Журнал технической физики

Методы возбуждения параметрического резонанса в схеме оптического магнитометрического датчика

Anton K. Vershovskii, М.В. Петренко

Published Jan 1, 2021• 3 citations
Publisher

Abstract

An experimental comparison of the methods for modulating the parameters of resonant transverse to the external magnetic field pumping radiation in a two-beam optical magnetometric sensor (so called Bell-Bloom scheme) is carried out, as well as a comparison of these methods with the standard method of radio-frequency excitation of magnetic resonance under conditions of strong laser pumping. It is shown that although the standard method allows one to achieve a greater suppression of the spin-exchange broadening of the magnetic resonance line by pumping light, the Bell-Bloom scheme has advantages that make it possible to obtain similar sensitivity values upon modulation of both the intensity and polarization of the pump light; at the same time, the Bell-Bloom scheme is potentially characterized by higher speed, which is essential for the problems of magnetoencephalography and ultra-low field magnetic resonance imaging.

DOI
10.21883/jtf.2021.05.50698.337-20

Related papers

Ultimate parameters of an all-optical MX resonance in Cs in ultraweak magnetic field

Mikhail V. Petrenko, Anatoly S. Pazgalev, Anton K. Vershovskii

We present the results of studying the parameters of the magnetic ${M}_{X}$ resonance in an all-optical sensor built according to the two-beam Bell-Bloom scheme in nonzero ultraweak magnetic fields in which the effects of spin-exchange broadening suppression are partially manifested. We report on the features of the resonance under these conditions. We also optimize the resonance parameters to achieve maximum sensitivity in magnetoencephalographic sensors. We demonstrate an improvement in the ultimate achievable sensitivity of an all-optical ${M}_{X}$ sensor by a factor of 4 or more, which in our experiment corresponds to a decrease from 13 to $3\phantom{\rule{0.2em}{0ex}}{\text{fT/Hz}}^{1/2}$ in a volume of $0.13\phantom{\rule{0.2em}{0ex}}{\mathrm{cm}}^{3}$. We also report the effect of incomplete suppression of spin-exchange broadening under conditions of strong transverse modulated optical pumping, and propose a semiempirical model to describe it.

arXiv:2306.16924•Physical Review Applied•Sep 29, 2023•1 citation

Трехуровневое приближение при расчете параметров оптически детектируемого магнитного резонанса в условиях сильной лазерной накачки

Anton K. Vershovskii, Mikhail V. Petrenko

An algorithm is proposed for the approximate solution of a purely nonlinear problem of the parameters of optically detected magnetic resonance in the ground state of alkali atoms in an optically dense medium under conditions of intense narrow-band optical pumping, which causes, first, bleaching of the atomic medium, and, second, partial suppression of spin-exchange broadening. The direct solution of the Liouville equation under these conditions is complicated by the fact that the relaxation time of each of the levels of the hyperfine and Zeeman structure of the ground state is determined by the populations of the remaining levels, which leads to the necessity of solving a self-consistent problem in a multilevel system, and, as a rule, requires the use of supercomputers. In this paper, approximations are proposed that make it possible to significantly simplify and speed up the calculation by orders of magnitude, and the results are compared with experimental data using the example of a two-beam Mx scheme of a magnetometric sensor

Оптика и спектроскопия•Jan 1, 2021•2 citations

Frequency transfer of an optically detected magnetic resonance and observation of the Hanle effect in a nonzero magnetic field

Anton K. Vershovskii, Mikhail V. Petrenko

The method of transferring the frequency of an optically detected magnetic resonance both up and down by an arbitrary value is implemented in a single-beam optical pumping scheme by modulating the linearly polarized beam component. The possibility of observing the Hanle resonance in a magnetic field virtually zeroed upon transition to a rotating coordinate system is demonstrated. A model experiment was carried out, confirming the fundamental feasibility and effectiveness of the method. Keywords: Optically detectable magnetic resonance, optical resonance frequency transfer, Hanle effect, Bell-Bloom scheme, quantum magnetometer.

Оптика и спектроскопия•Jan 1, 2023•1 citation

All-Optical Magnetometric Sensor for Magnetoencephalography and Ultralow Field Tomography

Anton K. Vershovskii, Anatoly S. Pazgalev, Mikhail V. Petrenko

Abstract A variant of the scheme of a magnetometric sensor based on cesium atomic vapor is proposed and experimentally investigated. The sensor uses magnetic resonance excitation by modulated light of an hyperfine optical pumping that is transverse to the magnetic field. It is shown that, for a cell with a volume of 0.125 cm3, the variational sensitivity of such a scheme, estimated from the ratio of the steepness of the signal in the center of the magnetic resonance to the shot noise of the detecting radiation, reaches a level of <10 fT/Hz1/2 in the frequency band width of the order of 850 Hz. The sensor, which does not emit radio frequency fields, is designed for use in magnetoencephalographic complexes. Possible ways to improve the performance of the scheme for detecting relatively fast (~4.2 kHz in a field of 0.1 mT) signals of the precession of proton magnetic moments in promising ultra-weak field tomography schemes are considered.

Technical Physics Letters•Sep 1, 2020•12 citations

Realization and Optimization of all-optical Cs atom magnetometer based on Bell-Bloom structure

Zhang Jun-ha

In order to availably solve the crosstalk of the traditionally optically pumped magnetometer,in which an oscillating magnetic field drives transitions between Zeeman magnetic sublevels,an all-optical cesium(Cs)atomic magnetometer based on Bell-Bloom structure has been investigated and demonstrated experimentally in this paper.When there is an external magnetic field,by analyzing the Larmor precession spectrum of the polarized ground state atoms induced by a modulated pump laser,the atomic magnetometer with high sensitivity is achieved.First,we present the operating principle of the magnetometer with amplitude modulation and set up the experimental system by means of a three-layer magnetic shield.Then in the case of 100 nT field by optimizing the experimental parameters,i.e.,the frequency and intensity of the pump laser,we have obtained the atomic precession spectrum with line width(full width at half maximum)of 22 Hz and signal noise ratio(SNR)of 3.2×104 per unit hertz band width,and then have evaluated the magnetometer with an intrinsic sensitivity of 100fT/Hz1/2 and spatial resolution smaller than 1.2cm,while the experimentally closed loop test shows that the cesium atomic magnetometer would have the sensitivity of 120fT/Hz1/2 and response time of 300 ms.

Journal of Optoelectronics·laser•Jan 1, 2015•0 citations

Three-Level Approximation upon Calculation of Parameters of Optically Detected Magnetic Resonance under the Conditions of Strong Laser Pumping

Anton K. Vershovskii, Mikhail V. Petrenko

Abstract An algorithm of approximate solution of an essentially nonlinear problem of parameters of optically detected magnetic resonance in the ground state of alkali atoms in an optically dense medium under the conditions of strong narrow-band optical pumping that induces transparency of atomic medium and partial suppression of spin-exchange broadening is proposed. Straightforward solution of the Liouville’s equation is complicated in this case by the fact that relaxation time of each of the levels of the hyperfine and Zeeman structures of the ground state is determined by populations of all other levels, which results in the necessity of solving a self-consistent problem in a multilevel system and requires using supercomputers, as a rule. In the present work, approximations that allow substantially simplifying and accelerating calculation are proposed. A two-beam MX-scheme of a magnetometric sensor is used as an example to compare the results of calculations with experimental data.

Optics and Spectroscopy•May 1, 2021•3 citations
An experimental comparison of the methods for modulating the parameters of resonant transverse to the external magnetic field pumping radiation in a two-beam optical magnetometric sensor (so called Bell-Bloom scheme) is carried out, as well as a comparison of these methods with the standard method of radio-frequency excitation of magnetic resonance under conditions of strong laser pumping. It is shown that although the standard method allows one to achieve a greater suppression of the spin-exchange broadening of the magnetic resonance line by pumping light, the Bell-Bloom scheme has advantages that make it possible to obtain similar sensitivity values upon modulation of both the intensity and polarization of the pump light; at the same time, the Bell-Bloom scheme is potentially characterized by higher speed, which is essential for the problems of magnetoencephalography and ultra-low field magnetic resonance imaging.
We present the results of studying the parameters of the magnetic MX{M}_{X}MX​ resonance in an all-optical sensor built according to the two-beam Bell-Bloom scheme in nonzero ultraweak magnetic fields in which the effects of spin-exchange broadening suppression are partially manifested. We report on the features of the resonance under these conditions. We also optimize the resonance parameters to achieve maximum sensitivity in magnetoencephalographic sensors. We demonstrate an improvement in the ultimate achievable sensitivity of an all-optical MX{M}_{X}MX​ sensor by a factor of 4 or more, which in our experiment corresponds to a decrease from 13 to 3fT/Hz1/23\phantom{\rule{0.2em}{0ex}}{\text{fT/Hz}}^{1/2}3fT/Hz1/2 in a volume of 0.13cm30.13\phantom{\rule{0.2em}{0ex}}{\mathrm{cm}}^{3}0.13cm3. We also report the effect of incomplete suppression of spin-exchange broadening under conditions of strong transverse modulated optical pumping, and propose a semiempirical model to describe it.
An algorithm is proposed for the approximate solution of a purely nonlinear problem of the parameters of optically detected magnetic resonance in the ground state of alkali atoms in an optically dense medium under conditions of intense narrow-band optical pumping, which causes, first, bleaching of the atomic medium, and, second, partial suppression of spin-exchange broadening. The direct solution of the Liouville equation under these conditions is complicated by the fact that the relaxation time of each of the levels of the hyperfine and Zeeman structure of the ground state is determined by the populations of the remaining levels, which leads to the necessity of solving a self-consistent problem in a multilevel system, and, as a rule, requires the use of supercomputers. In this paper, approximations are proposed that make it possible to significantly simplify and speed up the calculation by orders of magnitude, and the results are compared with experimental data using the example of a two-beam Mx scheme of a magnetometric sensor
The method of transferring the frequency of an optically detected magnetic resonance both up and down by an arbitrary value is implemented in a single-beam optical pumping scheme by modulating the linearly polarized beam component. The possibility of observing the Hanle resonance in a magnetic field virtually zeroed upon transition to a rotating coordinate system is demonstrated. A model experiment was carried out, confirming the fundamental feasibility and effectiveness of the method. Keywords: Optically detectable magnetic resonance, optical resonance frequency transfer, Hanle effect, Bell-Bloom scheme, quantum magnetometer.
Abstract A variant of the scheme of a magnetometric sensor based on cesium atomic vapor is proposed and experimentally investigated. The sensor uses magnetic resonance excitation by modulated light of an hyperfine optical pumping that is transverse to the magnetic field. It is shown that, for a cell with a volume of 0.125 cm3, the variational sensitivity of such a scheme, estimated from the ratio of the steepness of the signal in the center of the magnetic resonance to the shot noise of the detecting radiation, reaches a level of <10 fT/Hz1/2 in the frequency band width of the order of 850 Hz. The sensor, which does not emit radio frequency fields, is designed for use in magnetoencephalographic complexes. Possible ways to improve the performance of the scheme for detecting relatively fast (~4.2 kHz in a field of 0.1 mT) signals of the precession of proton magnetic moments in promising ultra-weak field tomography schemes are considered.
In order to availably solve the crosstalk of the traditionally optically pumped magnetometer,in which an oscillating magnetic field drives transitions between Zeeman magnetic sublevels,an all-optical cesium(Cs)atomic magnetometer based on Bell-Bloom structure has been investigated and demonstrated experimentally in this paper.When there is an external magnetic field,by analyzing the Larmor precession spectrum of the polarized ground state atoms induced by a modulated pump laser,the atomic magnetometer with high sensitivity is achieved.First,we present the operating principle of the magnetometer with amplitude modulation and set up the experimental system by means of a three-layer magnetic shield.Then in the case of 100 nT field by optimizing the experimental parameters,i.e.,the frequency and intensity of the pump laser,we have obtained the atomic precession spectrum with line width(full width at half maximum)of 22 Hz and signal noise ratio(SNR)of 3.2×104 per unit hertz band width,and then have evaluated the magnetometer with an intrinsic sensitivity of 100fT/Hz1/2 and spatial resolution smaller than 1.2cm,while the experimentally closed loop test shows that the cesium atomic magnetometer would have the sensitivity of 120fT/Hz1/2 and response time of 300 ms.
Abstract An algorithm of approximate solution of an essentially nonlinear problem of parameters of optically detected magnetic resonance in the ground state of alkali atoms in an optically dense medium under the conditions of strong narrow-band optical pumping that induces transparency of atomic medium and partial suppression of spin-exchange broadening is proposed. Straightforward solution of the Liouville’s equation is complicated in this case by the fact that relaxation time of each of the levels of the hyperfine and Zeeman structures of the ground state is determined by populations of all other levels, which results in the necessity of solving a self-consistent problem in a multilevel system and requires using supercomputers, as a rule. In the present work, approximations that allow substantially simplifying and accelerating calculation are proposed. A two-beam MX-scheme of a magnetometric sensor is used as an example to compare the results of calculations with experimental data.