Applied Superconductivity

A resistive d.c. SQUID noise thermometer

S. Menkel, D. Drung, C. Aßmann, T. Schurig

Published Jul 1, 1998• 11 citations
Publisher
DOI
10.1016/s0964-1807(98)00109-4

Related papers

Frequency noise in hysteretic resistive SQUIDs

S.L. Thomasson

Physica B Condensed Matter•Apr 7, 2003•0 citations

Improving noise performance of the SQUID in presence of high resistive signal leads

Narayanan Mariyappa, C. Parasakthi, S. Sengottuvel +5 more

AIP conference proceedings•Jan 1, 2012•0 citations

The quantum roulette noise thermometer

John C. Gallop, L. Hao, R.P. Reed

Applied Superconductivity•Jul 1, 1997•9 citations

Influence of the Josephson junction on the impedance and noise of a resistive superconductive quantum interference device

Robert J. Soulen, William E. Fogle, Jack H. Colwell +2 more

The impedance and noise of a resistive superconductive quantum interference device (R-SQUID) have been measured as a function of the dc and rf currents applied to it. The Josephson junction was adjustable so that data were also taken for several values of the junction critical current. The results were compared with the predictions of a resistively shunted junction (RSJ) model which takes into account the influence of the Josephson junction on the impedance and noise. The agreement was found to be quite good and demonstrates that the noise in the circuit is well understood. Use of the R-SQUID as a noise thermometer below 1 K is assessed in terms of corrections due to the RSJ model. It is demonstrated how the dc and rf currents may be adjusted so that the total noise of the R-SQUID is reduced to within 0.1% of the Johnson noise generated by the resistor alone. Under these conditions, the R-SQUID may be used as a noise thermometer to determine thermodynamic temperature to this inaccuracy from 6 to 700 mK.

Journal of Applied Physics•Oct 15, 1993•4 citations

A Compact Very Low Noise DC-SQUID Magnetometer

R. Cantor, Tapani Ryhänen, Heikki Seppä

Springer proceedings in physics•Jan 1, 1992•5 citations

Current noise measured in the dc SQUID

John M. Martinis, John L. Clarke

Journal of Low Temperature Physics•Dec 1, 1986•12 citations
The impedance and noise of a resistive superconductive quantum interference device (R-SQUID) have been measured as a function of the dc and rf currents applied to it. The Josephson junction was adjustable so that data were also taken for several values of the junction critical current. The results were compared with the predictions of a resistively shunted junction (RSJ) model which takes into account the influence of the Josephson junction on the impedance and noise. The agreement was found to be quite good and demonstrates that the noise in the circuit is well understood. Use of the R-SQUID as a noise thermometer below 1 K is assessed in terms of corrections due to the RSJ model. It is demonstrated how the dc and rf currents may be adjusted so that the total noise of the R-SQUID is reduced to within 0.1% of the Johnson noise generated by the resistor alone. Under these conditions, the R-SQUID may be used as a noise thermometer to determine thermodynamic temperature to this inaccuracy from 6 to 700 mK.