arXiv:0905.3147 · Journal of Physics B Atomic Molecular and Optical Physics

Positioning of the rf potential minimum line of a linear Paul trap with micrometer precision

Peter F. Herskind, Aurélien Dantan, Magnus Albert, Joan P. Marler, Michael Drewsen

Published Jul 15, 2009• 39 citations
arXivHTMLPDF

Abstract

We demonstrate a general technique to achieve a precise radial displacement of the nodal line of the radiofrequency (rf) field in a linear Paul trap. The technique relies on selective adjustment of the load capacitance of the trap electrodes, achieved through the addition of capacitors to the basic resonant rf-circuit used to drive the trap. Displacements of up to 100 micrometer with micrometer precision are measured using a combination of fluorescence images of ion Coulomb crystals and coherent coupling of such crystals to a mode of an optical cavity. The displacements are made without measurable distortion of the shape or structure of the Coulomb crystals, as well as without introducing excess heating commonly associated with the radial displacement of crystals by adjustment through static potentials. We expect this technique to be of importance for future developments of microtrap architectures and ion-based cavity QED.

DOI
10.1088/0953-4075/42/15/154008

Related papers

Measurement of ion displacement via RF power variation for excess micromotion compensation

Ryoichi Saito, Kota Saito, Takashi Mukaiyama

We demonstrate a method of micromotion minimization of a trapped ion in a linear Paul trap based on the precision measurement of the ion trapping position displacement due to a stray electric field in the radial plane by ion fluorescence imaging. The amount of displacement in the radial plane is proportional to the strength of a stray electric field. Therefore, we evaluated the micromotion compensation condition by measuring the ion displacements from the ion equilibrium position using two different radial trap frequencies with various combinations of the compensation voltage. The residual electric field uncertainty of this technique reached a few volts per meter. This compensation technique does not depend on the orientation of the incident cooling laser or the detuning and imaging direction. Therefore, this method is suitable for a planar ion trap, a stylus ion trap, which limits the propagation angle of lasers, or miniaturized ion trap systems for sensing and metrological applications.

arXiv:2102.00788•Journal of Applied Physics•Mar 25, 2021•10 citations

Ion-trajectory analysis for micromotion minimization and the measurement of small forces

Timm F. Gloger, Peter Kaufmann, Delia Kaufmann +4 more

For experiments with ions confined in a Paul trap, minimization of micromotion is often essential. In order to diagnose and compensate micromotion we have implemented a method that allows for finding the position of the radio-frequency (rf) null reliably and efficiently, in principle, without any variation of direct current (dc) voltages. We apply a trap modulation technique and focus-scanning imaging to extract three-dimensional ion positions for various rf drive powers and analyze the power dependence of the equilibrium position of the trapped ion. In contrast to commonly used methods, the search algorithm directly makes use of a physical effect as opposed to efficient numerical minimization in a high-dimensional parameter space. Using this method we achieve a compensation of the residual electric field that causes excess micromotion in the radial plane of a linear Paul trap down to $0.09\phantom{\rule{4pt}{0ex}}\text{V}\phantom{\rule{0.16em}{0ex}}{\text{m}}^{\ensuremath{-}1}$. Additionally, the precise position determination of a single harmonically trapped ion employed here can also be utilized for the detection of small forces. This is demonstrated by determining light pressure forces with a precision of 135 yN. As the method is based on imaging only, it can be applied to several ions simultaneously and is independent of laser direction and thus well suited to be used with, for example, surface-electrode traps.

arXiv:1503.07031•Physical Review A•Oct 27, 2015•45 citations

Precise positioning of an ion in an integrated Paul trap-cavity system using radiofrequency signals

Ezra Kassa, Hiroki Takahashi, Costas Christoforou +1 more

We report a novel miniature Paul ion trap design with an integrated optical fibre cavity which can serve as a building block for a fibre-linked quantum network. In such cavity quantum electrodynamic set-ups, the optimal coupling of the ions to the cavity mode is of vital importance and this is achieved by moving the ion relative to the cavity mode. The trap presented herein features an endcap-style design complemented with extra electrodes on which additional radiofrequency voltages are applied to fully control the pseudopotential minimum in three dimensions. This method lifts the need to use three-dimensional translation stages for moving the fibre cavity with respect to the ion and achieves high integrability, mechanical rigidity and scalability. Not based on modifying the capacitive load of the trap, this method leads to precise control of the pseudopotential minimum allowing the ion to be moved with precisions limited only by the ion’s position spread. We demonstrate this by coupling the ion to the fibre cavity and probing the cavity mode profile.

arXiv:1712.04011•Journal of Modern Optics•Dec 6, 2017•10 citations

Ion-trap cavity QED system for probabilistic entanglement

Nicolas Seymour-Smith

Laser systems and a linear radiofrequency (rf) Paul trap with an integrated co-axial cavity have been developed for experiments in cavity QED and probabilistic entanglement. Single 40Ca+ ions and large Coulomb crystals have been trapped routinely and laser cooled with long trapping lifetimes. A technique to achieve precise overlap of the pseudopotential minimum of the rf-field with the cavity mode has been implemented through variable capacitors in the resonant rf-circuit used to drive the trap. Three-dimensional micromotion compensation has been implemented. An 894 nm laser has been frequency stabilised to a Pound-Drever-Hall cavity which is in turn stabilised to atomic Cs using polarisation spectroscopy. The Allan variance of the error signal has been reduced to less than a kilohertz on timescales greater than a second. A novel implementation of the scanning cavity transfer lock has been developed to transfer the stability of the 894 nm laser to the 397 nm ion cooling and 866 nm repumping lasers. The bandwidth of the system has been increased to 380 Hz and the Allan variance of the error signal has been reduced to less than ten kilohertz on timescales of greater than a second. The pseudopotential minimum of the rf field has been overlapped optimally with the optical cavity mode through mapping of the fluorescence from cavity-field repumped ions as a function of their displacement. Coupling to the cavity mode has been confirmed by observation of resonant fluorescence into the cavity mode using the cavity-assisted Raman transition process. The thesis demonstrates that the setup is ready for the controlled production of single photons with pre-determined polarisation states, and progression onto new schemes to entangle multiple ions that are coupled to the optical cavity mode.

•May 21, 2012•1 citation

A compact radiofrequency drive based on interdependent resonant circuits for precise control of ion traps

Amelia Detti, Marco De Pas, Lucia Duca +2 more

Paul traps are widely used to confine electrically charged particles like atomic and molecular ions by using an intense radiofrequency (RF) field, typically obtained by a voltage drop on capacitative electrodes placed in vacuum. We present a RF drive realized on a compact printed circuit board and providing a high-voltage RF signal to a quadrupole Paul trap. The circuit is formed by using four interdependent resonant circuits - each of which is connected to an electrode of a Paul trap - fed by low-noise amplifiers, leading to an output voltage of peak-to-peak amplitude up to 200 V at 3.23 MHz. The presence of a single resonant circuit for each electrode ensures a strong control on the voltage drop on each electrode, e.g., by applying a DC field through a bias tee. Additionally, the moderate quality factor Q = 67 of the resonant circuits ensures a fast operation of the drive, which can be turned on and off in less than 10 μs. Finally, the RF lines are equipped with pickups that sample the RF in phase and amplitude, thus providing a signal that can be used to actively control the voltage drop at the trap's electrodes.

arXiv:1812.02018•Review of Scientific Instruments•Feb 1, 2019•9 citations

Endcap-type Paul trap for precision spectroscopy and studies of ion-crystal dynamics

Anand Prakash, Akhil Ayyadevara, E. Krishnakumar +5 more

We present the design and fabrication of an endcap-type Paul trap. The trap is designed for studies with Ca+ and Yb+. The design, fabrication process, and characterization are presented in detail, with a focus on trapping a single compensated ion at the rf node. A custom-built imaging system of NA = 0.14 and magnification of ≈22× performs close to the diffraction limit and resolves multi-ion clusters. Controlled ion loading and characterization of the trap are performed using 40Ca+. The experimentally determined quadrupole coefficient of the trap is ≈0.3, which is very close to the design value. The relative frequency shift along the spectroscopy beam due to excess micromotion is at the level of 3.5 × 10−18 for 40Ca+. Applications of this trap encompass single-ion-based optical frequency standards, tests of fundamental physics, the study of mesoscopic Coulomb clusters, and the controlled interaction of a single ion with co-trapped atoms.

arXiv:2601.07328•AIP Advances•Jul 1, 2026•0 citations
We demonstrate a general technique to achieve a precise radial displacement of the nodal line of the radiofrequency (rf) field in a linear Paul trap. The technique relies on selective adjustment of the load capacitance of the trap electrodes, achieved through the addition of capacitors to the basic resonant rf-circuit used to drive the trap. Displacements of up to 100 micrometer with micrometer precision are measured using a combination of fluorescence images of ion Coulomb crystals and coherent coupling of such crystals to a mode of an optical cavity. The displacements are made without measurable distortion of the shape or structure of the Coulomb crystals, as well as without introducing excess heating commonly associated with the radial displacement of crystals by adjustment through static potentials. We expect this technique to be of importance for future developments of microtrap architectures and ion-based cavity QED.
We demonstrate a method of micromotion minimization of a trapped ion in a linear Paul trap based on the precision measurement of the ion trapping position displacement due to a stray electric field in the radial plane by ion fluorescence imaging. The amount of displacement in the radial plane is proportional to the strength of a stray electric field. Therefore, we evaluated the micromotion compensation condition by measuring the ion displacements from the ion equilibrium position using two different radial trap frequencies with various combinations of the compensation voltage. The residual electric field uncertainty of this technique reached a few volts per meter. This compensation technique does not depend on the orientation of the incident cooling laser or the detuning and imaging direction. Therefore, this method is suitable for a planar ion trap, a stylus ion trap, which limits the propagation angle of lasers, or miniaturized ion trap systems for sensing and metrological applications.
For experiments with ions confined in a Paul trap, minimization of micromotion is often essential. In order to diagnose and compensate micromotion we have implemented a method that allows for finding the position of the radio-frequency (rf) null reliably and efficiently, in principle, without any variation of direct current (dc) voltages. We apply a trap modulation technique and focus-scanning imaging to extract three-dimensional ion positions for various rf drive powers and analyze the power dependence of the equilibrium position of the trapped ion. In contrast to commonly used methods, the search algorithm directly makes use of a physical effect as opposed to efficient numerical minimization in a high-dimensional parameter space. Using this method we achieve a compensation of the residual electric field that causes excess micromotion in the radial plane of a linear Paul trap down to $0.09\phantom{\rule{4pt}{0ex}}\text{V}\phantom{\rule{0.16em}{0ex}}{\text{m}}^{\ensuremath{-}1}$. Additionally, the precise position determination of a single harmonically trapped ion employed here can also be utilized for the detection of small forces. This is demonstrated by determining light pressure forces with a precision of 135 yN. As the method is based on imaging only, it can be applied to several ions simultaneously and is independent of laser direction and thus well suited to be used with, for example, surface-electrode traps.
We report a novel miniature Paul ion trap design with an integrated optical fibre cavity which can serve as a building block for a fibre-linked quantum network. In such cavity quantum electrodynamic set-ups, the optimal coupling of the ions to the cavity mode is of vital importance and this is achieved by moving the ion relative to the cavity mode. The trap presented herein features an endcap-style design complemented with extra electrodes on which additional radiofrequency voltages are applied to fully control the pseudopotential minimum in three dimensions. This method lifts the need to use three-dimensional translation stages for moving the fibre cavity with respect to the ion and achieves high integrability, mechanical rigidity and scalability. Not based on modifying the capacitive load of the trap, this method leads to precise control of the pseudopotential minimum allowing the ion to be moved with precisions limited only by the ion’s position spread. We demonstrate this by coupling the ion to the fibre cavity and probing the cavity mode profile.
Laser systems and a linear radiofrequency (rf) Paul trap with an integrated co-axial cavity have been developed for experiments in cavity QED and probabilistic entanglement. Single 40Ca+ ions and large Coulomb crystals have been trapped routinely and laser cooled with long trapping lifetimes. A technique to achieve precise overlap of the pseudopotential minimum of the rf-field with the cavity mode has been implemented through variable capacitors in the resonant rf-circuit used to drive the trap. Three-dimensional micromotion compensation has been implemented. An 894 nm laser has been frequency stabilised to a Pound-Drever-Hall cavity which is in turn stabilised to atomic Cs using polarisation spectroscopy. The Allan variance of the error signal has been reduced to less than a kilohertz on timescales greater than a second. A novel implementation of the scanning cavity transfer lock has been developed to transfer the stability of the 894 nm laser to the 397 nm ion cooling and 866 nm repumping lasers. The bandwidth of the system has been increased to 380 Hz and the Allan variance of the error signal has been reduced to less than ten kilohertz on timescales of greater than a second. The pseudopotential minimum of the rf field has been overlapped optimally with the optical cavity mode through mapping of the fluorescence from cavity-field repumped ions as a function of their displacement. Coupling to the cavity mode has been confirmed by observation of resonant fluorescence into the cavity mode using the cavity-assisted Raman transition process. The thesis demonstrates that the setup is ready for the controlled production of single photons with pre-determined polarisation states, and progression onto new schemes to entangle multiple ions that are coupled to the optical cavity mode.
Paul traps are widely used to confine electrically charged particles like atomic and molecular ions by using an intense radiofrequency (RF) field, typically obtained by a voltage drop on capacitative electrodes placed in vacuum. We present a RF drive realized on a compact printed circuit board and providing a high-voltage RF signal to a quadrupole Paul trap. The circuit is formed by using four interdependent resonant circuits - each of which is connected to an electrode of a Paul trap - fed by low-noise amplifiers, leading to an output voltage of peak-to-peak amplitude up to 200 V at 3.23 MHz. The presence of a single resonant circuit for each electrode ensures a strong control on the voltage drop on each electrode, e.g., by applying a DC field through a bias tee. Additionally, the moderate quality factor Q = 67 of the resonant circuits ensures a fast operation of the drive, which can be turned on and off in less than 10 μs. Finally, the RF lines are equipped with pickups that sample the RF in phase and amplitude, thus providing a signal that can be used to actively control the voltage drop at the trap's electrodes.
We present the design and fabrication of an endcap-type Paul trap. The trap is designed for studies with Ca+ and Yb+. The design, fabrication process, and characterization are presented in detail, with a focus on trapping a single compensated ion at the rf node. A custom-built imaging system of NA = 0.14 and magnification of ≈22× performs close to the diffraction limit and resolves multi-ion clusters. Controlled ion loading and characterization of the trap are performed using 40Ca+. The experimentally determined quadrupole coefficient of the trap is ≈0.3, which is very close to the design value. The relative frequency shift along the spectroscopy beam due to excess micromotion is at the level of 3.5 × 10−18 for 40Ca+. Applications of this trap encompass single-ion-based optical frequency standards, tests of fundamental physics, the study of mesoscopic Coulomb clusters, and the controlled interaction of a single ion with co-trapped atoms.