IEEE Sensors Journal

Towards the Non-Zero Field Cesium Magnetic Sensor Array for Magnetoencephalography

Mikhail V. Petrenko, Sergei P. Dmitriev, Anatoly S. Pazgalev, Alexei E. Ossadtchi, Anton K. Vershovskii

Published Jun 15, 2021• 24 citations
Publisher

Abstract

Magnetic sensors developed for application in magnetoencephalography must meet a number of requirements; the main ones are compactness, sensitivity and response speed. We present a quantum optically pumped atomic sensor with cell volume of 0.5 cm3that meets these requirements and is operable in nonzero magnetic fields. The ultimate sensitivity of the sensor was estimated as (using the criteria of the ratio of the slope of the magnetic resonance signal to the shot noise spectral density) to be better than 5 fT/ √Hz. The actual sensitivity, measured in a gradiometric scheme, reaches 13 fT/ √Hz per sensor. We also present a novel and fast algorithm for optimization of the geometric properties of non-zero field sensor array with respect to maximization of the information transfer rate for cortical sources.

DOI
10.1109/jsen.2021.3089455

Related papers

All-Optical Magnetometric Sensor for Magnetoencephalography and Ultralow Field Tomography

Anton K. Vershovskii, Anatoly S. Pazgalev, Mikhail V. Petrenko

Abstract A variant of the scheme of a magnetometric sensor based on cesium atomic vapor is proposed and experimentally investigated. The sensor uses magnetic resonance excitation by modulated light of an hyperfine optical pumping that is transverse to the magnetic field. It is shown that, for a cell with a volume of 0.125 cm3, the variational sensitivity of such a scheme, estimated from the ratio of the steepness of the signal in the center of the magnetic resonance to the shot noise of the detecting radiation, reaches a level of <10 fT/Hz1/2 in the frequency band width of the order of 850 Hz. The sensor, which does not emit radio frequency fields, is designed for use in magnetoencephalographic complexes. Possible ways to improve the performance of the scheme for detecting relatively fast (~4.2 kHz in a field of 0.1 mT) signals of the precession of proton magnetic moments in promising ultra-weak field tomography schemes are considered.

Technical Physics Letters•Sep 1, 2020•12 citations

Optically pumped non-zero field magnetometric sensor for the magnetoencephalographic systems using intra-cavity contacted VCSELs with rhomboidal oxide current aperture

Mikhail A. Bobrov, Sergey A. Blokhin, Nikolay A. Maleev +10 more

Abstract We demonstrate the possibility of using vertical-cavity surface emitting lasers with intracavity contacts and a rhomboidal oxide current aperture for creating compact optically pumped 133Cs atomic magnetometer operating in non-zero magnetic fields, which is promising to use in magnetoencephalographic systems. The magnetic resonance parameters were studied in a two-beam MX optically pumped magnetic field sensor scheme based on the effect of the magnetic resonance line narrowing at high laser pumping and high concentrations of alkali-metal atoms. The ultimate sensitivity of OPM was estimated by the ratio of the magnetic resonance steepness to the probe light shot noise level. It is shown that a magnetic sensor based on the vertical-cavity surface emitting lasers and a compact (0.125 cm3 volume) Cs vapor cell can achieve a shot noise-limited sensitivity better than 15 fT in 1 Hz bandwidth. Developed intracavity contacted VCSELs suitable for use in compact atomic magnetometers for magnetoencephalographic systems.

Journal of Physics Conference Series•Dec 1, 2020•6 citations

Magnetoencephalography with a two-color pump-probe, fiber-coupled atomic magnetometer

Cort Johnson, Peter D. D. Schwindt, Michael P. Weisend

The authors have detected magnetic fields from the human brain with a compact, fiber-coupled rubidium spin-exchange-relaxation-free magnetometer. Optical pumping is performed on the D1 transition and Faraday rotation is measured on the D2 transition. The beams share an optical axis, with dichroic optics preparing beam polarizations appropriately. A sensitivity of <5 fT/Hz is achieved. Evoked responses resulting from median nerve and auditory stimulation were recorded with the atomic magnetometer. Recordings were validated by comparison with those taken by a commercial magnetoencephalography system. The design is amenable to arraying sensors around the head, providing a framework for noncryogenic, whole-head magnetoencephalography.

Applied Physics Letters•Dec 13, 2010•213 citations

A modular optically pumped magnetometer system

Thomas Coussens, Aikaterini Gialopsou, C. Abel +9 more

Abstract To address the demands in healthcare and industrial settings for spatially resolved magnetic imaging, we present a modular optically pumped magnetometer (OPM) system comprising a multi-sensor array of highly sensitive quantum magnetometers. This system is designed and built to facilitate fast prototyping and testing of new measurement schemes by enabling quick reconfiguration of the self-contained laser and sensor modules as well as allowing for the construction of various array layouts with a shared light source. The modularity of this system facilitates the development of methods for managing high-density arrays for magnetic imaging. The magnetometer sensitivity and bandwidth are first characterised in both individual channel and differential gradiometer configurations before testing in a real-world magnetoencephalography environment by measuring alpha rhythms from the brain of a human participant. We demonstrate the OPM system in a first-order axial gradiometer configuration with a magnetic field gradient sensitivity of 10 fT / cm / Hz at a baseline of 4.5 cm. Single-channel operation achieved a sensitivity of 65 fT / Hz . Bandwidths exceeding 200 Hz were achieved for two independent modules. The system’s increased temporal resolution allows for the measurement of spinal cord signals, which we demonstrate by using phantom signal trials and comparing with an existing commercial sensor.

arXiv:2106.05877•Quantum Science and Technology•May 21, 2024•13 citations

Miniaturized atomic magnetometer employing pump-beam modulation

Ziqian Yue, Junjian Tang, Yueyang Zhai

Atomic magnetometers with magnetic field modulation have emerged as pivotal sensors for biomagnetic measurements. However, challenges such as crosstalk between adjacent modulation fields and the complexity of signal circuitry become increasingly serious in the development of array-based equipment with the magnetometers, especially in the miniaturized atomic magnetometer. This paper proposes an all-optical, fully integrated fiber-coupled atomic magnetometer that employs pump beam modulation. A miniaturized atomic magnetometer with a volume of 10 cm³ has been designed, incorporating an active measurement area of 3 × 3 × 3 mm3. The magnetometer operates in the spinexchange relaxation-free (SERF) regime with an 87Rb vapor cell. An amplitude-modulated pump beam directly modulates the rubidium atomic ensembles, rather than the magnetic field modulation. The optical rotation angle is detected by an unmodulated probe beam oriented orthogonally to the pump beam. In light of the discontinuous pumping characteristics, we developed a modified theoretical model to clarify the output response of the proposed atomic magnetometer. The experimental results demonstrate that the new design achieves the same sensitivity as conventional magnetometer configurations with a smaller volume and without crosstalk of the magnetic field. This research highlights the significant potential for advancing the development of highly sensitive, miniaturized atomic magnetometers, making them particularly suitable for applications in magnetocardiography (MCG) and magnetoencephalography (MEG).

Advanced Sensor Systems and Applications•Oct 11, 2024•0 citations

Towards a multi-channel zero-field optically pumped magnetometer for imaging

Ronja Rasser, Peter A. Koss, Svenja A. Knappe +1 more

Abstract We present a design for a multi-channel optically pumped zero-field magnetometer utilizing a 200-μm-thick Rubidium vapor cell. The vapor cell and its housing are designed to reduce the minimal distance between a magnetic sample and the sensing volume to about 1 mm, to optimize the effective spatial resolution. The thin vapor cell, filled with 2 atm of nitrogen as a buffer gas reduces the volume across which the magnetic field is averaged. The vapor cell is fully illuminated by a single laser beam, and the transmitted light is imaged onto a 4 x 4 photodiode array, allowing for simultaneous measurement of a magnetic field distribution with up to 16 channels. The performance of the magnetometer is studied for all channels. It is shown that the sensor can operate in the spin-exchange relaxation-free regime with a projected photon-shot noise limited noise floor of about 1 pT/Hz1/2 for a sensitive voxel size of approximately 600 μm x 600 μm x 200 μm.

EPJ Quantum Technology•Aug 13, 2025•7 citations
Magnetic sensors developed for application in magnetoencephalography must meet a number of requirements; the main ones are compactness, sensitivity and response speed. We present a quantum optically pumped atomic sensor with cell volume of 0.5 cm3that meets these requirements and is operable in nonzero magnetic fields. The ultimate sensitivity of the sensor was estimated as (using the criteria of the ratio of the slope of the magnetic resonance signal to the shot noise spectral density) to be better than 5 fT/ √Hz. The actual sensitivity, measured in a gradiometric scheme, reaches 13 fT/ √Hz per sensor. We also present a novel and fast algorithm for optimization of the geometric properties of non-zero field sensor array with respect to maximization of the information transfer rate for cortical sources.
Abstract A variant of the scheme of a magnetometric sensor based on cesium atomic vapor is proposed and experimentally investigated. The sensor uses magnetic resonance excitation by modulated light of an hyperfine optical pumping that is transverse to the magnetic field. It is shown that, for a cell with a volume of 0.125 cm3, the variational sensitivity of such a scheme, estimated from the ratio of the steepness of the signal in the center of the magnetic resonance to the shot noise of the detecting radiation, reaches a level of <10 fT/Hz1/2 in the frequency band width of the order of 850 Hz. The sensor, which does not emit radio frequency fields, is designed for use in magnetoencephalographic complexes. Possible ways to improve the performance of the scheme for detecting relatively fast (~4.2 kHz in a field of 0.1 mT) signals of the precession of proton magnetic moments in promising ultra-weak field tomography schemes are considered.
Abstract We demonstrate the possibility of using vertical-cavity surface emitting lasers with intracavity contacts and a rhomboidal oxide current aperture for creating compact optically pumped 133Cs atomic magnetometer operating in non-zero magnetic fields, which is promising to use in magnetoencephalographic systems. The magnetic resonance parameters were studied in a two-beam MX optically pumped magnetic field sensor scheme based on the effect of the magnetic resonance line narrowing at high laser pumping and high concentrations of alkali-metal atoms. The ultimate sensitivity of OPM was estimated by the ratio of the magnetic resonance steepness to the probe light shot noise level. It is shown that a magnetic sensor based on the vertical-cavity surface emitting lasers and a compact (0.125 cm3 volume) Cs vapor cell can achieve a shot noise-limited sensitivity better than 15 fT in 1 Hz bandwidth. Developed intracavity contacted VCSELs suitable for use in compact atomic magnetometers for magnetoencephalographic systems.
The authors have detected magnetic fields from the human brain with a compact, fiber-coupled rubidium spin-exchange-relaxation-free magnetometer. Optical pumping is performed on the D1 transition and Faraday rotation is measured on the D2 transition. The beams share an optical axis, with dichroic optics preparing beam polarizations appropriately. A sensitivity of <5 fT/Hz is achieved. Evoked responses resulting from median nerve and auditory stimulation were recorded with the atomic magnetometer. Recordings were validated by comparison with those taken by a commercial magnetoencephalography system. The design is amenable to arraying sensors around the head, providing a framework for noncryogenic, whole-head magnetoencephalography.
Abstract To address the demands in healthcare and industrial settings for spatially resolved magnetic imaging, we present a modular optically pumped magnetometer (OPM) system comprising a multi-sensor array of highly sensitive quantum magnetometers. This system is designed and built to facilitate fast prototyping and testing of new measurement schemes by enabling quick reconfiguration of the self-contained laser and sensor modules as well as allowing for the construction of various array layouts with a shared light source. The modularity of this system facilitates the development of methods for managing high-density arrays for magnetic imaging. The magnetometer sensitivity and bandwidth are first characterised in both individual channel and differential gradiometer configurations before testing in a real-world magnetoencephalography environment by measuring alpha rhythms from the brain of a human participant. We demonstrate the OPM system in a first-order axial gradiometer configuration with a magnetic field gradient sensitivity of 10 fT / cm / Hz at a baseline of 4.5 cm. Single-channel operation achieved a sensitivity of 65 fT / Hz . Bandwidths exceeding 200 Hz were achieved for two independent modules. The system’s increased temporal resolution allows for the measurement of spinal cord signals, which we demonstrate by using phantom signal trials and comparing with an existing commercial sensor.
Atomic magnetometers with magnetic field modulation have emerged as pivotal sensors for biomagnetic measurements. However, challenges such as crosstalk between adjacent modulation fields and the complexity of signal circuitry become increasingly serious in the development of array-based equipment with the magnetometers, especially in the miniaturized atomic magnetometer. This paper proposes an all-optical, fully integrated fiber-coupled atomic magnetometer that employs pump beam modulation. A miniaturized atomic magnetometer with a volume of 10 cm³ has been designed, incorporating an active measurement area of 3 × 3 × 3 mm3. The magnetometer operates in the spinexchange relaxation-free (SERF) regime with an 87Rb vapor cell. An amplitude-modulated pump beam directly modulates the rubidium atomic ensembles, rather than the magnetic field modulation. The optical rotation angle is detected by an unmodulated probe beam oriented orthogonally to the pump beam. In light of the discontinuous pumping characteristics, we developed a modified theoretical model to clarify the output response of the proposed atomic magnetometer. The experimental results demonstrate that the new design achieves the same sensitivity as conventional magnetometer configurations with a smaller volume and without crosstalk of the magnetic field. This research highlights the significant potential for advancing the development of highly sensitive, miniaturized atomic magnetometers, making them particularly suitable for applications in magnetocardiography (MCG) and magnetoencephalography (MEG).
Abstract We present a design for a multi-channel optically pumped zero-field magnetometer utilizing a 200-μm-thick Rubidium vapor cell. The vapor cell and its housing are designed to reduce the minimal distance between a magnetic sample and the sensing volume to about 1 mm, to optimize the effective spatial resolution. The thin vapor cell, filled with 2 atm of nitrogen as a buffer gas reduces the volume across which the magnetic field is averaged. The vapor cell is fully illuminated by a single laser beam, and the transmitted light is imaged onto a 4 x 4 photodiode array, allowing for simultaneous measurement of a magnetic field distribution with up to 16 channels. The performance of the magnetometer is studied for all channels. It is shown that the sensor can operate in the spin-exchange relaxation-free regime with a projected photon-shot noise limited noise floor of about 1 pT/Hz1/2 for a sensitive voxel size of approximately 600 μm x 600 μm x 200 μm.