We report a noise bandwidth of 7 GHz in the new generation of NbN hot-electron bolometer (HEB) mixers that are being developed for the space observatory Millimetron. The HEB receiver driven by a 2.5-THz local oscillator offered a noise temperature of 600 K in a 50-MHz final detection bandwidth. As the filter center frequency was swept this value remained nearly constant up to the cutoff frequency of the cryogenic amplifier at 7 GHz. We believe that such a low value of the noise temperature is due to reduced radio frequency (RF) loss at the interface between the superconducting film and the gold contacts. We have also performed gain bandwidth measurements at the superconducting transition on HEB mixers with various lengths and found them to be in excellent agreement with the results of the analytical and numerical models developed for the HEB mixer with both diffusion and phonon cooling of hot electrons.
Diffusion-cooled Nb hot-electron bolometer (HEB) mixers have the potential to simultaneously achieve high intermediate frequency (IF) bandwidths and low mixer noise temperatures for operation at THz frequencies (above the superconductive gap energy). We have measured the IF signal bandwidth at 630 GHz of Nb devices with lengths L = 0.3, 0.2, and 0.1 micrometer in a quasioptical mixer configuration employing twin-slot antennas. The 3-dB EF bandwidth increased from 1.2 GHz for the 0.3 gm long device to 9.2 GHz for the 0.1 gm long device. These results demonstrate the expected 1/L squared dependence of the IF bandwidth at submillimeter wave frequencies for the first time, as well as the largest EF bandwidth obtained to date. For the 0.1 gm device, which had the largest bandwidth, the double sideband (DSB) noise temperature of the receiver was 320-470 K at 630 GHz with an absorbed LO power of 35 nW, estimated using the isothermal method. A version of this mixer with the antenna length scaled for operation at 2.5 THz has also been tested. A DSB receiver noise temperature of 1800 plus or minus 100 K was achieved, which is about 1,000 K lower than our previously reported results. These results demonstrate that large EF bandwidth and low-noise operation of a diffusion-cooled HEB mixer is possible at THz frequencies with the same device geometry.
We demonstrate that the performance of NbN lattice cooled hot electron bolometer mixers depends strongly on the interface quality between the bolometer and the contact structure. We show experimentally that both the receiver noise temperature and the gain bandwidth can be improved by more than a factor of 2 by cleaning the interface and adding an additional superconducting interlayer to the contact pad. Using this we obtain a double sideband receiver noise temperature TN,DSB=950 K at 2.5 THz and 4.3 K, uncorrected for losses in the optics. At the same bias point, we obtain an IF gain bandwidth of 6 GHz.
NbN phonon-cooled superconducting hot-electron bolometric mixers with quasi-optical coupling were fabricated in a self-aligned process by means of e-beam-lithography. The mixer element consists of a 5 nm thick NbN film sputtered on a fused quartz substrate. Noise performance was investigated in the frequency range of 798 to 813 GHz for intermediate frequencies from 1.0 to 1.75 GHz. Uncorrected double sideband receiver noise temperatures of about 900 K were measured at an intermediate frequency of 1000 MHz with a 50 MHz bandpass filter. The gain bandwidth was determined by superposing two local oscillator signals and measuring the height of the discrete line at the intermediate frequency output. A 3dB-roll-off-frequency of some 1.3 GHz was obtained.
Sensitivity and gain bandwidth measurements of phonon-cooled NbN superconducting hot-electron bolometer mixers are presented. The best receiver noise temperatures are: 700 K at 1.6 THz and 1100 K at 2.5 THz. Parylene as an antireflection coating on silicon has been investigated and used in the optics of the receiver. The dependence of the mixer gain bandwidth (GBW) on the bias voltage has been measured. Starting from low bias voltages, close to operating conditions yielding the lowest noise temperature, the GBW increases towards higher bias voltages, up to three times the initial value. The highest measured GBW is 9 GHz within the same bias range the noise temperature increases by a factor of two.
In this article, we measure the double sideband (DSB) receiver noise temperature (
TrecDSB)of an NbN hot electron bolometer (HEB) mixer at three local oscillator frequencies of 1.6, 2.5, and 5.3 THz. The HEB has cleaned contact interfaces with a 200-nm-thick Au layer. The measured
TrecDSBvalues are 530 ± 11 K, 640 ± 18 K, and 2190 ±150 K at 1.6, 2.5, and 5.3 THz, respectively, using an air setup with total optical losses of 2.60 ± 0.04, 2.63 ± 0.16, and 4.70 ± 0.24 dB, respectively. We derived low mixer noise temperatures (
TmixerDSB)of 240 ± 6 K at 1.6 THz and 290 ± 13 K at 2.5 THz, achieving over 30% improvement compared to published NbN HEB mixers. This enhancement can reduce the integration time of a heterodyne instrument by roughly a factor of 2. At 5.3 THz,
TmixerDSBis 620 ± 55 K, showing limited improvement due to nonoptimized antenna geometry. These results also contribute to understanding the device physics of a wide HEB (4 μm) at high frequencies. The mixer was developed for the proposed Orbiting Astronomical Satellite for Investigating Stellar Systems and Single Aperture Large Telescope for Universe Studies (concept) missions.