IEEE Transactions on Instrumentation and Measurement
Hg/sup +/ trapped ion standard with the superconducting cavity maser oscillator
John D. Prestage, Robert L. Tjoelker, R.T. Wang, G.J. Dick, Lute Maleki
Abstract
The frequency stability of an atomic standard based on /sup 199/Hg/sup +/ ions confined in a hybrid RF/DC linear trap is described. The 40.5-GHz clock transition has been measured to be 17 mHz wide, representing a quality factor greater than 2*10/sup 12/. A 160-mHz line is used to steer the output of a 5-MHz crystal oscillator to obtain a stability of 2*10/sup -15/ for 24000-s averaging times. In a separate measurement, a 37-mHz line is used to steer the output of the superconducting cavity maser oscillator to reach 1*10/sup -15/ stability at 10000 s.>