Nuclear-magnetic-resonance gyroscopes that detect rotation as a shift in the precession frequency of nuclear spins have attracted a lot of attention. Under a feedback-generated drive, the precession frequency in the steady state is supposed to be dependent only on the angular momentum and an applied magnetic field. However, nuclei with spins larger than 1/2 experience electric quadrupole interaction with electric-field gradients at the cell walls. This quadrupole interaction shifts the precession frequencies of the nuclear spins, which brings inaccuracy to the rotation measurement as the quadrupole interaction constant
Cq is difficult to precisely measure. In this paper, the effects of quadrupole interaction on nuclear-magnetic-resonance gyroscopes are theoretically studied. We find that, when the constant
Cq is small compared to the characteristic decay rate of the system, which is the case in most gyroscopes for practical use, as the strength of the feedback-driving field increases, the quadrupole shift monotonically decreases, regardless of the system parameters in including
Cq. Furthermore, the larger driving amplitude can reduce the quadrupole shift's fluctuation caused by fluctuations of the system parameters, and, thus, it can reduce systematic errors resulted from the quadrupole interaction. In the large-
Cq regime, more than one precession frequency exists, and the nuclear spins may precess with a single frequency or multifrequencies depending on initial conditions. In this regime with large driving amplitudes, the nuclear spin can restore the single-frequency precession. These results are obtained by solving an effective master equation of the nuclear spins in a rotating frame from which both the steady-state solutions and the dynamics of the system are shown.
Abstract We report theoretical investigations on the transient dynamics of the ground-state Hanle effect in the Voigt geometry, where a time-varying scanning magnetic field B (t) is applied perpendicular to the light propagation direction. Transient Hanle signals are calculated for different light polarizations using a theoretical model based on the density matrix approach. First, we study the ellipticity dependence of transient Hanle signals in two configurations in which the major axis of the light polarization ellipse ( ε a ) is aligned either perpendicular or parallel to the field B . For ε a ⊥ B , elliptical polarization of the light produces oscillations of frequencies 2ΩL(t) and ΩL(t) simultaneously, where ΩL(t) is the Larmor frequency of the corresponding field B (t). On the other hand, in the case of ε a || B , only ΩL(t) oscillation is observed using elliptically polarized light. Second, we vary the tilt angle between the polarization vector of a linearly polarized light and the orientation of the field B and study its effect on the transient response of Hanle signals. Both the amplitude and the decay rate of the oscillations show a strong dependence on the light ellipticity and tilt angle of the polarization vector. In addition, the influence of the magnetic field sweep rate on the transient signals is demonstrated for given values of light ellipticity and polarization angle.
Spin alignment based Rb-87 magnetometry with free spin precession has been proposed. A 795-nm linearly polarized laser beam, serving as the pump beam and the probe beam, propagates through a cylindrical Rb-87 enriched atomic vapor cell along the Z axis, with the polarization orientation aligned along the Y axis. Simultaneously, the static magnetic field B0⃗ is applied along the Y axis, and the π/2 pulse radio-frequency (RF) magnetic field BRF⃗ is applied along the Z axis. Given that the laser polarization direction aligns with the static magnetic field, the pump beam exhibits π polarization. This induces a symmetric population distribution of ground-state atoms across the Zeeman sublevels. This leads to the generation of a magnetic quadrupole moment, thereby facilitating the formation of the spin alignment state. The π/2 pulse RF magnetic field aligns the magnetic quadrupole moment with the direction of the RF magnetic field. Upon the cessation of the RF magnetic field, the magnetic quadrupole moment undergoes precession around the static magnetic field. The linearly polarized probe beam, in conjunction with a balanced polarimeter comprising a true zero-order half-wave plate, a Wollaston prism, and a balanced differential photodiode, is employed to measure the free precession decay signal. In this magnetometry system, both the pump power and the RF magnetic field strength have been optimized. The sensitivity of the spin alignment based Rb-87 magnetometry with free spin precession is about 1.7 pT/Hz1/2.
We have observed coherence among the four magnetic sublevels of freely precessing
21Ne for times of 4.5 h. This has allowed direct detection of the quadrupole component of coherence relaxation and a test of linearity of quantum mechanics. The quadrupole component of relxation is evident in the different relaxation rates for the three lines of the NMR spectrum. Nonlinear corrections to quantum mechanics are found to be less than 1.6\ifmmode\times\else\texttimes\fi{}${10}^{\mathrm{\ensuremath{-}}26}$ of the binding energy per nucleon of
21Ne.
Optically pumped magnetometers are important instruments for imaging biological magnetic signals without the need for cryogenic cooling. These magnetometers are presently available in the commercial market and utilize the principles of atomic alignment or orientation, enabling remarkable sensitivity and precision in the measurement of magnetic fields. This research focuses on utilizing a spin-aligned atomic ensemble for magnetometry at zero field. An approach is introduced which involves evaluating how the linear polarization of light rotates as it passes through the atomic vapor to null the magnetic field. Analytical expressions are derived for the resulting spin alignment and photodetection signals. Experimental results are provided, demonstrating good agreement with the theoretical predictions. The sensitivity and bandwidth of the magnetometer are characterized based on the detected polarization rotation signal. Lastly, the practical utility of the magnetometer for medical applications is demonstrated by successfully detecting a synthetic cardiac signal.