Journal of the Optical Society of America B

Theory of saturated-absorption cavity ring-down: radiocarbon dioxide detection, a case study

Giovanni Giusfredi, Iacopo Galli, Davide Mazzotti, Pablo Cancio, Paolo De Natale

Published Oct 1, 2015• 41 citations
Publisher

Abstract

The theoretical bases of saturated-absorption cavity ring-down (SCAR) spectroscopy are treated and described in detail in different transition broadening regimes. All the approximations that are introduced to achieve a model as simple and accurate as possible are discussed. In the Voigt broadening regime, a calibration factor that relates the measured spectral areas in the saturated and linear absorption regimes is retrieved. A relevant application of this technique to the measurement of radiocarbon dioxide concentration at both enriched and natural abundance (∼10−12) is presented. The population relaxation rate of one ro-vibrational state of the ν3 band is measured for this highly diluted CO2 isotopologue.

DOI
10.1364/josab.32.002223

Related papers

Saturation cavity ring-down spectrometry using a dynamical relaxation model

Lei Yang, H. Lin, X. J. Feng +2 more

We propose a new simple approximate solution to the two-state rate equation model for analyzing decay signals of saturation cavity ring-down spectrometry in the adiabatic and low-saturation regime. It helps obtain baseline-immune Doppler-free spectra for hyperfine transitions and linear absorption coefficients of a gas in the saturation regime. To demonstrate it, a baseline-immune Lamb dip spectrum of the R1A2 transitions in the 2v2 + v3 band of methane was recorded. The line position was determined to be 6 076.108 457 7(11) cm-1, the relative uncertainty being 1.8 × 10-10.

Optics Express•Jan 22, 2019•7 citations

Theoretical detection limit of saturated absorption cavity ring-down spectroscopy (SCAR) and two-photon absorption cavity ring-down spectroscopy

Kevin K. Lehmann

Applied Physics B•Oct 10, 2013•25 citations

Theoretical description of Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectroscopy under optically saturated conditions

Weiguang Ma, Aleksandra Foltynowicz, Ove Axner

A theoretical description of Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectroscopy (NICE-OHMS) under optically saturated conditions is presented. Expressions for the strength and shape of the Doppler-broadened NICE-OHMS signals are given for both the absorption and the dispersion phase, in the Voigt regime as well as in the Doppler limit. It is shown that Doppler-broadened NICE-OHMS is affected less by optical saturation than other cavity-enhanced techniques; in the Doppler limit the absorption signal decreases by a factor of (1+G±1)−1/2, where G±1 is the degree of saturation for one of the frequency modulation sidebands, whereas the dispersion signal is virtually unaffected by optical saturation. In the Voigt regime both signals show additional dependence on optical saturation. The concept of saturation-insensitive detection is introduced and its conditions are identified.

Journal of the Optical Society of America B•Jun 24, 2008•57 citations

Theoretical model analysis of the dynamic saturation in cavity ringdown spectroscopy

Jae Yong Lee, Jae Won HAHN

Transient saturation in cavity ringdown spectroscopy is theoretically modeled with coupled rate equations accounting for the dynamics of intracavity photons and absorber population, which permits a recipe for retrieving original spectra in the saturation regime.

Laser Applications to Chemical, Security and Environmental Analysis•Jan 1, 2002•0 citations

Optical feedback frequency stabilized cavity ring-down spectroscopy

Johannes Burkart, D. Romanini, S. Kassi

We introduce optical feedback frequency stabilized cavity ring-down spectroscopy (OFFS-CRDS), a near-shot-noise-limited technique that combines kilohertz resolution with an absorption detection sensitivity of 5×10(-13) cm(-1) Hz(-1/2). Its distributed feedback laser source is stabilized to a highly stable V-shaped reference cavity by optical feedback and fine-tuned by means of single-sideband modulation. The stability of this narrow laser is transferred to a ring-down (RD) cavity using a new fibered Pound-Drever-Hall (PDH) locking scheme without a dedicated electro-optic phase modulator, yielding several hundred RD events per second. We demonstrate continuous coverage of more than 7 nm with a baseline noise of 5×10(-12) cm(-1) and a dynamic range spanning six decades. With its resonant intracavity light intensity on the order of 1 kW/cm2, the spectrometer was used for observing a Lamb dip in a transition of carbon dioxide involving four vibrational quanta. Saturating such a weak transition at 160 μW input power, OFFS-CRDS paves the way to Doppler-free molecular overtone spectroscopy for precision measurements of hyperfine structures and pressure shifts.

Optics Letters•Aug 4, 2014•66 citations

Cavity Ring-down Spectroscopy with behavior of Hybrid Cavity Structures

Muhammad Junaid Khan, Rida Batool Sheraliyat

Cavity ring-down (CRD) spectroscopy represents a direct absorption technique of sample absorption measurement. Instead of measuring the amount of the absorbed light, this technique determines the rate at which light intensity decays inside an optical cavity. When using a pulsed or continuous-wave light source, CRD spectroscopy offers considerably higher sensitivity as compared with conventional spectroscopy of absorption, making even very weak absorptions easy to detect.

physics.opticsCross-listedarXiv:2512.12241•Dec 16, 2025
The theoretical bases of saturated-absorption cavity ring-down (SCAR) spectroscopy are treated and described in detail in different transition broadening regimes. All the approximations that are introduced to achieve a model as simple and accurate as possible are discussed. In the Voigt broadening regime, a calibration factor that relates the measured spectral areas in the saturated and linear absorption regimes is retrieved. A relevant application of this technique to the measurement of radiocarbon dioxide concentration at both enriched and natural abundance (∼10−12) is presented. The population relaxation rate of one ro-vibrational state of the ν3 band is measured for this highly diluted CO2 isotopologue.
We propose a new simple approximate solution to the two-state rate equation model for analyzing decay signals of saturation cavity ring-down spectrometry in the adiabatic and low-saturation regime. It helps obtain baseline-immune Doppler-free spectra for hyperfine transitions and linear absorption coefficients of a gas in the saturation regime. To demonstrate it, a baseline-immune Lamb dip spectrum of the R1A2 transitions in the 2v2 + v3 band of methane was recorded. The line position was determined to be 6 076.108 457 7(11) cm-1, the relative uncertainty being 1.8 × 10-10.
A theoretical description of Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectroscopy (NICE-OHMS) under optically saturated conditions is presented. Expressions for the strength and shape of the Doppler-broadened NICE-OHMS signals are given for both the absorption and the dispersion phase, in the Voigt regime as well as in the Doppler limit. It is shown that Doppler-broadened NICE-OHMS is affected less by optical saturation than other cavity-enhanced techniques; in the Doppler limit the absorption signal decreases by a factor of (1+G±1)−1/2, where G±1 is the degree of saturation for one of the frequency modulation sidebands, whereas the dispersion signal is virtually unaffected by optical saturation. In the Voigt regime both signals show additional dependence on optical saturation. The concept of saturation-insensitive detection is introduced and its conditions are identified.
We introduce optical feedback frequency stabilized cavity ring-down spectroscopy (OFFS-CRDS), a near-shot-noise-limited technique that combines kilohertz resolution with an absorption detection sensitivity of 5×10(-13) cm(-1) Hz(-1/2). Its distributed feedback laser source is stabilized to a highly stable V-shaped reference cavity by optical feedback and fine-tuned by means of single-sideband modulation. The stability of this narrow laser is transferred to a ring-down (RD) cavity using a new fibered Pound-Drever-Hall (PDH) locking scheme without a dedicated electro-optic phase modulator, yielding several hundred RD events per second. We demonstrate continuous coverage of more than 7 nm with a baseline noise of 5×10(-12) cm(-1) and a dynamic range spanning six decades. With its resonant intracavity light intensity on the order of 1 kW/cm2, the spectrometer was used for observing a Lamb dip in a transition of carbon dioxide involving four vibrational quanta. Saturating such a weak transition at 160 μW input power, OFFS-CRDS paves the way to Doppler-free molecular overtone spectroscopy for precision measurements of hyperfine structures and pressure shifts.