstructed using sub-pixel resolution using the Drizzle algorithm, yielding focused waist radii ranging between 1.1 µm and 1.5 µm. Such beam radii were within the design tolerances required to maintain acceptable levels of crosstalk. Extrapolation of the beam steering demonstrated an addressable length of 549(5) µm, comfortably exceeding the 460 µm requirement. Finally, the setup was integrated into the main experiment and validated through a series of proof-of-concept operations on actual ion chains. The measured crosstalk across a 100-ion chain remained below the 0.3 threshold, peaking at 0.186(2) for one AOD and 0.206(3) for the other. The setup successfully allows for the preparation of benchmark states such as Néel states with roughly 90% fidelity. Furthermore, the ability of the system to create arbitrary product states and to reconstruct them via quantum state tomography across the whole crystal was tested by preparing |xyzxyz . . . ⟩ states and performing both single-site and pairwise state tomography on them.