A high-finesse optical cavity was employed to perform highly sensitive spectroscopy of molecular oxygen at wavelengths near 763 nm. An equivalent absorption length of ∼1 km was obtained by a 26-cm-long optical cavity with a finesse of 6000. An extended cavity diode laser was frequency locked to the cavity, and pure absorption profiles were recovered by monitoring of the cavity transmission during continuous scans of the cavity resonance through O2 rotational lines, allowing a detailed investigation of the line shapes. Phase modulation of the laser at a frequency equal to the cavity free-spectral-range frequency was employed for detection of weak absorption signals inside the cavity. A minimum detectable absorption coefficient of 6.9×10-11 cm-1 Hz-1/2 was measured. Finally, a test of the symmetrization postulate in 16O nuclei was demonstrated.
Cavity-enhanced absorption spectroscopy is now widely used as an ultrasensitive technique in observing weak spectroscopic absorptions. Photons inside the cavity are reflected back and forth between the mirrors with reflectivities R close to one and thus (on average) exploit an absorption pathlength L that is 1/(1 − R ) longer than a single pass measurement. As suggested by the Beer-Lambert law, this increase in L results in enhanced absorbance A (given by αL with α being the absorption coefficient) which in turn favours the detection of weak absorptions. At the same time, however, only (1 − R ) of the incident light can enter the cavity [assuming that mirror transmission T is equal to (1 − R )], so that the reduction in transmitted light intensity Δ I caused by molecular absorption equates to that would be obtained if in fact no cavity were present . The enhancement in A = Δ I / I , where I is the total transmitted light intensity, achievable from CEAS therefore comes not from an increase in Δ I , but a sharp decrease in I. In this paper, we calculate the magnitudes of these two terms before and after a cavity is introduced, and aim at interpreting the sensitivity improvement offered by cavity-enhanced absorption spectroscopy from this observable-oriented (i.e. Δ I and I ) perspective. It is first shown that photon energy stored in the cavity is at best as intense as the input light source, implying that any absorbing sample within the cavity is exposed to the same or even lower light intensity after the cavity is formed. As a consequence, the intensity of the light absorbed or scattered by the sample, which corresponds to the Δ I term aforementioned, is never greater than would be the case in a single pass measurement. It is then shown that while this “numerator” term is not improved, the “denominator” term, I , is reduced considerably; therefore, the increase in contrast ratio Δ I / I is solely contributed by the attenuation of transmitted background light I and is ultimately down to the suppression of any measurement noise that is associated with it. The noise component that is most effectively suppressed is the type whose magnitude scales linearly with light intensity I , as is typical of noise caused by environmental instabilities, followed by the shot noise which scales as square root of I . No suppression is achievable for noise sources that are independent of I , a notable example being the thermal noise of a detector or of detection electronics. The usefulness of this “noise suppression” argument is that it links the sensitivity gain offered by a cavity with the property of measurement noise present in the system, and clearly suggests that the achievable sensitivity is dependent on how efficient the various noise components are “suppressed” by the cavity.
Abstract Measurements of ultrahigh-fidelity absorption spectra can help validate quantum theory, engineer ultracold chemistry, and remotely sense atmospheres. Recent achievements in cavity-enhanced spectroscopy using either frequency-based dispersion or time-based absorption approaches have set new records for accuracy with uncertainties at the sub-per-mil level. However, laser scanning or susceptibility to nonlinearities limits their ultimate performance. Here we present cavity buildup dispersion spectroscopy (CBDS), probing the CO molecule as an example, in which the dispersive frequency shift of a cavity resonance is encoded in the cavity’s transient response to a phase-locked non-resonant laser excitation. Beating between optical frequencies during buildup exactly localizes detuning from mode center, and thus enables single-shot dispersion measurements. CBDS can yield an accuracy limited by the chosen frequency standard and measurement duration and is currently 50 times less susceptible to detection nonlinearity compared to intensity-based methods. Moreover, CBDS is significantly faster than previous frequency-based cavity-enhanced methods. The generality of CBDS shows promise for improving fundamental research into a variety of light–matter interactions.
High sensitivity detection of quantum absorption is demonstrated at a sensitivity of
1times1011 integrated absorption, in 1 sec, using 10 mTorr total gas pressure. Achieving highly sensitive detection relies on simple basic principles: 1) The molecule-generated electric field in the forward direction (the "darkness wave") is most precious, and contains the maximum available information. It is the physical manifestation of the -
alpha∗L term appearing when beer#### absorption law is expanded for low opacity, 2) After generating the molecular signal, the detected unabsorbed light sets the "shotnoise" measurement floor. We use an optical transmission cavity for the gas cell, since the high internal "build-up" field will elicit a strong molecular "darkness wave", for a given transmitted de light on the detector, A more conventional explanation of Gavity Enhancement would be the extension of the effective cell length by the factor finesse
∗2/pi.3) To avoid excess low-frequency noise, such small signals should be measured by ac methods, comparing on-resonant and off-resonant cases in quick succession. 4) By SIMULTANEOUSLY obtaining and subtracting these cases, one provides a siginal channel with NO OUTPUT unless there is a resonance. This is conveniently accomplished by using the FM detection method with the optical heterodyne sidebands and carrier being transmitted through the cavity via adjacent axial orders 5). A remarkable property of this configuration, when the modulation frequency= cavity mode spacing, is the suppression of ANY detection of laser frequency noise with the transmitted light. We can refer to this property as (laser FM-) noise-Immune detection. This enables profitable use of cavity finesse in the range above 10,000, without a nois penalty, Altogether, we refer to this new spectroscopy as Noise-Immune, Cavity-Enhanced, Optical Heterodyne Molecular spectroscopy, i.e. "NICE OHMS". We have measured excellent saturated dispersion signals from HCCH in the band near 790 mm (
nu1+3nu3), and recently, using a Nd:YAG source, from the HCCD P(5) line (
nu2+3nu3) at 1064 mm. The saturation peaks broaden from the transit limit of 270 kHz near zero pressure at the rate 34.7 kHz/m Torr. We observe interesting narrow linwidths 8-fold below the transit limit, basically using optical selection of slow molecules by low power and pressure (2 mW and 2 m Torr). The absolute frequency is measurred to be 281, 635,363,960 MHz
pm45 kHz.
We study the response of a high-finesse optical cavity to a cw laser during the laser frequency passage through resonance. For a laser that is spectraly larger than the cavity resonance, laser-field phase fluctuations are converted into amplitude fluctuations, and cavity injection is intrinsically noisy. We develop a model based on Schawlow-Townes spontaneous-emission laser broadening and discuss in detail its effects on high-sensitivity spectroscopic techniques such as cavity-enhanced absorption or cavity ring-down spectroscopy. We present realistic simulations of cavity injection during a sweep through resonance and calculation of statistical quantities such as the average injection efficiency. Agreement with experimental observations is established.
It is experimentally demonstrated that a narrow band continuous wave (cw) light source can be used in combination with a high-finesse optically stable cavity to perform sensitive, high-resolution direct absorption and optical rotation spectroscopy in an amazingly simple experimental setup, using ideas from the field of cavity ring down spectroscopy. Light from a scanning narrow band cw laser is coupled into the cavity via accidental coincidences of the laser frequency with the frequency of one of the multitude of modes of the cavity. The absorption and polarization rotation information is extracted from a measurement of the time-integrated light intensity leaking out of the cavity as a function of laser wavelength.