Optically pumped magnetometers usually rely on optical pumping using circularly or linearly polarized light. We study here zero-field magnetometers pumped with elliptically polarized light, preparing both atomic orientation and alignment with complementary geometries. We start by extending the three-step approach for elliptically polarized pumping. This allows us to study the Hanle effect in elliptical polarization by comparing the analytical absorption signals with experiments made on 4He metastable state. We then study parametric resonance magnetometers based on elliptical polarization by using the dressed-atom formalism with one and two radio-frequency fields. The results show good agreement with the experimental measurements and interesting perspectives for magnetometry where symmetry breaking by pumping light is mitigated.
Quantum sensing has appeared as a promising technique to push the sensitivities of sensors to inherent classical physical limitations. In particular, quantum magnetometers opens the possibilities of new range of applications where extreme sensitivities are needed. A zero field Optically Pumped Magnetometer (OPM), coupled to a conical magnetic guide (MG) is known as a promising technique to improve the spatial resolution of the sensor. This coupling, though, can produce disturbances in the optimal sensing point. In this work a set of compensating coils is designed to reduce the impact of the magnetic guide in the initial sub-15 $\fT/\sqrt{Hz}$ sensitivity of the commercial QuSpin QZFM. First, the system and different strategies are studied theoretically and later implemented experimentally, improving the sensitivity in the OPM-MG by an order of magnitude, from 100 fT/Hz originally, to 10 fT/Hz.
We report the development and performance of an atomic magnetometer (AM) utilizing the Raman-driven spin noise spectroscopy (RDSNS) technique, which measures the intrinsic spin fluctuations of an atomic medium subjected to a coherent two-photon optical drive. The RDSNS signal is noninvasively detected using a far-off resonant weak probe light utilizing a polarimetric detection scheme. This overcomes the limitations of the small dynamic range and low bandwidth of traditional optically pumped magnetometers (OPMs) or AMs operating in the spin exchange relaxation-free (SERF) regime. We experimentally demonstrate, using both paraffin-coated and buffer-gas-filled vapor cell containing neutral rubidium atoms, that the dynamic range of such magnetometers can reach beyond 150 G while maintaining a dc magnetometer resolution of around 30 pT and a magnetic noise sensitivity of around30pT/Hzat a bandwidth of 100 Hz and a magnetic field of about 0.6 mT. Being an all-optical setup, this instrument is largely immune to interference with stray radio frequency fields and has the potential to be miniaturized and field deployable. One of the practical applications would be to measure variations in geomagnetic fields precisely without having to shield the magnetometer.