We report high-precision frequency ratio measurements between optical atomic clocks based on
27Al
+,
171Yb, and
87Sr. With total fractional uncertainties at or below
3.2×10−18, these measurements meet an important milestone criterion for redefinition of the second in the International System of Units. Discrepancies in
87Sr ratios at approximately
1×10−16 and the Al
+/Yb ratio at
1.6×10−17 in fractional units compared to our previous measurements underscore the importance of repeated, high-precision comparisons by different laboratories. A key innovation in this work is the use of a common ultrastable reference delivered to all clocks via a 3.6 km phase-stabilized fiber link between two institutions. Derived from a cryogenic single-crystal silicon cavity, this reference improves comparison stability by a factor of 2 to 3 over previous systems, with an optical lattice clock ratio achieving a fractional instability of
1.3×10−16 at 1 second. By enabling faster comparisons, this stability will improve sensitivity to non-white noise processes and other underlying limits of state-of-the-art optical frequency standards.
We report a measurement of the optical frequency ratio between the
2S1/2(F=0)--
2F7/2(F=3) electric-octupole (E3) transition of
171Yb
+ and the
1S0--
3P0 transition of
87Sr,
νYb+/νSr=1.4959916185449005881(65). Reaching a fractional uncertainty of
4.3×10−18, this result improves upon the previous best by more than a factor of three and is among the few that meet the requirements for interspecies clock comparisons specified by the roadmap towards the redefinition of the SI second. The comparison is between a transportable optical lattice clock and a stationary single-ion clock. It spans a period of nearly two years, during which the transportable clock was intermittently operated off-campus. The ratio was reproducibly measured during four separate campaigns, which are consistent within their statistical uncertainties. The results demonstrate reproducible
10−18 level operation of the transportable clock and thus validate its application for chronometric geodesy and as a transfer standard for inter-institute clock comparisons, e.g., in the absence of optical fiber links.