Abstract This work proposes a novel method for in situ measurement of the non‐orthogonal angles of magnetic field coils based on single‐beam magnetic resonance. The orthogonality of magnetic field coils is crucial for ensuring the accuracy of atomic magnetometers. The theoretical principles are derived to establish a measurement method for the non‐orthogonal angles. The proposed method requires only one single beam of light to perform in situ measurements of non‐orthogonal angles in the YZ and XZ directions. The experimental results demonstrate that the angle between the Y and Z coils is 89.3001 ± 0.0045°, a value that remains stable despite variations in experimental parameters such as temperature and light power, thereby confirming the robustness of the method. Additionally, the measured gyromagnetic ratio is 6.9963 ± 0.0005 Hz nT −1 , which closely matches the theoretical value of 6.9958 Hz nT −1 , further validating the method's accuracy. The proposed method provides a more straightforward approach and has significant potential for widespread application across various atomic magnetometers, especially in miniaturized configurations. Furthermore, this method may be extended to the field of atomic gyroscopes.
Abstract This work proposes a novel method for in situ measurement of the non‐orthogonal angles of magnetic field coils based on single‐beam magnetic resonance. The orthogonality of magnetic field coils is crucial for ensuring the accuracy of atomic magnetometers. The theoretical principles are derived to establish a measurement method for the non‐orthogonal angles. The proposed method requires only one single beam of light to perform in situ measurements of non‐orthogonal angles in the YZ and XZ directions. The experimental results demonstrate that the angle between the Y and Z coils is 89.3001 ± 0.0045°, a value that remains stable despite variations in experimental parameters such as temperature and light power, thereby confirming the robustness of the method. Additionally, the measured gyromagnetic ratio is 6.9963 ± 0.0005 Hz nT −1 , which closely matches the theoretical value of 6.9958 Hz nT −1 , further validating the method's accuracy. The proposed method provides a more straightforward approach and has significant potential for widespread application across various atomic magnetometers, especially in miniaturized configurations. Furthermore, this method may be extended to the field of atomic gyroscopes.
An in situ simultaneous measurement of magnetic coil constants and nonorthogonal angles method is proposed using the atomic magnetometer. Based on the transient response of the atomic spin polarization under the different magnetic fields, we establish the corresponding measurement model of magnetic coil constants and nonorthogonal angles through the relationship between the Larmor precession frequency and the external magnetic field. By this model, we experimentally obtained the coil constants of x , y, and z axes are, respectively, 9.04 ± 0.10, 8.94 ± 0.07, and 4.34 ± 0.02 nT / V in our system. The nonorthogonal angles are 89.04 deg between x axis and y axis, 88.60 deg between y axis and z axis, and 91.15 deg between z axis and x axis. This method can be used to simultaneously measure magnetic coil constants and nonorthogonal angles without relying on external instruments, which provides a basic guarantee for the accurate calculation sensitivity of spin-exchange-relaxation-free atomic magnetometers, especially for miniaturized atomic magnetometers.
The spin exchange relaxation free (SERF) atomic magnetometer has been a research hotspot due to its ultra-high sensitivity. The SERF atomic magnetometer's sensitivity is subject to the accuracy of its compensating magnetic field and the calibration signal. These two processes need a tri-axial coil system to generate the required magnetic fields, which are directly affected by the non-orthogonality of the tri-axial coil system, especially the non-orthogonality between the x- and y-axes coils. In order to solve this problem, an ingenious and effective method for measuring the non-orthogonal angle between the x- and y-axes in a magnetic coil system has been demonstrated in this paper for the first time. This measurement method was only based on the dynamics of the SERF atomic magnetometer and did not need any additional calibration equipment. A large dc magnetic field Bywas applied to the y-axis, such that the magnetic field Byxprojected from Byon to the x-axis could be measured. By then solving the relevant trigonometric function, we could obtain the non-orthogonal angle. Using this method, we experimentally obtained the non-orthogonal angle of about 84.08° between the x- and y-axes saddle-shaped coils, with a variance of about 0.14°. Finally, the correctness of this method for calculating the non-orthogonal angle was analyzed.
In this article, we propose an in situ method to calibrate the magnetic-field coil systems in optically-pumped magnetometers. The method is based on parametric resonance induced by nonresonant radio-frequency (RF) fields, and is especially suitable for microfabricated single-beam zero-field atomic magnetometers. We have experimentally and theoretically demonstrated this method with a single-beam spin-exchange-relaxation-free (SERF) magnetometer, and the experiment results are in line with theoretical expectations.
The calibration results of the triaxial coils affect the atomic magnetometer’s measurement sensitivity. For precise in situ calibration, we first studied the transient atomic dynamics of a dual-beam spin-exchange relaxation free (SERF) atomic magnetometer under alternating magnetic fields. We obtained the corresponding dynamic free-induction-decay (FID) curves by alternately applying square wave signals to thex-,y-, andz-axis coil, and then extracted the precession frequencies by fast Fourier transform (FFT) to achieve a decoupled calibration of the triaxial coil constants. Our calibration results were in better agreement under different experimental conditions than traditional parameter fitting methods with a relative error of less than 1.5% compared to conventional fluxgate calibration methods. Furthermore, we simultaneously measured nonorthogonal angles between each pair of axes: x–y, y–z, and z–x, respectively, which are used to correct the magnetic fields error. This method is suitable for coil calibration and nonorthogonal correction of integrated and miniaturized dual beam magnetometers.
Abstract In this study, an in situ measurement method of spin polarization in single‐beam atomic magnetometer is presented. Modeling magnetic linewidth by incorporating nonlinear attenuation along the propagation direction of light and employing the output response equation of magnetometers at zero‐field resonance. Utilizing the model for fitting to extract relaxation rate information, the photon absorption cross‐sections at three temperatures are estimated to be , , and , respectively. Further, the spin polarization curves are obtained by introducing the spin polarization equations. Given the diminishing marginal utility resulting from the enhancement of performance through increased spin polarization, long‐term sensitivity is employed to determine optimal spin polarization range. The optimal long‐term sensitivity for four displayed states is (@31.5Hz), while the ultimate minimum sensitivity is 10.7 (@31.5Hz). The optimal spin polarization ranges at three temperatures are 78.9%–87.1%, 82.3%–88.5%, and 84.5%–88.4%, respectively. The measurement method applied to single‐beam magnetometer can be easily extended to multi‐axis vector magnetometer. The determination of optimal spin polarization range is of great significance for the performance and stability of the magnetometer system, improving the sensitivity of magnetic field measurement, and ensuring its widespread application in bio‐magnetic measurement.
Electrical heating elements, which are widely used to heat the vapor cell of ultrasensitive atomic magnetometers, inevitably produce a magnetic field interference. In this paper, we propose a novel measurement method of the amplitude of electrical-heating-induced magnetic field for an atomic magnetometer. In contrast to conventional methods, this method can be implemented in the atomic magnetometer itself without the need for extra magnetometers. It can distinguish between different sources of magnetic fields sensed by the atomic magnetometer, and measure the three-axis components of the magnetic field generated by the electrical heater and the temperature sensor. The experimental results demonstrate that the measurement uncertainty of the heater’s magnetic field is less than 0.2 nT along the x-axis, 1.0 nT along the y-axis, and 0.4 nT along the z-axis. The measurement uncertainty of the temperature sensor’s magnetic field is less than 0.02 nT along all three axes. This method has the advantage of measuring the in-situ magnetic field, so it is especially suitable for miniaturized and chip-scale atomic magnetometers, where the cell is extremely small and in close proximity to the heater and the temperature sensor.