Abstract In closed-loop dual-atom interferometers, the initial phase shift is a non-inertial phase induced by Raman pulses. This initial optical phase affects the measurements of both acceleration and rotation rate. In this study, we present a method to modulate the initial optical phase of light pulse atom interferometers (LPAIs). Employing this modulation method, we investigated the influence of the initial optical phase on a closed-loop dual-atom interferometer. The results reveal a better short-term sensitivity in the rotation rate measurement at an initial optical phase of 0. Our closed-loop dual-atom interferometer allows for decoupled measurements of rotation rate and acceleration. With a short interrogation time of 2T = 1.73 ms, the short-term sensitivities of the rotation rate and acceleration measurements are 2.73 × 10 − 5 ( rad / s ) / Hz and 272 μ g / Hz , respectively.
We demonstrate a closed-loop light-pulse atom-interferometer inertial sensor that can realize continuous decoupled measurements of acceleration and rotation rate. The sensor operates with double-loop atom interferometers, which share the same Raman light pulses in a spatially separated Mach-Zehnder configuration and use continuous cold atomic beams propagating in opposite directions from two
2D+ magneto-optical trappings. Acceleration and the rotation rate are decoupled and simultaneously measured by the sum and difference of dual-atom-interferometer signals, respectively. The sensitivities of inertial measurements are also increased to be approximately 1.86 times higher than that of a single atom interferometer. The acceleration phase shift is compensated in real time by phase locking these interferometers via the Raman laser phases from the sum-interferometer signal, and the gyroscope performance is improved. We achieve long-term stabilities of $6.1\phantom{\rule{0.2em}{0ex}}\text{\ensuremath{\mu}}\text{g}$ and 840 nrad/s for the acceleration and the rotation rate, respectively, using a short interrogation time of 0.87 ms (interference area
A=0.097mm2). This work provides a building block for an atomic interferometer-based inertial measurement unit for use in field applications that require a high data rate and high stability.