High-precision gyroscopes in space are essential for fundamental physics research and navigation. Due to its potential high precision, the cold atom gyroscope is expected to be one of the next generation of gyroscopes in space. Here, we report the first realization of a cold atom gyroscope, which was demonstrated by the atom interferometer installed in the China Space Station (CSS) as a payload. By compensating for the CSS's high dynamic rotation rate using a built-in piezoelectric mirror, spatial interference fringes in the interferometer are successfully obtained. Then, the optimized ratio of the Raman laser's angles is derived, the coefficients of the piezoelectric mirror are self-calibrated in orbit, and various systemic effects are corrected. We achieve a rotation measurement resolution of 50 μrad/s for a single shot and 17 μrad/s for an average number of 32. The measured rotation is -1142 ± 29 μrad/s and is compatible with that recorded by the classical gyroscope of the CSS. This study paves the way for developing high-precision cold atom gyroscopes in space.
The cold atom interference gyroscope (CAIG) offer substantial potential for rotation measurement due to the high sensitivity and stability. The CAIG with a fountain configuration realized by four-pulse could provide a larger interference-loop area, and enhanced performance. However, this kind of CAIG is usually much larger and higher since it requires a vacuum chamber with sufficient height to achieve a long Raman pulse interval. We demonstrate a four-pulse CAIG based on a miniaturized vacuum chamber that enables portable and transportable rotation rate measurements. The main vacuum chamber is realized by whole glass material. The height of the vacuum unit is 0.7 m, the volume is
8.86×104 cm3, and the mass is 75 kg. Then, we estimated the performance of our portable CAIG in the environment of the underground laboratory and the fifth-floor office building. In the laboratory, the sensitivities of the homemade CAIG is
4.44×10−6rad/s
/(Hz)1/2with an interrogation time of 55 ms and an interference area of 25 mm2. In addition, we measured the angular velocity of the Earth, the relative error is 2.4%. Furthermore, we transported the CAIG to the fifth floor of the office building, which is near the subway. Ultimately, the CAIG achieved a sensitivity of
5.47×10−4rad/s
/(Hz)1/2. Additionally, the theoretical maximum interrogation time achieved by this CAIG is 124 ms corresponding to a sensitivity of
5.62×10−8 /(Hz)1/2. Our new design of the compact CAIG could provide novel insights into the miniaturization of CAIGs, while also pointing out areas for further improvement.