arXiv:2611.00561
Ultra-broadband quantum infrared spectroscopy extending to 20 {\textmu}m
Yu Mukai, Ryo Okamoto, Shigeki Takeuchi
Abstract
Quantum infrared spectroscopy (QIRS), which exploits quantum interference between photon-pair generation processes, enables infrared spectroscopy using only visible and near-infrared (VNIR) light sources and detectors. Extending the spectral coverage of QIRS into the long-wavelength infrared region up to 20 {\textmu}m is crucial for the identification and characterization of a broad range of materials. However, implementing QIRS beyond 10 {\textmu}m has been challenging because the nonlinear materials used to date do not provide suitable photon-pair sources in this spectral region. Here, we design a wavelength-tunable photon-pair source based on a gallium selenide crystal and develop a QIRS system covering a broad spectral range from 8.6 to 20 {\textmu}m. Using this system, we measure the transmittance spectrum of a polystyrene film and observe characteristic absorption bands beyond 10 {\textmu}m relevant to identifying polystyrene and to probing local chain conformation. These results demonstrate that QIRS is capable of covering the entire molecular fingerprint region.
AI Summary
Quantum infrared spectroscopy measures a sample’s infrared response through interference between photon-pair generation processes while detecting only visible/near-infrared photons. This paper addresses extending demonstrated operation beyond approximately $10.5\,\mu\mathrm{m}$ into the long-wavelength molecular-fingerprint region, where commonly used nonlinear crystals have insufficient transparency. A $100\,\mu\mathrm{m}$-thick GaSe crystal, pumped continuously at $785\,\mathrm{nm}$, provides angle-tunable Type-II spontaneous parametric down-conversion. Three incidence-angle settings cover $8.6$–$20\,\mu\mathrm{m}$ ($1164$–$500\,\mathrm{cm}^{-1}$). Measured near-infrared signal spectra and detection-efficiency-corrected count rates agree with phase-matching and brightness calculations to within approximately 25% for the rates. A double-pass Michelson-type nonlinear interferometer produces signal-photon interference sensitive to the infrared idler path, establishing coherent operation out to $20\,\mu\mathrm{m}$ without direct idler detection. Fourier-transform measurements of a $30\,\mu\mathrm{m}$ polystyrene film resolve characteristic absorption bands, including features near $760$, $700$, and $540\,\mathrm{cm}^{-1}$. At $4\,\mathrm{cm}^{-1}$ resolution, the retrieved transmission agrees with conventional FTIR within statistical uncertainty over most of the range. The demonstration requires approximately 42 hours for 20 reference–sample pairs at each angle, versus approximately one minute for the conventional FTIR comparison. Interference visibility is approximately 7% at the reported $61^\circ$ setting, and baseline noise is close to shot noise. Limited idler collection, spatial-mode mismatch, broadband dispersion, and momentum–spectrum correlations constrain performance. Thicker crystals could increase brightness but narrow bandwidth and aggravate walk-off and dispersion; larger-acceptance optics and spectral-domain acquisition are discussed as possible improvements.
Written by gpt-6.1-sol from the paper's full text. Automated: it can be wrong.
- Comments
- Submitted to Optica Quantum on 6 July 2026