arXiv:1607.04980 · Review of Scientific Instruments

Cryogenic setup for trapped ion quantum computing

M. F. Brandl, M. W. van Mourik, L. Postler, A. Nolf, K. Lakhmanskiy, R. R. Paiva, S. Möller, N. Daniilidis, H. Häffner, V. Kaushal, T. Ruster, C. Warschburger, H. Kaufmann, U. G. Poschinger, F. Schmidt-Kaler, P. Schindler, T. Monz, R. Blatt

Published Nov 1, 2016• 81 citations
arXivHTMLPDF

Abstract

We report on the design of a cryogenic setup for trapped ion quantum computing containing a segmented surface electrode trap. The heat shield of our cryostat is designed to attenuate alternating magnetic field noise, resulting in 120 dB reduction of 50 Hz noise along the magnetic field axis. We combine this efficient magnetic shielding with high optical access required for single ion addressing as well as for efficient state detection by placing two lenses each with numerical aperture 0.23 inside the inner heat shield. The cryostat design incorporates vibration isolation to avoid decoherence of optical qubits due to the motion of the cryostat. We measure vibrations of the cryostat of less than ±20 nm over 2 s. In addition to the cryogenic apparatus, we describe the setup required for an operation with 40Ca+ and 88Sr+ ions. The instability of the laser manipulating the optical qubits in 40Ca+ is characterized by yielding a minimum of its Allan deviation of 2.4 ⋅ 10−15 at 0.33 s. To evaluate the performance of the apparatus, we trapped 40Ca+ ions, obtaining a heating rate of 2.14(16) phonons/s and a Gaussian decay of the Ramsey contrast with a 1/e-time of 18.2(8) ms.

DOI
10.1063/1.4966970

Related papers

Design and characterization of a cryogenic vacuum chamber for ion trapping experiments

Darian M. Hartsell, J. M. Gray, Chris Shappert +5 more

We present the design and characterization of a cryogenic vacuum chamber incorporating mechanical isolation from vibrations, a high numerical-aperture in-vacuum imaging objective, in-vacuum magnetic shielding, and an antenna for global radio frequency manipulation of trapped ions. The cold shield near 4 K is mechanically referenced to an underlying optical table via thermally insulating supports and exhibits root mean square vibrations less than 7.61(4) nm. Using the in-vacuum objective, we can detect 397 nm photons from a trapped 40Ca+ ion with 1.77% efficiency and achieve 99.9963(4)% single-shot state-detection fidelity in 50 μs. To characterize the efficacy of the magnetic shields, we perform Ramsey experiments on the ground-state qubit and obtain a coherence time of 24(2) ms, which extends to 0.25(1) s with a single spin-echo pulse. XY4 and XY32 dynamical decoupling sequences driven via the radio frequency antenna extend the coherence to 0.72(2) and 0.81(3) s, respectively.

arXiv:2510.01557•Applied Physics Letters•Jan 19, 2026•2 citations

Ultra-low vibration closed-cycle cryogenic surface-electrode ion trap apparatus

T. Dubielzig

Trapped ions are one of the leading platforms for building scalable quantum information processors. Current ion traps run on less than 100 qubits. A useful quantum computer needs to operate thousands or even millions of qubits. If systems reach gate fidelities that allow for quantum error correction, gate errors below the physical fidelity can be reached, which is important for scaling. The most promising platform for scaling up ion-based quantum computers are surface electrode traps in connection with the so-called QCCD (Quantum Charge Coupled Device) architecture. In an evolution of the standard (laser-based) approach to quantum logic gates, gates can be performed by applying an oscillating magnetic near-field gradient to the ions. Current fidelities are about one order of magnitude away from reaching the fault tolerant threshold with no fundamental limitations in sight. Scaling QCCDs is impeded by heating rates of the ions as well as vacuum quality. Ions have to be stored for at least as long as it takes for an algorithm to run and subsequent readout. Collisions with background gas and heating can expel ions from the trap. The microwave near-field approach of driving gates benefits from ions that are trapped close to the trap surface. Unfortunately, heating becomes more prominent near the surface. A cryogenic environment reduces the background pressure by several orders of magnitude and the kinetic energy of gas molecules by two orders of magnitude. Heating rates in a cryogenic trap are about two orders of magnitude lower than in a room temperature trap with otherwise equal physical properties. To operate a QCCD processor, it is required to employ phase-stable Raman laser beams. The cryogenic design of the apparatus needs to take into account the resulting requirements for ultra-low vibration amplitudes. This is particularly relevant for closed-cycle cooling systems, which are desirable from an economic point of view compared to liquid cryostats that boil off helium, but which also feature moving mechanical parts that can introduce a serious amount of vibrations. Here, we present a vibration isolated closed cycle Gifford-McMahon cryocooler with record low vibration amplitudes of 29 nm (RMS=7.8 nm) in the vertical and 51 nm (RMS=13.5 nm) in the horizontal direction. It contains a self-sustained inner vacuum chamber that houses the trap. It features 100 DC connections, 10 of which are able to carry up to 1 A of current and 8 high frequency lines. At 5 K, we can apply continuous 3.4 W of microwave power continuously at 1 GHz to the trap electrodes without heating the system. This is about 5.8 times more than state of the art near-field gates require. We also present the laser systems required to for single-shot ablation loading and Doppler cooling of Be ions: a 70 mW 313 nm Dopper cooling laser, a 1064 nm ns-pulse laser for ablation of neutral beryllium atoms and a 235 nm laser for ionizing them. The detection system features an in vacuum imaging system on a cryogenic 3D translation stage that is able to scan and image the complete surface of an ion trap. The light can be detected on a photomultiplier tube or spatially resolved on an EMCCD camera. Our trap features a meander structure for driving gates and can trap up to 5 ions of which we, so far, did not lose any due to background gas collisions. This hints at a vacuum pressure of 1e-12 mbar or lower. The demonstrated system could hold a cryogenic QCCD chip with a few 10 qubits

•Jan 1, 2021•26 citations

High-Stability Cryogenic System for Quantum Computing With Compact Packaged Ion Traps

Robert Spivey, I. V. Inlek, Zhubing Jia +9 more

Cryogenic environments benefit ion trapping experiments by offering lower motional heating rates, collision energies, and an ultrahigh vacuum (UHV) environment for maintaining long ion chains for extended periods of time. Mechanical vibrations caused by compressors in closed-cycle cryostats can introduce relative motion between the ion and the wavefronts of lasers used to manipulate the ions. Here, we present a novel ion trapping system where a commercial low-vibration closed-cycle cryostat is used in a custom monolithic enclosure. We measure mechanical vibrations of the sample stage using an optical interferometer, and observe a root-mean-square relative displacement of 2.4 nm and a peak-to-peak displacement of 17 nm between free-space beams and the trapping location. We packaged a surface ion trap in a cryopackage assembly that enables easy handling while creating a UHV environment for the ions. The trap cryopackage contains activated carbon getter material for enhanced sorption pumping near the trapping location, and source material for ablation loading. Using$^{171}$Yb$^{+}$as our ion, we estimate the operating pressure of the trap as a function of package temperature using phase transitions of zig-zag ion chains as a probe. We measured the radial mode heating rate of a single ion to be 13 quanta/s on average. The Ramsey coherence measurements yield 330-ms coherence time for counter-propagating Raman carrier transitions using a 355-nm mode-locked pulse laser, demonstrating the high optical stability.

arXiv:2108.05290•IEEE Transactions on Quantum Engineering•Nov 8, 2021•37 citations

A Cryogenic Linear Paul Trap for Quantum Simulation

Paul Hess, Harvey Kaplan, Aaron Lee +5 more

statistics. We present progress towards the construction of a cryogenic ion trap apparatus, designed to cryopump background gas within the 4K chamber. The resulting reduction in pressure will allow robust trapping of up to 100 ions in a single chain. Cooling is provided by a closed cycle cryostat with a gas mediated thermal linkage which mechanically decouples the ion trap from the vibrating cold head. A spherical octagon surrounding the ion trap allows optical access for global and individual addressing beams and high numerical aperture uorescence collection.

Bulletin of the American Physical Society•Jun 11, 2015•0 citations

Stability of ion chains in a cryogenic surface-electrode ion trap

Grahame Vittorini

Cold, trapped atomic ions have enabled the investigation of fundamental physics and generated a rich field of applications. Foremost among these is quantum computation which has recently driven the development of the sophisticated, scalable surface-electrode trap. Despite the many advantages of surface-electrode traps, the typically smaller ion-electrode distance, d, in these traps results in an increased ion heating rate that is proportional to d^(-4) and a decreased trap well-depth that is proportional to d^(-2). These shortcomings can be simultaneously addressed by installing the trap into a cryogenic environment. With this in mind, a closed-cycle, cryogenic ion trapping apparatus that maintains excellent vacuum, is highly modular, has increased optical access, and uses a simple vibration isolation system has been developed. \n \nSingle ions are trapped and used to characterize system properties such as the motion of the vibration isolation stage. In order to compare this system to a similar room temperature apparatus, the ion trapping lifetime and heating rate are determined. A single ion also serves as a sensitive electric field probe that is used to measure and compensate stray electric fields across the trap. \n \nDue to the long dark ion lifetimes in this system, it is well-suited to probing the stability of small, linear ion crystals. Linear ion crystals of arbitrary length are built in an automated fashion using transport waveforms and the scaling of dark lifetime with ion number for N <= 6 is investigated. These data are then used to consider the relevance of various loss channels.

Bulletin of the American Physical Society•Sep 10, 2013•1 citation

Surface-electrode ion traps for scalable quantum computing

D. T. C. Allcock

The major challenges in trapped-ion quantum computation are to scale up few-ion experiments to many qubits and to improve control techniques so that quantum logic gates can be carried out with higher fidelities. This thesis reports experimental progress in both of these areas. In the early part of the thesis we describe the fabrication of a surface-electrode ion trap, the development of the apparatus and techniques required to operate it and the successful trapping of 40Ca+ ions. Notably we developed methods to control the orientation of the principal axes and to minimise ion micromotion. We propose a repumping scheme that simplifies heating rate measurements for ions with low-lying D levels, and use it to characterise the electric field noise in the trap. Surface-electrode traps are important because they offer a route to dense integration of electronic and optical control elements using existing microfabrication technology. We explore this scaling route by testing a series of three traps that were microfabricated at Sandia National Laboratories. Investigations of micromotion and charging of the surface by laser beams were carried out and improvements to future traps are suggested. Using one of these traps we also investigated anomalous electrical noise from the electrode surfaces and discovered that it can be reduced by cleaning with a pulsed laser. A factor of two decrease was observed; this represents the first in situ removal of this noise source, an important step towards higher gate fidelities. In the second half of the thesis we describe the design and construction of an experiment for the purpose of replacing laser-driven multi-qubit quantum logic gates with microwave-driven ones. We investigate magnetic-field-independent hyperfine qubits in 40Ca+ as suitable qubits for this scheme. We make a design study of how best to integrate an ion trap with the microwave conductors required to implement the gate and propose a novel integrated resonant structure. The trap was fabricated and ions were successfully loaded. Single-qubit experiments show that the microwave fields above the trap are in excellent agreement with software simulations. There are good prospects for demonstrating a multi-qubit gate in the near future. We conclude by discussing the possibilities for larger-scale quantum computation by combining microfabricated traps and microwave control.

•Jan 1, 2011•6 citations
We report on the design of a cryogenic setup for trapped ion quantum computing containing a segmented surface electrode trap. The heat shield of our cryostat is designed to attenuate alternating magnetic field noise, resulting in 120 dB reduction of 50 Hz noise along the magnetic field axis. We combine this efficient magnetic shielding with high optical access required for single ion addressing as well as for efficient state detection by placing two lenses each with numerical aperture 0.23 inside the inner heat shield. The cryostat design incorporates vibration isolation to avoid decoherence of optical qubits due to the motion of the cryostat. We measure vibrations of the cryostat of less than ±20 nm over 2 s. In addition to the cryogenic apparatus, we describe the setup required for an operation with 40Ca+ and 88Sr+ ions. The instability of the laser manipulating the optical qubits in 40Ca+ is characterized by yielding a minimum of its Allan deviation of 2.4 ⋅ 10−15 at 0.33 s. To evaluate the performance of the apparatus, we trapped 40Ca+ ions, obtaining a heating rate of 2.14(16) phonons/s and a Gaussian decay of the Ramsey contrast with a 1/e-time of 18.2(8) ms.
We present the design and characterization of a cryogenic vacuum chamber incorporating mechanical isolation from vibrations, a high numerical-aperture in-vacuum imaging objective, in-vacuum magnetic shielding, and an antenna for global radio frequency manipulation of trapped ions. The cold shield near 4 K is mechanically referenced to an underlying optical table via thermally insulating supports and exhibits root mean square vibrations less than 7.61(4) nm. Using the in-vacuum objective, we can detect 397 nm photons from a trapped 40Ca+ ion with 1.77% efficiency and achieve 99.9963(4)% single-shot state-detection fidelity in 50 μs. To characterize the efficacy of the magnetic shields, we perform Ramsey experiments on the ground-state qubit and obtain a coherence time of 24(2) ms, which extends to 0.25(1) s with a single spin-echo pulse. XY4 and XY32 dynamical decoupling sequences driven via the radio frequency antenna extend the coherence to 0.72(2) and 0.81(3) s, respectively.
Trapped ions are one of the leading platforms for building scalable quantum information processors. Current ion traps run on less than 100 qubits. A useful quantum computer needs to operate thousands or even millions of qubits. If systems reach gate fidelities that allow for quantum error correction, gate errors below the physical fidelity can be reached, which is important for scaling. The most promising platform for scaling up ion-based quantum computers are surface electrode traps in connection with the so-called QCCD (Quantum Charge Coupled Device) architecture. In an evolution of the standard (laser-based) approach to quantum logic gates, gates can be performed by applying an oscillating magnetic near-field gradient to the ions. Current fidelities are about one order of magnitude away from reaching the fault tolerant threshold with no fundamental limitations in sight. Scaling QCCDs is impeded by heating rates of the ions as well as vacuum quality. Ions have to be stored for at least as long as it takes for an algorithm to run and subsequent readout. Collisions with background gas and heating can expel ions from the trap. The microwave near-field approach of driving gates benefits from ions that are trapped close to the trap surface. Unfortunately, heating becomes more prominent near the surface. A cryogenic environment reduces the background pressure by several orders of magnitude and the kinetic energy of gas molecules by two orders of magnitude. Heating rates in a cryogenic trap are about two orders of magnitude lower than in a room temperature trap with otherwise equal physical properties. To operate a QCCD processor, it is required to employ phase-stable Raman laser beams. The cryogenic design of the apparatus needs to take into account the resulting requirements for ultra-low vibration amplitudes. This is particularly relevant for closed-cycle cooling systems, which are desirable from an economic point of view compared to liquid cryostats that boil off helium, but which also feature moving mechanical parts that can introduce a serious amount of vibrations. Here, we present a vibration isolated closed cycle Gifford-McMahon cryocooler with record low vibration amplitudes of 29 nm (RMS=7.8 nm) in the vertical and 51 nm (RMS=13.5 nm) in the horizontal direction. It contains a self-sustained inner vacuum chamber that houses the trap. It features 100 DC connections, 10 of which are able to carry up to 1 A of current and 8 high frequency lines. At 5 K, we can apply continuous 3.4 W of microwave power continuously at 1 GHz to the trap electrodes without heating the system. This is about 5.8 times more than state of the art near-field gates require. We also present the laser systems required to for single-shot ablation loading and Doppler cooling of Be ions: a 70 mW 313 nm Dopper cooling laser, a 1064 nm ns-pulse laser for ablation of neutral beryllium atoms and a 235 nm laser for ionizing them. The detection system features an in vacuum imaging system on a cryogenic 3D translation stage that is able to scan and image the complete surface of an ion trap. The light can be detected on a photomultiplier tube or spatially resolved on an EMCCD camera. Our trap features a meander structure for driving gates and can trap up to 5 ions of which we, so far, did not lose any due to background gas collisions. This hints at a vacuum pressure of 1e-12 mbar or lower. The demonstrated system could hold a cryogenic QCCD chip with a few 10 qubits
Cryogenic environments benefit ion trapping experiments by offering lower motional heating rates, collision energies, and an ultrahigh vacuum (UHV) environment for maintaining long ion chains for extended periods of time. Mechanical vibrations caused by compressors in closed-cycle cryostats can introduce relative motion between the ion and the wavefronts of lasers used to manipulate the ions. Here, we present a novel ion trapping system where a commercial low-vibration closed-cycle cryostat is used in a custom monolithic enclosure. We measure mechanical vibrations of the sample stage using an optical interferometer, and observe a root-mean-square relative displacement of 2.4 nm and a peak-to-peak displacement of 17 nm between free-space beams and the trapping location. We packaged a surface ion trap in a cryopackage assembly that enables easy handling while creating a UHV environment for the ions. The trap cryopackage contains activated carbon getter material for enhanced sorption pumping near the trapping location, and source material for ablation loading. Using171^{171}171Yb+^{+}+as our ion, we estimate the operating pressure of the trap as a function of package temperature using phase transitions of zig-zag ion chains as a probe. We measured the radial mode heating rate of a single ion to be 13 quanta/s on average. The Ramsey coherence measurements yield 330-ms coherence time for counter-propagating Raman carrier transitions using a 355-nm mode-locked pulse laser, demonstrating the high optical stability.
statistics. We present progress towards the construction of a cryogenic ion trap apparatus, designed to cryopump background gas within the 4K chamber. The resulting reduction in pressure will allow robust trapping of up to 100 ions in a single chain. Cooling is provided by a closed cycle cryostat with a gas mediated thermal linkage which mechanically decouples the ion trap from the vibrating cold head. A spherical octagon surrounding the ion trap allows optical access for global and individual addressing beams and high numerical aperture uorescence collection.
Cold, trapped atomic ions have enabled the investigation of fundamental physics and generated a rich field of applications. Foremost among these is quantum computation which has recently driven the development of the sophisticated, scalable surface-electrode trap. Despite the many advantages of surface-electrode traps, the typically smaller ion-electrode distance, d, in these traps results in an increased ion heating rate that is proportional to d^(-4) and a decreased trap well-depth that is proportional to d^(-2). These shortcomings can be simultaneously addressed by installing the trap into a cryogenic environment. With this in mind, a closed-cycle, cryogenic ion trapping apparatus that maintains excellent vacuum, is highly modular, has increased optical access, and uses a simple vibration isolation system has been developed. \n \nSingle ions are trapped and used to characterize system properties such as the motion of the vibration isolation stage. In order to compare this system to a similar room temperature apparatus, the ion trapping lifetime and heating rate are determined. A single ion also serves as a sensitive electric field probe that is used to measure and compensate stray electric fields across the trap. \n \nDue to the long dark ion lifetimes in this system, it is well-suited to probing the stability of small, linear ion crystals. Linear ion crystals of arbitrary length are built in an automated fashion using transport waveforms and the scaling of dark lifetime with ion number for N <= 6 is investigated. These data are then used to consider the relevance of various loss channels.
The major challenges in trapped-ion quantum computation are to scale up few-ion experiments to many qubits and to improve control techniques so that quantum logic gates can be carried out with higher fidelities. This thesis reports experimental progress in both of these areas. In the early part of the thesis we describe the fabrication of a surface-electrode ion trap, the development of the apparatus and techniques required to operate it and the successful trapping of 40Ca+ ions. Notably we developed methods to control the orientation of the principal axes and to minimise ion micromotion. We propose a repumping scheme that simplifies heating rate measurements for ions with low-lying D levels, and use it to characterise the electric field noise in the trap. Surface-electrode traps are important because they offer a route to dense integration of electronic and optical control elements using existing microfabrication technology. We explore this scaling route by testing a series of three traps that were microfabricated at Sandia National Laboratories. Investigations of micromotion and charging of the surface by laser beams were carried out and improvements to future traps are suggested. Using one of these traps we also investigated anomalous electrical noise from the electrode surfaces and discovered that it can be reduced by cleaning with a pulsed laser. A factor of two decrease was observed; this represents the first in situ removal of this noise source, an important step towards higher gate fidelities. In the second half of the thesis we describe the design and construction of an experiment for the purpose of replacing laser-driven multi-qubit quantum logic gates with microwave-driven ones. We investigate magnetic-field-independent hyperfine qubits in 40Ca+ as suitable qubits for this scheme. We make a design study of how best to integrate an ion trap with the microwave conductors required to implement the gate and propose a novel integrated resonant structure. The trap was fabricated and ions were successfully loaded. Single-qubit experiments show that the microwave fields above the trap are in excellent agreement with software simulations. There are good prospects for demonstrating a multi-qubit gate in the near future. We conclude by discussing the possibilities for larger-scale quantum computation by combining microfabricated traps and microwave control.